Agile™ Esophageal M00517240Recalls & reported problems
Boston Scientific Corporation · Partially Covered Stent System · Class 2 · Implant
All models of Agile Esophageal (16)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729972945), received between Nov 24, 2020 and Nov 16, 2023 — classified as 3 injuries.
The most frequently reported specific device problem is “Human-Device Interface Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Human-Device Interface Problem2
- Activation, Positioning or Separation Problem1
- Device Stenosis1
- Difficult or Delayed Positioning1
- Material Deformation1
- Positioning Problem1
- Use of Device Problem1
- Wrong Label1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Dysphagia/ Odynophagia1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryNov 16, 2023Human-Device Interface Problem; Device Stenosis; Human-Device Interface Problem; Device Stenosis
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AGILE ESOPHAGEAL PARTIALLY COVERED STENT WAS IMPLANTED IN THE MIDDLE ESOPHAGUS TO TREAT A MALIGNANT STRICTURE DURING A STENT PLACEMENT PROCEDURE PERFORMED ON AN UNKNOWN DATE. THE PATIENT'S ANATOMY WAS TORTUOUS AND WAS NOT DILATED PRIOR TO STENT PLACEMENT. ON OCTOBER 27, 2023, DURING A ROUTINE FOLLOW-UP, IT WAS OBSERVED THAT THERE WAS AN INGROWTH ON THE UNCOVERED PART OF THE STENT. THE PHYSICIAN NOTICED THAT THE STENT INGROWTH OCCURRED MUCH EARLIER THAN WITH THE WALLFLEX ESOPHAGEAL STENT. SUBSEQUENTLY, TO RESTORE PATENCY, A WALLFLEX ESOPHAGEAL STENT WAS PLACED, AND THE PROCEDURE WAS COMPLETED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION RECEIVED ON JANUARY 05, 2024. NOTE: THIS REPORT PERTAINS TO ONE OF TWO DEVICES USED DURING THE SAME PROCEDURE. REFER TO MANUFACTURER REPORTS # 3005099803-2023-06204 AND 3005099803-2024-00099 FOR THE ASSOCIATED DEVICE INFORMATION. ON (B)(6) 2023, THE FIRST AGILE ESOPHAGEAL STENT WAS IMPLANTED (THE SUBJECT OF MFR. REPORT # 3005099803-2024-00099); HOWEVER, ON AN UNKNOWN DATE, THE PATIENT PRESENTED WITH DYSPHAGIA WHICH PROMPTED THE PHYSICIAN TO CHE
Patient problems: Dysphagia/ Odynophagia; No Clinical Signs, Symptoms or Conditions
MAUDE report 18155469 ↗InjuryFeb 24, 2023Wrong Label; Wrong Label
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON FEBRUARY 02, 2023 THAT AN AGILE ESOPHAGEAL PARTIALLY COVERED STENT WAS TO BE IMPLANTED TO TREAT A 7CM MALIGNANT STRICTURE IN THE LOWER ESOPHAGUS DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) WITH STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2023. THE PATIENT'S ANATOMY WAS NOT TORTUOUS AND WAS NOT DILATED PRIOR TO STENT PLACEMENT. DURING THE PROCEDURE, THE AGILE ESOPHAGEAL STENT WAS SUCCESSFULLY PLACED INSIDE THE PATIENT; HOWEVER, IT WAS NOTICED THAT THE STENT WAS UNABLE TO PROVIDE THE NEEDED COVERAGE. SUBSEQUENTLY, THE AGILE ESOPHAGEAL STENT WAS REMOVED WITH FORCEPS AND THE PROCEDURE WAS RESCHEDULED. UPON CHECKING, A 10MM X 6CM AGILE ESOPHAGEAL STENT WAS INCORRECTLY LABELED AS 18MM X 12.5CM AGILE ESOPHAGEAL STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16436213 ↗InjuryNov 24, 2020Difficult or Delayed Positioning; Use of Device Problem; Activation, Positioning or Separation Problem; Human-Device Interface Problem; Material Deformation; Positioning Problem; Difficult or Delayed Positioning; Use of Device Problem; Activation, Positioning or Separation Problem; Human-Device Interface Problem; Material Deformation; Positioning Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2020 THAT AN AGILE ESOPHAGEAL PARTIALLY COVERED STENT WAS IMPLANTED BELOW THE CRICOPHARYNGEAL MUSCLE TO TREAT AN ESOPHAGEAL FISTULA. REPORTEDLY, THE PATIENT'S ANATOMY WAS NOT TORTUOUS AND WAS NOT DILATED PRIOR TO STENT PLACEMENT PROCEDURE. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, WHILE ATTEMPTING TO DEPLOY THE STENT, STRONG RESISTANCE WAS ENCOUNTERED. REPORTEDLY, THE DELIVERY SYSTEM WAS IN FULLY DEPLOYED POSITION; HOWEVER, THE STENT WAS PARTIALLY DEPLOYED. REPORTEDLY, THE STENT WAS PARTIALLY DEPLOYED ON THE DELIVERY SYSTEM WHEN IT WAS REMOVED FROM THE PATIENT AND THE STENT WAS THEN RE-CONSTRAINED MANUALLY. THEN THE SCOPE AND STENT WERE RE-INSERTED BACK INTO THE PATIENT AND THE STENT WAS FULLY DEPLOYED. THE PHYSICIAN THEN USED A RAT TOOTH FORCEP TO ADJUST THE STENT INTO THE CORRECT POSITION AND THE PROCEDURE WAS COMPLETED. AFTER THE PROCEDURE, A KINK IN THE DEPLOYMENT CATHETER WAS NOTED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10898255 ↗
Other PROSTHESIS, ESOPHAGEAL devices
- WallFlex™ Esophageal M00516730Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall29 reports
- WallFlex™ Esophageal M00516740Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall1 death22 reports
- WallFlex™ Esophageal M00516750Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall1 death19 reports
- WallFlex™ Esophageal M00516700Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall18 reports
- WallFlex™ Esophageal M00516910Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall15 reports
- WallFlex™ Esophageal M00516930Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall14 reports
- WallFlex™ Esophageal M00516710Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall13 reports
- WallFlex™ Esophageal M00516940Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall12 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page