Agile™ Esophageal M00517400Recalls & reported problems

Boston Scientific Corporation · Fully Covered Stent System · Class 2 · Implant

All models of Agile Esophageal (16)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08714729973027), received between Apr 7, 2021 and Nov 25, 2024 — classified as 3 injuries, 1 malfunction.

The most frequently reported specific device problem is “Premature Activation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Premature Activation2
  • Adverse Event Without Identified Device or Use Problem1
  • Difficult to Remove1
  • Material Deformation1
  • Migration1
  • Positioning Problem1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Discomfort1
  • Fistula1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 25, 2024Premature Activation; Use of Device Problem; Premature Activation; Use of Device Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AGILE ESOPHAGEAL FULLY COVERED STENT WAS TO BE IMPLANTED IN THE ESOPHAGUS TO TREAT SEVERE DYSPHAGIA SECONDARY TO ANASTOMOTIC STRICTURE DURING A STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2024. THE PATIENT'S ANATOMY WAS NOT TORTUOUS AND WAS NOT DILATED PRIOR TO STENT PLACEMENT. DURING THE PROCEDURE, THE STENT WAS DEPLOYED BEFORE REACHING THE "POINT OF NO RETURN" MARKER. THE STENT WAS POSITIONED DISTALLY AT THE SITE OF THE STRICTURE AND WAS REMOVED WITH RETRIEVAL FORCEPS. THE PROCEDURE WAS COMPLETED WITH ANOTHER AGILE ESOPHAGEAL STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. NOTE: IT WAS REPORTED THAT THE STENT WAS TO BE IMPLANTED TO TREAT SEVERE DYSPHAGIA SECONDARY TO ANASTOMOTIC STRICTURE. HOWEVER, PER THE AGILE ESOPHAGEAL FULLY COVERED STENT SYSTEM INSTRUCTIONS FOR USE (IFU), THE STENT IS INTENDED FOR MAINTAINING ESOPHAGEAL LUMINAL PATENCY IN ESOPHAGEAL STRICTURES CAUSED BY INTRINSIC AND/OR EXTRINSIC MALIGNANT TUMORS, AND OCCLUSION OF CONCURRENT ESOPHAGEAL FISTULAS. THE DEVICE IS NOT INDICATED FOR THE TREATMENT OF SEVERE DYSPHAGIA SECONDARY TO ANASTOMOTIC STRICTURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20775540 ↗
  • MalfunctionSep 26, 2023Difficult to Remove; Difficult to Remove

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AGILE ESOPHAGEAL STENT WAS TO BE IMPLANTED TO TREAT A 2CM MALIGNANT STRICTURE IN THE UPPER ESOPHAGUS DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE PERFORMED ON (B)(6) 2023. THE PATIENT'S ANATOMY WAS NOT TORTUOUS AND WAS NOT DILATED PRIOR TO STENT PLACEMENT. DURING THE PROCEDURE, THE STENT WAS FULLY DEPLOYED BUT GOT CAUGHT ON THE DELIVERY SYSTEM DURING REMOVAL, WHICH RESULTED IN THE STENT MOVING OUT OF ITS POSITION. THE STENT WAS REMOVED ALONG WITH THE DELIVERY SYSTEM. THE PROCEDURE WAS COMPLETED USING ANOTHER AGILE ESOPHAGEAL STENT. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17818603 ↗
  • InjurySep 14, 2023Premature Activation; Material Deformation; Positioning Problem; Premature Activation; Material Deformation; Positioning Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AGILE ESOPHAGEAL STENT WAS TO BE IMPLANTED TO TREAT AN ANASTOMATIC STRICTURE IN THE ESOPHAGUS DURING AN ESOPHAGEAL STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2023. DURING THE PROCEDURE, THE STENT PREMATURELY DEPLOYED BEFORE THE DISTAL HANDLE WAS COMPLETELY PULLED BACK, WHICH CAUSED THE STENT TO BECOME DEFORMED AND MIGRATE INTO THE PATIENT'S STOMACH. THE STENT WAS REMOVED, AND THE PHYSICIAN COMPLETED THE PROCEDURE BY ADMINISTERING STEROIDS AND USING A CRE BALLOON. THE PATIENT WILL BE RESCHEDULED FOR ANOTHER STENT PLACEMENT PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17744275 ↗
  • InjuryApr 7, 2021Adverse Event Without Identified Device or Use Problem; Migration; Adverse Event Without Identified Device or Use Problem; Migration

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON MARCH 15, 2021 THAT AN AGILE ESOPHAGEAL FULLY COVERED STENT WAS IMPLANTED TO TREAT A LESS THAN 6CM MALIGNANT STRICTURE IN THE ESOPHAGUS DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) WITH STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2020. POST-STENT PLACEMENT THE PATIENT UNDERWENT RADIATION. REPORTEDLY, THE PATIENT'S ANATOMY WAS TORTUOUS AND WAS NOT DILATED PRIOR TO STENT PLACEMENT. ON MARCH 08, 2021, DURING EMERGENT FOLLOW-UP, THE PATIENT PRESENTED WITH TRACHEOESOPHAGEAL FISTULA (TEF) AND DISCOMFORT. THE TEF STARTED AT THE DISTAL END OF THE STENT. CONSEQUENTLY, THE STENT WAS REMOVED FROM THE PATIENT USING RAT-TOOTH FORCEPS AND A NON-BSC DEVICE WAS IMPLANTED OVER THE TEF TO COMPLETE THE PROCEDURE. IN THE PHYSICIAN'S ASSESSMENT THE DISCOMFORT AND THE TEF WERE RELATED TO THE DISTAL END OF THE STENT AND THE RADIATION TREATMENT. IT WAS REPORTED THAT AN IMAGE SHOWED THAT THE AGILE ESOPHAGEAL STENT MIGRATED IN THE AIRWAY.

    Patient problems: Discomfort; Fistula

    MAUDE report 11627301 ↗

FDA clearance and classification

The FDA regulates product code ESW (prosthesis, esophageal) under 21 CFR 878.3610 as a Class II device. 21 CFR 878.3610 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Agile™ Esophageal M00517400

Has the Agile™ Esophageal M00517400 been recalled?

No FDA recall record in our data names this exact device identifier (DI 08714729973027), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Agile™ Esophageal M00517400?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Apr 7, 2021 and Nov 25, 2024. In those reports, the most common specific device problems are “Premature Activation” (2), “Difficult to Remove” (1) and “Material Deformation” (1); the most common patient problems are “Discomfort” (1) and “Fistula” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Agile™ Esophageal M00517400?

None of the reports are classified as deaths, 3 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Agile™ Esophageal M00517400 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the Agile™ Esophageal M00517400, and is it still sold?

The FDA's device database lists Boston Scientific Corporation as the company responsible for it (brand Agile™ Esophageal, model M00517400). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other PROSTHESIS, ESOPHAGEAL devices

All PROSTHESIS, ESOPHAGEAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page