Skip to content

Agile™ Esophageal M00517470Recalls & reported problems

Boston Scientific Corporation · Fully Covered Stent System · Class 2 · Implant

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729973096), received between Sep 29, 2023 and Jun 28, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Activation, Positioning or Separation Problem” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Activation, Positioning or Separation Problem3
  • Break2
  • Adverse Event Without Identified Device or Use Problem1
  • Detachment of Device or Device Component1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 28, 2024Activation, Positioning or Separation Problem; Detachment of Device or Device Component; Material Deformation; Activation, Positioning or Separation Problem; Detachment of Device or Device Component; Material Deformation

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AGILE ESOPHAGEAL FULLY COVERED STENT WAS TO BE IMPLANTED TO TREAT A MALIGNANT STRICTURE IN THE ESOPHAGUS DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) WITH STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2024. THE PATIENT'S ANATOMY WAS TORTUOUS AND WAS DILATED PRIOR TO STENT PLACEMENT. DURING THE PROCEDURE, THE YELLOW INNER SHEATH BROKE, AND THE STENT COULD NOT BE RESHEATHED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19632273 ↗
  • MalfunctionApr 11, 2024Break; Activation, Positioning or Separation Problem; Adverse Event Without Identified Device or Use Problem; Break; Activation, Positioning or Separation Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AGILE ESOPHAGEAL TTS FULY COVERED STENT WAS TO BE IMPLANTED IN THE ESOPHAGUS TO TREAT A MALIGNANT STRICTURE DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) WITH STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE SHEATH BROKE DURING ATTEMPTED DEPLOYMENT. THE STENT WAS REMOVED FROM THE PATIENT PARTIALLY DEPLOYED AND ANOTHER AGILE ESOPHAGEAL STENT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19089460 ↗
  • MalfunctionSep 29, 2023Break; Activation, Positioning or Separation Problem; Break; Activation, Positioning or Separation Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AGILE ESOPHAGEAL FULLY COVERED STENT WAS USED TO TREAT A MALIGNANT ESOPHAGEAL STRICTURE DURING A STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2023. THE PATIENT'S ANATOMY WAS NOT TORTUOUS AND WAS NOT DILATED PRIOR TO STENT PLACEMENT. DURING THE PROCEDURE, WHEN THE STENT WAS DEPLOYED, THE BLUE CATHETER WAS DETACHED FROM THE DELIVERY SYSTEM. THE STENT WAS PARTIALLY DEPLOYED WHEN IT WAS REMOVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER AGILE ESOPHAGEAL STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17844902 ↗

Other PROSTHESIS, ESOPHAGEAL devices

All PROSTHESIS, ESOPHAGEAL devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.