Ranger™ H74939419400680Recalls & reported problems
Boston Scientific Corporation · Paclitaxel-Coated PTA Balloon Catheter · Class 3
All models of Ranger (50)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 08714729975793), received between Apr 23, 2021 and Mar 26, 2025 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture5
- Leak/Splash1
- Material Integrity Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Obstruction/Occlusion1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 26, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED DISTAL LEFT SUPERFICIAL FEMORAL ARTERY. AFTER PERFORMING AN IPSILATERAL PUNCTURE, A GUIDEWIRE WAS PASSED THROUGH THE LESION FOLLOWED BY ADVANCING A 4.0 X 60MM, 80CM RANGER BALLOON CATHETER FOR DILATATION. HOWEVER, DURING THE FIRST INFLATION AT 6 ATMOSPHERES FOR 10 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A 135CM-SHAFT RANGER BALLOON CATHETER. THE LESION WAS TREATED SUCCESSFULLY. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21699609 ↗MalfunctionMar 18, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 4.0 X 60MM, 80CM RANGER DRUG-COATED BALLOON WAS ADVANCED FOR DILATION. HOWEVER, DURING THE FIRST INFLATION AT 14 ATMOSPHERES FOR 2 MINUTES, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18926775 ↗MalfunctionJul 31, 2023Leak/Splash; Material Rupture; Leak/Splash; Material Rupture
IT WAS REPORTED THAT THE BALLOON RUPTURED AND HAD A PINHOLE. THIS 4.0 X 60MM, 80CM RANGER PACLITAXEL-COATED PTA BALLOON CATHETER WAS SELECTED FOR USE IN AN ENDOVASCULAR THERAPY PROCEDURE. THE 99% STENOSED, MODERATELY CALCIFIED, AND MODERATELY TORTUOUS LESION WAS LOCATED IN THE RIGHT SUPERFICIAL FEMORAL ARTERY. THE BALLOON RUPTURED AND EXHIBITED A PINHOLE SLIGHTLY DISTAL FROM THE BALLOON CATHETER CENTER. THIS WAS NOTED DURING THE FIRST INFLATION AT 8 ATM FOR APPROXIMATELY FIVE SECONDS. THE BALLOON WAS REMOVED INTACT, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THE PHYSICIAN NOTED THIS WAS EFFECTS OF CALCIFICATION. THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17424155 ↗MalfunctionMay 11, 2022Material Rupture; Material Rupture
IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN A MODERATELY TORTUOUS AND MODERATELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 4.0 X 60MM, 80CM RANGER PACLITAXEL-COATED PTA BALLOON CATHETER WAS SELECTED FOR THE ENDOVASCULAR THERAPY PROCEDURE. THE RANGER PACLITAXEL-COATED PTA BALLOON CATHETER WAS ADVANCED FOR DILATION; HOWEVER, DURING THE FIRST INFLATION AT 6 ATMOSPHERES FOR 30 SECONDS INFLATION TIME, THE BALLOON RUPTURED. AFTER DECOMPRESSION, THE DEVICE WAS REMOVED FROM THE BODY AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14360723 ↗MalfunctionApr 29, 2022Material Rupture; Material Rupture
IT WAS REPORTED THAT THE BALLOON RUPTURED. THE 95% STENOSED TARGET LESION WAS MODERATELY CALCIFIED, MILDLY TORTUOUS AND LOCATED IN THE SUPERFICIAL FEMORAL ARTERY (SFA). A RANGER DRUG-COATED BALLOON WAS SELECTED FOR USE. DURING THE FIRST INFLATION, THE BALLOON WAS INFLATED TO 6ATM FOR 10 TEN SECONDS. HOWEVER, THE BALLOON THEN RUPTURED. THE PROCEDURE WAS COMPLETED USING THIS DEVICE. THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14256008 ↗InjuryApr 23, 2021Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT SLOW FLOW OCCURRED. THE LESION LOCATED IN THE SUPERFICIAL FEMORAL ARTERY (SFA) WAS DILATED THREE TIMES WITH A 3.5 MM X 4 CM COYOTE BALLOON. THE LESION WAS THEN DILATED FOR THREE MINUTES WITH THE 4MM X 60MM X 80CM RANGER DRUG COATED BALLOON. AFTERWARDS, WHEN CONTRAST WAS PERFORMED, A SLOW FLOW OF BLOOD WAS OBSERVED. THE PROCEDURE WAS COMPLETED WITH IMPROVED SLOW FLOW BY USING A VASODILATOR. IT WAS THOUGHT THAT THE PACLITAXEL APPLIED TO THE RANGER MAY HAVE COME OFF TO THE PERIPHERY.
Patient problems: Obstruction/Occlusion
MAUDE report 11715800 ↗
Other Drug-Eluting Peripheral Transluminal Angioplasty Catheter devices
- LUTONIX® Drug Coated Balloon LX351305150Lutonix, Inc. · Drug-Eluting Peripheral Transluminal Angioplasty Catheter117 reports
- LUTONIX® Drug Coated Balloon LX351306150Lutonix, Inc. · Drug-Eluting Peripheral Transluminal Angioplasty Catheter75 reports
- Ranger™ H74939419502010Boston Scientific Corporation · Drug-Eluting Peripheral Transluminal Angioplasty Catheter71 reports
- Ranger™ H74939419602010Boston Scientific Corporation · Drug-Eluting Peripheral Transluminal Angioplasty Catheter71 reports
- Lutonix® Drug Coated Balloon LX351305100Lutonix, Inc. · Drug-Eluting Peripheral Transluminal Angioplasty Catheter66 reports
- IN.PACT™ Admiral™ ADM06004013PMedtronic, Inc. · Drug-Eluting Peripheral Transluminal Angioplasty Catheter1 recall12 reports
- IN.PACT™ Admiral™ ADM06015013PMedtronic, Inc. · Drug-Eluting Peripheral Transluminal Angioplasty Catheter1 recall11 reports
- IN.PACT™ Admiral™ ADM04015013PMedtronic, Inc. · Drug-Eluting Peripheral Transluminal Angioplasty Catheter1 recall8 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.