Skip to content

Ranger™ H74939419500410Recalls & reported problems

Boston Scientific Corporation · Paclitaxel-Coated PTA Balloon Catheter · Class 3

All models of Ranger (50)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08714729975922), received between Sep 29, 2022 and Dec 18, 2025 — classified as 3 injuries, 1 malfunction.

The most frequently reported specific device problem is “Material Rupture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Material Rupture1

Patient problems

  • Restenosis2
  • Thrombosis/Thrombus2
  • Arteriosclerosis/ Atherosclerosis1
  • Embolism/Embolus1
  • Fistula1
  • Ischemia1
  • No Clinical Signs, Symptoms or Conditions1
  • Obstruction/Occlusion1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PERFORATION AND FISTULA OCCURRED. THE SUBJECT UNDERWENT TREATMENT WITH THE RANGER DRUG-COATED BALLOON AND ELUVIA DRUG-ELUTING STENTS ON (B)(6) 2021 AS PART OF THE CLINICAL TRIAL. TARGET LESION #001 WAS LOCATED IN THE RIGHT MID SUPERFICIAL FEMORAL ARTERY (SFA), RIGHT DISTAL SFA, AND RIGHT MID POPLITEAL ARTERY, WITH A PROXIMAL REFERENCE VESSEL DIAMETER OF 6 MM, A DISTAL REFERENCE VESSEL DIAMETER OF 5 MM, A LESION LENGTH OF 240 MM, AND 90% STENOSIS. THE LESION WAS CLASSIFIED AS A TASC II B LESION. PRIOR TO TREATMENT OF THE TARGET LESION, BALLOON DILATION WAS PERFORMED USING 4 MM X 100 MM AND 4 MM X 80 MM STERLING OTW PTA BALLOONS. TREATMENT CONSISTED OF DILATION WITH A 5 MM X 40 MM RANGER DRUG-COATED BALLOON STUDY DEVICE, FOLLOWED BY PLACEMENT OF 6 MM X 120 MM AND 6 MM X 80 MM ELUVIA DRUG-ELUTING STENT STUDY DEVICES. POST-TREATMENT DILATION WAS PERFORMED USING A 6 MM X 100 MM MUSTANG OTW PTA BALLOON, AND THE FINAL RESIDUAL STENOSIS WAS DOCUMENTED AS 0%. ON (B)(6) 2021, THE SUBJECT WAS DISCHARGED FROM THE HOSPITAL ON DUAL ANTIPLATELET THERAPY. PER BASELINE CORE LAB FORM DATED (B)(6) 2021, DURING THE INDEX PROCEDURE FOR TREATMENT OF TARGET LESION #001, POST-STUDY DE

    Patient problems: Fistula; Perforation of Vessels

    MAUDE report 23839120 ↗
  • MalfunctionJun 21, 2023Material Rupture; Material Rupture

    IT WAS REPORTED THAT THE BALLOON BURST. A 5.0 X 40MM, 135CM RANGER PACLITAXEL-COATED PTA BALLOON CATHETER WAS SELECTED FOR USE IN AN ANGIOPLASTY PROCEDURE IN THE RIGHT PROFUNDA. THE 60% TARGET LESION WAS MILDLY CALCIFIED AND LOCATED IN NON-TORTUOUS ANATOMY. AFTER ATHERECTOMY WAS PERFORMED, THE RANGER BALLOON WAS PUT IN PLACE. UPON INFLATION, THE BALLOON BURST AT 4ATM IN LESS THAN 5 SECONDS. THE BALLOON WAS TAKEN OUT AND INSPECTED. NO MISSING PIECES FROM THE BALLOON, AND THE SHAFT WAS STILL INTACT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A 6.0 X 40MM, 135CM RANGER PACLITAXEL-COATED PTA BALLOON CATHETER. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17176043 ↗
  • InjuryNov 18, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ELEGANCE STUDY. IT WAS REPORTED THAT THE PATIENT EXPERIENCED TARGET-LESION REVASCULARIZATION. THE SUBJECT UNDERWENT TREATMENT WITH THREE RANGER DRUG COATED BALLOONS ON (B)(6) 2021 AS A PART OF THE ELEGANCE CLINICAL TRIAL. THE TARGET LESION WAS IN THE LEFT PROXIMAL COMMON FEMORAL ARTERY AND FEMORAL-ANTERIOR TIBIAL BYPASS WITH PROXIMAL REFERENCE VESSEL DIAMETER OF 10 MM AND DISTAL REFERENCE VESSEL DIAMETER OF 5 MM. THE LESION WAS CLASSIFIED AS TASC II D, WITH A LENGTH OF 360 MM AND 90% STENOSIS. PRIOR TO TARGET LESION TREATMENT, BALLOON DILATATION WAS PERFORMED USING 3.0 MM X 60 MM STERLING PTA BALLOON. THE TREATMENT OF TARGET LESION WAS PERFORMED BY DILATATION USING STUDY DEVICES 5 MM X 40 MM, 6 MM X 100 MM AND 6 MM X 200 MM RANGER DRUG COATED BALLOONS. FOLLOWING POST DILATATION WAS PERFORMED WITH 7.0 MM X 220 MM STERLING PTA BALLOON AND THE FINAL RESIDUAL STENOSIS WAS NOTED TO BE 10%. ON (B)(6) 2022, 315 DAYS POST-INDEX PROCEDURE THE SUBJECT DEVELOPED WORSENING PERIPHERAL ARTERY DISEASE IN LEFT COMMON FEMORAL ARTERY. IN RESPONSE, THE LEFT COMMON FEMORAL ARTERY TO ANTERIOR TIBIAL ARTERY WAS BYPASSED WITH 6 MM RINGED PROPATEN PTFE. FOLLOWING THAT, A LEFT LOWER EXTREMITY ANGIOGRAM WAS

    Patient problems: Embolism/Embolus; Obstruction/Occlusion; Restenosis; Swelling/ Edema; Thrombosis/Thrombus

    MAUDE report 15822997 ↗
  • InjurySep 29, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ELEGANCE STUDY. IT WAS REPORTED THAT THE PATIENT EXPERIENCED CLAUDICATION RELATED TO WORSENING ISCHEMIA OF THE TARGET VESSEL. THE SUBJECT UNDERWENT TREATMENT WITH THREE RANGER DRUG COATED BALLOONS ON (B)(6) 2021 AS A PART OF THE ELEGANCE CLINICAL TRIAL. THE TARGET LESION WAS IN THE LEFT PROXIMAL COMMON FEMORAL ARTERY AND FEMORAL-ANTERIOR TIBIAL BYPASS WITH PROXIMAL REFERENCE VESSEL DIAMETER OF 10 MM AND DISTAL REFERENCE VESSEL DIAMETER OF 5 MM. THE LESION WAS CLASSIFIED AS TASC II D, WITH A LENGTH OF 360 MM AND 90% STENOSIS. PRIOR TO TARGET LESION TREATMENT, BALLOON DILATATION WAS PERFORMED USING 3.0 MM X 40 MM STERLING PTA BALLOON. THE TREATMENT OF TARGET LESION WAS PERFORMED BY DILATATION USING STUDY DEVICES 5 MM X 40 MM, 6 MM X 100 MM AND 6 MM X 200 MM RANGER DRUG COATED BALLOONS. FOLLOWING POST DILATATION WAS PERFORMED WITH 7.0 MM X 220 MM STERLING PTA BALLOON AND THE FINAL RESIDUAL STENOSIS WAS NOTED TO BE 10%. ON (B)(6) 2022, SUBJECT WAS NOTED WITH THE CLAUDICATION SYMPTOMS RELATED TO WORSENING ISCHEMIA NOTED IN BILATERAL LOWER EXTREMITY. ON THE SAME DAY, SUBJECT UNDERWENT LEFT LOWER EXTREMITY DUPLEX WHICH REVEALED THAT THE LEFT EXTERNAL ILIAC ARTERY WAS PATENT, COMMON FEMORA

    Patient problems: Arteriosclerosis/ Atherosclerosis; Ischemia; Restenosis; Thrombosis/Thrombus

    MAUDE report 15506376 ↗

Other Drug-Eluting Peripheral Transluminal Angioplasty Catheter devices

All Drug-Eluting Peripheral Transluminal Angioplasty Catheter devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page