TRUSELECT™ M001394101750Recalls & reported problems
Boston Scientific Corporation · Microcatheter · Class 2
All models of TRUSELECT (7)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08714729976639), received between Sep 7, 2021 and Nov 4, 2024 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Difficult to Remove” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult to Remove2
- Failure to Advance1
- Fracture1
- Material Deformation1
- Material Integrity Problem1
- Material Perforation1
- Stretched1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 4, 2024Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT THERE WAS A HOLE PRESENT IN THE MICROCATHETER. A TRUSELECT WAS SELECTED FOR USE. DURING THE PROCEDURE, IT WAS OBSERVED BY THE PHYSICIAN THAT THERE WAS A HOLE IN THE CATHETER LUMEN. NO FURTHER INFORMATION WAS PROVIDED REGARDING THE PROCEDURE OR PATIENT STATUS, DESPITE REQUEST BY BOSTON SCIENTIFIC.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20600732 ↗MalfunctionMar 6, 2024Fracture; Difficult to Remove; Failure to Advance; Fracture; Difficult to Remove; Failure to Advance
IT WAS REPORTED THAT FRACTURES OCCURRED. A 175CM TRUSELECT AND A 014X200CM FATHOM GUIDEWIRE WERE SELECTED FOR USE IN A Y90 PROCEDURE TO TREAT HEPATOCELLULAR CARCINOMA (HCC). THE CATHETER WAS ADVANCED WITH THE GUIDEWIRE THROUGH THE TORTUOUS ANATOMY. THE GUIDEWIRE WAS REMOVED TO TAKE AN ANGIOGRAM. THE GUIDEWIRE WAS ADVANCED AGAIN IN ATTEMPT TO ADVANCE THE CATHETER A LITTLE FURTHER IN THE VESSEL WHERE TREATMENT WAS PLANNED; HOWEVER, ADVANCEMENT DIFFICULTY OCCURRED DUE TO THE "CURLY Q" IN THE VESSEL. AFTER MULTIPLE UNSUCCESSFUL ATTEMPTS, THE FATHOM AND TRUSELECT WERE REMOVED WITH RESISTANCE. MULTIPLE FRACTURES WERE NOTED ON THE TRUSELECT. A NEW TRUSELECT AND FATHOM WERE USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18852310 ↗MalfunctionNov 30, 2021Material Perforation; Material Deformation; Material Perforation; Material Deformation
IT WAS REPORTED THE GUIDEWIRE PUNCTURED THROUGH THE CATHETER AND THE PROCEDURE WAS CANCELLED. A 175CM TRUSELECT CATHETER WAS SELECTED FOR A PROSTATE ARTERY EMBOLIZATION PROCEDURE. DURING THE PROCEDURE IT WAS NOTED THAT THE CATHETER WAS KINKED AND THE WIRE WAS GOING THROUGH THE SIDEWALL OF THE CATHETER AT THE KINK. THE PHYSICIAN ATTEMPTED TO PULL THE CATHETER BACK AND THEN REMOVED THE CATHETER AND NOTED THE KINK. THEY DID NOT HAVE ANOTHER CATHETER TO COMPLETE THE PROCEDURE, SO THE CASE WAS CANCELLED AND RESCHEDULED. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12902605 ↗MalfunctionSep 7, 2021Difficult to Remove; Stretched; Difficult to Remove; Stretched
IT WAS REPORTED THAT THE DEVICE WAS DIFFICULT TO REMOVE. A TRUSELECT 175CM STRAIGHT MICROCATHETER WAS SELECTED FOR USE IN A PROSTATE ARTERY EMBOLIZATION PROCEDURE. UPON REMOVAL FROM THE PATIENT, IT WAS REPORTED THAT THERE WAS DIFFICULTY REMOVING THE DEVICE, CAUSING THE DEVICE TO STRETCH NEAR THE HUB AND BECOME UNUSABLE. A NEW MICROCATHETER WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12434653 ↗
FDA clearance and classification
The FDA regulates product code KRA (catheter, continuous flush) under 21 CFR 870.1210 as a Class II device. 21 CFR 870.1210 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the TRUSELECT™ M001394101750
Has the TRUSELECT™ M001394101750 been recalled?
No FDA recall record in our data names this exact device identifier (DI 08714729976639), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the TRUSELECT™ M001394101750?
The FDA's MAUDE database lists 4 adverse event reports for this device, received between Sep 7, 2021 and Nov 4, 2024. In those reports, the most common specific device problems are “Difficult to Remove” (2), “Failure to Advance” (1) and “Fracture” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the TRUSELECT™ M001394101750?
None of the reports are classified as deaths, 0 as injuries and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the TRUSELECT™ M001394101750, and is it still sold?
The FDA's device database lists Boston Scientific Corporation as the company responsible for it (brand TRUSELECT™, model M001394101750). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Catheter, continuous flush devices
- PROWLER SELECT 606-S255XMedos International Sàrl · Catheter, continuous flush2 deaths973 reports
- PhenomTM 27 Catheter FG15150-0615-1SMicro Therapeutics, Inc. · Catheter, continuous flush2 deaths139 reports
- MarathonTM 105-5056Micro Therapeutics, Inc. · Catheter, continuous flush124 reports
- Renegade™ HI-FLO™ Kit M001183020Boston Scientific Corporation · Catheter, continuous flush109 reports
- MarksmanTM FA-55150-1030Micro Therapeutics, Inc. · Catheter, continuous flush109 reports
- EkoSonic® 6779-002Boston Scientific Corporation · Catheter, Continuous Flush1 recall4 deaths55 reports
- Chameleon CH08-40-75USMozarc Medical US LLC · Catheter, continuous flush2 recalls4 reports
- RebarTM 105-5081-153Micro Therapeutics, Inc. · Catheter, continuous flush2 deaths102 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page