Skip to content

TRUSELECT™ M001394111550Recalls & reported problems

Boston Scientific Corporation · Microcatheter · Class 2

All models of TRUSELECT (7)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08714729976660), received between Nov 5, 2020 and Apr 27, 2026 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component2
  • Adverse Event Without Identified Device or Use Problem1
  • Deformation Due to Compressive Stress1
  • Entrapment of Device1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 27, 2026Deformation Due to Compressive Stress; Adverse Event Without Identified Device or Use Problem

    REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON (B)(6) 2026. THE RETURNED DEVICE EVALUATION REVEALED THE SHAFT WAS CRUSHED APPROXIMATELY 7.5CM FROM THE DISTAL TIP. A 155CM BERN TRUSELECT CATHETER WAS USED FOR A PROSTATE ARTERY EMBOLIZATION. ACCESSING A SMALL PENILE BRANCH TO COIL TOOK OVER AN HOUR, BUT ONE PUSHABLE, AND TWO EMBOLD COILS WERE SUCCESSFULLY DEPLOYED. THE MICROCATHETER PROLAPSED PROXIMALLY AND WAS REMOVED TO REGAIN ACCESS INTO THE PROSTATIC ARTERY. THEN IT WAS NOTED THAT A KINK WAS OBSERVED; THEREFORE, THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE AND THERE WERE NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25009586 ↗
  • MalfunctionJul 23, 2025Entrapment of Device; Entrapment of Device

    IT WAS REPORTED THAT THE DEVICES WERE ENTRAPPED. A TRUSELECT MICROCATHETER AND A FATHOM 14 GUIDEWIRE WERE SELECTED FOR USE. DURING THE PROCEDURE, THE GUIDEWIRE WOULD NOT TRACK THROUGH THE MICROCATHETER EASILY, AND THERE WAS DIFFICULTY REMOVING THE GUIDEWIRE FROM THE MICROCATHETER. BOTH DEVICES HAD TO BE REMOVED TOGETHER FROM THE PATIENT. THE PROCEDURE WAS COMPLETED WITH A NEW TRUSELECT AND A NEW FATHOM. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22579796 ↗
  • MalfunctionNov 22, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT TIP DETACHMENT OCCURRED. A 155CM TRUSELECT WAS SELECTED FOR USE. DURING PREPARATION, WHILE TRYING TO PULL THE STYLET OUT OF THE MICROCATHETER, THE TIP OF THE CATHETER CAME OFF WITH IT. THE PROCEDURE WAS COMPLETED WITH A NEW MICROCATHETER. THERE WERE NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18191586 ↗
  • MalfunctionAug 3, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT A TIP BREAK OCCURRED. A 155CM TRUSELECT WAS SELECTED FOR USE. THE HOUSING HOOP WAS FLUSHED AND DURING REMOVAL OF THE CATHETER FROM THE HOOP, THE CATHETER TIP BROKE OFF. THE DEVICE NEVER ENTERED THE PATIENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17454680 ↗
  • MalfunctionNov 5, 2020Leak/Splash; Leak/Splash

    REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 16OCT2020. IT WAS REPORTED THAT A CATHETER LEAK OCCURRED. THE TARGET LESION WAS LOCATED IN SEGMENT 4 OF THE LIVER. A 155CM TRUSELECT WAS SELECTED FOR A PRE Y90 MAPPING OF COLORECTAL METASTASES. DURING LASER MAGNETIC RESONANCE (LMR), PRIOR TO ADMINISTERING THE MELANOMA-ASSOCIATED ANTIGEN (MAA), A COUPLE OF POWER INJECTIONS AT 700 PSI WERE PERFORMED. AFTER THE INJECTIONS, SALINE WAS PUSHED THROUGH THE CATHETER. A POOLING OF SALINE WAS NOTED AT APPROXIMATELY 3CM FROM THE HUB ALONG THE VESTIMID COATING OUTSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH A 2.4 MICROCATHETER. THERE WERE NO PATIENT COMPLICATIONS. HOWEVER, DEVICE ANALYSIS REVEALED A SLICE IN THE CATHETER SHAFT 23CM FROM THE HUB.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10791373 ↗

FDA clearance and classification

The FDA regulates product code KRA (catheter, continuous flush) under 21 CFR 870.1210 as a Class II device. 21 CFR 870.1210 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the TRUSELECT™ M001394111550

Has the TRUSELECT™ M001394111550 been recalled?

No FDA recall record in our data names this exact device identifier (DI 08714729976660), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the TRUSELECT™ M001394111550?

The FDA's MAUDE database lists 5 adverse event reports for this device, received between Nov 5, 2020 and Apr 27, 2026. In those reports, the most common specific device problems are “Detachment of Device or Device Component” (2), “Deformation Due to Compressive Stress” (1) and “Entrapment of Device” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the TRUSELECT™ M001394111550?

None of the reports are classified as deaths, 0 as injuries and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the TRUSELECT™ M001394111550, and is it still sold?

The FDA's device database lists Boston Scientific Corporation as the company responsible for it (brand TRUSELECT™, model M001394111550). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Catheter, continuous flush devices

All Catheter, continuous flush devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page