Flexiva Pulse ID M006L8406940Recalls & reported problems
Boston Scientific Corporation · Flexiva Pulse and Flexiva Pulse TracTip laser fibers are high power single-use fiber · Class 2
All models of Flexiva Pulse ID (5)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 08714729978701), received between Jun 15, 2023 and Jun 1, 2026 — classified as 15 malfunctions.
The most frequently reported specific device problem is “Break” (12 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break12
- Connection Problem1
- Material Deformation1
- Power Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions15
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 1, 2026
FROM OPERATIVE REPORT: FIRST ATTEMPTED TO PERFORM LITHOTRIPSY USING THE TRILOGY DEVICE AT THE HIGHEST SETTING BUT THIS WAS MAKING VERY LITTLE IMPACT ON THE STONE DUE TO ITS VERY HIGH DENSITY. THE DECISION WAS THEN MADE TO USE A 1000 MICRON HOLMIUM LASER FIBER FOR LITHOTRIPSY INSTEAD. LITHOTRIPSY WAS THEN BEGUN USING A POWER OF 100 W. HOWEVER, AFTER ABOUT 15 MINUTES OF LITHOTRIPSY THE FIBER BROKE DUE TO THE HIGH POWER. A NEW FIBER WAS THEN PLACED AND USING A SLIGHTLY LOWER POWER OF 75 W LITHOTRIPSY WAS CONTINUED. AFTER LITHOTRIPSY WAS COMPLETED ON ONE OF THE LARGE STONES, OPERATIVE TIME WAS APPROACHING 3 HOURS AND THE DECISION WAS MADE TO STOP HERE FOR THE DAY. THE TRILOGY WAS THEN REINSERTED AND USED TO CLEAR THE REMAINING DUST AND FRAGMENTS. THE NEPHROSCOPE WAS THEN REMOVED. A 22 FRENCH COUNCIL TIP FOLEY CATHETER WAS THEN INSERTED THROUGH THE ACCESS SHEATH INTO THE POUCH AND THE BALLOON INFLATED WITH 10 ML OF STERILE WATER. THE ACCESS SHEATH WAS THEN CUT SO IT COULD BE REMOVED OVER THE CATHETER. CONTRAST AND SALINE WERE THEN INJECTED THROUGH THE CATHETER INTO THE POUCH TO CONFIRM PLACEMENT. ALL WIRES WERE THEN REMOVED. THE SUPRAPUBIC TUBE WAS SECURED USING A 2-0 NYLON DRAIN STITCH
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25347558 ↗MalfunctionDec 23, 2025Break; Break
IT WAS REPORTED THAT DURING PROCEDURE, THE TIP OF THE LASER FIBER BROKE AT THE EXPOSED GLASS TIP. THE SAME EVENT OCCURRED WITH TWO FIBERS. THE PROCEDURE WAS COMPLETED USING ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS. THIS EVENT IS FOR THE SECOND FIBER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23882797 ↗MalfunctionDec 23, 2025Break; Break
IT WAS REPORTED THAT DURING PROCEDURE, THE TIP OF THE LASER FIBER BROKE AT THE EXPOSED GLASS TIP. THE SAME EVENT OCCURRED WITH TWO FIBERS. THE PROCEDURE WAS COMPLETED USING ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS. THIS EVENT IS FOR THE FIRST FIBER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23882583 ↗MalfunctionJan 8, 2025Break; Break
IT WAS REPORTED THAT DURING THE PROCEDURE, THE FIBER TIP BROKE OFF IN THE MIDDLE OF THE CASE. A SECOND FIBER WAS USED BUT THE SAME PROBLEM WAS ENCOUNTERED. THE PROCEDURE WAS COMPLETED WITH A THIRD FIBER DEVICE. THERE WERE NO PATIENT COMPLICATIONS. THIS EVENT IS FOR THE SECOND FIBER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21106907 ↗MalfunctionJan 8, 2025Break; Break
IT WAS REPORTED THAT DURING THE PROCEDURE, THE FIBER TIP BROKE OFF IN THE MIDDLE OF THE CASE. A SECOND FIBER WAS USED BUT THE SAME PROBLEM WAS ENCOUNTERED. THE PROCEDURE WAS COMPLETED WITH A THIRD FIBER DEVICE. THERE WERE NO PATIENT COMPLICATIONS. THIS EVENT IS FOR THE FIRST FIBER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21106901 ↗MalfunctionDec 30, 2024Break; Break
IT WAS REPORTED THAT DURING A URETEROSCOPY PROCEDURE TO TREAT A KIDNEY STONE, THE FIBER BROKE. THE PROCEDURE WAS COMPLETED WITH ANOTHER FIBER. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21043472 ↗MalfunctionAug 14, 2024Break; Break
IT WAS REPORTED THAT DURING THE PROCEDURE, THE FIBER TIP WAS BROKEN AFTER A SHORT AMOUNT OF TIME LASING. THE PROCEDURE WAS COMPLETED WITH ANOTHER FIBER AND NO PATIENT COMPLICATIONS REPORTED. THIS IS THE SECOND THE TWO COMPLAINTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19988661 ↗MalfunctionAug 5, 2024Break; Power Problem; Break; Power Problem
IT WAS REPORTED THAT DURING A CYSTOLITHOPAXY PROCEDURE, THE GLASS TIP OF THE FIBER BROKE AND STOPPED WORKING. THE FIBER TIP WAS FLUSHED OUT FROM THE PATIENT BY USING AN ELLICK DEVICE. THE PROCEDURE WAS COMPLETED WITH ANOTHER FIBER. THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19913179 ↗MalfunctionAug 5, 2024Break; Break
IT WAS REPORTED THAT DURING THE PROCEDURE, THE FIBER TIP WAS BROKEN AFTER A SHORT AMOUNT OF TIME LASING. THE PROCEDURE WAS COMPLETED WITH ANOTHER FIBER AND NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19908525 ↗MalfunctionJul 23, 2024Break; Break
IT WAS REPORTED THAT THE LASER WIRE CAME APART FROM THE FIBER CONNECTOR. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE AND NO PATIENT COMPLICATIONS WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19811234 ↗MalfunctionJun 10, 2024Connection Problem; Connection Problem
IT WAS REPORTED THAT THE FIBER WAS NOT RECOGNIZED BY THE LASER. THE PROCEDURE WAS NOT COMPLETED DUE TO THIS EVENT. NO PATIENT COMPLICATIONS WERE REPORTED. THIS EVENT IS BEING REPORTED FOR ABORTED/CANCELLED PROCEDURE WITH A PATIENT UNDER ANESTHESIA OR SEDATION IS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19506021 ↗MalfunctionMay 22, 2024Break; Break
IT WAS REPORTED THAT DURING THE PROCEDURE, THE TIP OF THE FIBER WAS BROKEN. THE PROCEDURE WAS COMPLETED WITH ANOTHER FIBER WITH NO PATIENT COMPLICATOINS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19373022 ↗MalfunctionMay 22, 2024Break; Break
IT WAS REPORTED THAT DURING A PROCEDURE, THE FIBER TIP BROKE. A NEW FIBER WAS USED TO COMPLETE THE PROCEDURE WITH NO PATIENT COMPLICATIONS. THIS REPORT IS FOR THE FIRST FIBER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19373015 ↗MalfunctionJun 20, 2023Break; Break
IT WAS REPORTED THAT THE LASER FIBER TIP DEGRADED AFTER LESS THAN 1 MINUTE PEDAL TIME, AND THE TIP WAS ALSO BROKEN. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17167779 ↗MalfunctionJun 15, 2023Material Deformation; Material Deformation
IT WAS REPORTED THAT DURING A LITHOPAXEY PROCEDURE FOR BLADDER STONE, AFTER 10 MINUTES OF LASING, THE FIBER TIP BURNED BACK TO THE CLADDING. IT WAS NEEDED TO USE ANOTHER FIBER TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17139749 ↗
Other Powered laser surgical instrument devices
- Moxy 0010-2400Boston Scientific Corporation · Powered Laser Surgical Instrument4,572 reports
- SOLTIVE TFL-PLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls450 reports
- Flexiva TracTip M0068403960Boston Scientific Corporation · Powered laser surgical instrument562 reports
- Moses™ AC-10030100Lumenis Ltd. · Powered Laser Surgical Instrument532 reports
- Flexiva M0068403910Boston Scientific Corporation · Powered laser surgical instrument376 reports
- Visualase 9735542Medtronic Navigation, Inc. · Powered laser surgical instrument1 recall2 deaths311 reports
- SOLTIVE TFL-SLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls33 reports
- NeuroBlate® System 20980Monteris Medical Inc · Powered laser surgical instrument7 deaths296 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.