GreenLight XPS™ 0010-0210S-GZ0Recalls & reported problems
Boston Scientific Corporation · Laser System · Class 2
All models of GreenLight XPS (2)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 08714729979173), received between Sep 5, 2023 and Apr 23, 2026 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Device Displays Incorrect Message” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Displays Incorrect Message2
- Power Problem2
- Failure to Power Up1
- No Display/Image1
- Nonstandard Device1
- Physical Resistance/Sticking1
- Unexpected Shutdown1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 23, 2026Nonstandard Device; Failure to Power Up; Power Problem; Unexpected Shutdown
IT WAS REPORTED THAT DURING PROCEDURE, THE CONSOLE DISPLAYED ERROR CODE 213 (LASER POWER SUPPLY OFF SETPOINT) AFTER 30 MINUTES OF USE. WHEN "CLEAR PROBLEM" WAS PRESSED, THE POWER SHUT OFF AND THE CONSOLE WOULD NOT START UP AFTERWARD. RESTARTING FROM THE BREAKER DID NOT CHANGE THE SITUATION. THERE WAS A SLIGHT UNUSUAL ODOR WAS ALSO NOTED. THE PROCEDURE WAS COMPLETE USING A NON BSC PRODUCT. THERE WERE NO PATIENT COMPLICATIONS. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED AND SERVICED THE CONSOLE AT THE FACILITY. DURING EVALUATION OF THE CONSOLE, A BURNING SMELL WAS DETECTED; THEREFORE, REPORTABILITY CRITERIA IS MET BASED ON THIS FINDING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24974810 ↗MalfunctionMar 16, 2026Power Problem
IT WAS REPORTED THAT DURING A PROCEDURE, THE CONSOLE DISPLAYED ERRORS 211 (LASER POWER SUPPLY BRICK FAULT) AND 213 (LPS OFF SETPOINT). THE CONSOLE POWERED OFF AND DID NOT TURN POWER ON AGAIN. DUE TO THIS, THE PROCEDURE COULD NOT BE COMPLETED WHEN THE PATIENT WAS UNDER GENERAL SEDATION. THERE WERE NO PATIENT COMPLICATIONS. THIS EVENT IS BEING REPORTED FOR AN ABORTED/CANCELLED PROCEDURE WITH A PATIENT UNDER ANESTHESIA OR SEDATION STATUS IS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24614113 ↗MalfunctionJun 23, 2025Physical Resistance/Sticking; Physical Resistance/Sticking
IT WAS REPORTED THAT DURING PREPARATION PRIOR TO PROCEDURE, THE COAGULATION PEDAL REMAINED ACTIVATED, IT APPEARS THAT THE CONTACT IS DAMAGED. THE PROCEDURE WAS COMPLETED USING THE SAME GENERATOR WITH A DIFFERENT FOOT SWITCH WITH NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22308263 ↗MalfunctionMay 15, 2025Device Displays Incorrect Message; Device Displays Incorrect Message
IT WAS REPORTED THAT THE CONSOLE DISPLAYED ERROR CODE 660 DURING PROCEDURE. THE CONSOLE WAS TURNED OFF FOR SAFETY AND THEN ERROR CODED 611 AND 831 WERE DISPLAYED, THEREFORE, THE DAY WAS SUSPENDED. THERE WAS NO PROCEDURE AND PATIENT OUTCOME REPORTED. A FULL SERVICE WAS PERFORMED ON THE CONSOLE; HOWEVER NO ISSUES WERE OBSERVED DURING THE EVALUATION. ALSO, WAS INFORMED THAT THE CITY WHERE THE EQUIPMENT IS LOCATED HAS BEEN EXPERIENCING ISSUES WITH THE ELECTRICAL GRID, AND THAT ELECTRICAL IRREGULARITIES MAY BE CONTRIBUTING TO THE REPORTED ERRORS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22032100 ↗MalfunctionMar 26, 2024No Display/Image; No Display/Image
IT WAS REPORTED THAT, DURING PREPARATION FOR THE PHOTOSELECTIVE VAPORIZATION OF THE PROSTATE PROCEDURE, THE SCREEN DID NOT DISPLAY ANY IMAGES. THE PROCEDURE WAS THEN CANCELLED. THERE WERE NO PATIENT COMPLICATIONS. THIS EVENT IS BEING REPORTED FOR AN ABORTED/CANCELLED PROCEDURE WITH A PATIENT UNDER GENERAL ANESTHESIA OR WHOSE SEDATION STATUS WAS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18981212 ↗MalfunctionSep 5, 2023Device Displays Incorrect Message; Device Displays Incorrect Message
IT WAS REPORTED THAT DURING A PROSTATE ABLATION PROCEDURE, THE CONSOLE LAUNCHES AN EMERGENCY SHUTDOWN MESSAGE, AND THE PROCEDURE COULD NOT BE COMPLETED. THERE WERE NO PATIENT COMPLICATIONS. THIS EVENT IS BEING REPORTED FOR AN ABORTED/CANCELLED PROCEDURE WITH A PATIENT UNDER ANESTHESIA OR SEDATION STATUS IS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17687544 ↗
Other Powered laser surgical instrument devices
- Moxy 0010-2400Boston Scientific Corporation · Powered Laser Surgical Instrument4,572 reports
- SOLTIVE TFL-PLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls450 reports
- Flexiva TracTip M0068403960Boston Scientific Corporation · Powered laser surgical instrument562 reports
- Moses™ AC-10030100Lumenis Ltd. · Powered Laser Surgical Instrument532 reports
- Flexiva M0068403910Boston Scientific Corporation · Powered laser surgical instrument376 reports
- Visualase 9735542Medtronic Navigation, Inc. · Powered laser surgical instrument1 recall2 deaths311 reports
- SOLTIVE TFL-SLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls33 reports
- NeuroBlate® System 20980Monteris Medical Inc · Powered laser surgical instrument7 deaths296 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.