SYNERGY™ XD H7493941808350Recalls & reported problems
Boston Scientific Corporation · Everolimus-Eluting Platinum Chromium Coronary Stent System · Class 3 · Implant
All models of SYNERGY XD (65)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 08714729981022), received between May 12, 2021 and May 27, 2026 — classified as 1 death, 5 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Break2
- Device Contamination with Chemical or Other Material1
- Device Markings/Labelling Problem1
- Difficult to Advance1
- Difficult to Remove1
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Restenosis1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 27, 2026Fracture; Difficult to Remove; Difficult to Advance
IT WAS REPORTED THAT REMOVAL DIFFICULTY OCCURRED. THE PHYSICIAN INITIALLY PERFORMED BALLOON ANGIOPLASTY ON THE LESION AND ATTEMPTED TO DEPLOY A 3.50 X 8MM SYNERGY XD DRUG-ELUTING STENT; HOWEVER, DELIVERY WAS DIFFICULT DUE TO INSUFFICIENT LESION PREPARATION. THE STENT WAS WITHDRAWN WITH THE INTENT TO FURTHER DILATE THE LESION. DURING REMOVAL, THE STENT BECAME LODGED IN THE GUIDE CATHETER, REQUIRING THE PHYSICIAN TO REMOVE THE ENTIRE GUIDE CATHETER FROM THE PATIENT TO RETRIEVE IT. UPON INSPECTION AFTER REMOVAL, ALL COMPONENTS OF THE STENT WERE THERE BUT IT WAS FRACTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AND THE PATIENT FULLY RECOVERED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25308408 ↗DeathMar 8, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT DIED. THE STENOSED TARGET LESION WAS LOCATED IN THE LEFT MAIN ARTERY. A NON-BOSTON SCIENTIFIC HEART PUMP WAS PLACED, AND AN UNKNOWN SYNERGY XD DRUG-ELUTING STENT WAS IMPLANTED. THE HEART PUMP WAS REMOVED THE NEXT DAY. THE PATIENT EXPIRED FIVE DAYS LATER; THE PHYSICIAN SUSPECTED STENT CLOTTING OFF AS A POSSIBLE CAUSE. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Restenosis
MAUDE report 18871126 ↗MalfunctionFeb 22, 2023Break; Break
IT WAS REPORTED THAT SHAFT BREAK OCCURRED. THE PATIENT PRESENTED WITH CORONARY ARTERY DISEASE. A 3.50 X 8MM SYNERGY XD DRUG-ELUTING STENT WAS SELECTED FOR USE. HOWEVER, WHILE LOADING THE STENT ONTO A WIGGLE WIRE, APPROXIMATELY 1/4 FROM DISTAL TIP OF DELIVERY DEVICE WAS SEPARATED AND WAS NEVER INTRODUCED INTO THE GUIDE CATHETER. THE BREAK OCCURRED TOWARD DISTAL PORTION APPROXIMATELY AT THE WIRE EXIT PORT. THE PROXIMAL PORTION WAS REMOVED, AND THE DISTAL PORTION WAS PULLED FROM WIRE. ANOTHER SYNERGY STENT WAS USED AND COMPLETED THE PROCEDURE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16418426 ↗MalfunctionNov 9, 2022Device Contamination with Chemical or Other Material; Adverse Event Without Identified Device or Use Problem; Device Contamination with Chemical or Other Material; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DEVICE CONTAMINATION OCCURRED. DURING UNPACKING OF A 3.50 X 8MM SYNERGY XD DRUG-ELUTING STENT, IT WAS NOTED THAT A WHITE PARTICLE WAS PRESENT ON THE TIP OF THE STENT. THE PROCEDURE WAS COMPLETED VIA ALTERNATIVE METHOD. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15760002 ↗MalfunctionNov 15, 2021Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT IMPROPER LABELING OCCURRED. A 3.50 X 8MM SYNERGY XD DRUG ELUTING STENT WAS SELECTED FOR USE. HOWEVER, IT WAS NOTED THAT THE SIZE OF THE STENT WITHIN THE BOX WAS INCORRECT. THE BOX WAS LABELED FOR 3.50 X 8MM SYNERGY XD; HOWEVER, THE STENT POUCH WAS 3.5 X 20 XD. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE USED. THERE WERE NO COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12811269 ↗MalfunctionMay 12, 2021Break; Break
IT WAS REPORTED THAT SHAFT BREAK OCCURRED. A 3.50 X 8MM SYNERGY XD DRUG-ELUTING STENT WAS SELECTED TO USE. HOWEVER, DURING INSERTION, THE PHYSICIAN BROKE THE SHAFT OF THE STENT 6 INCHES FROM THE HUB OUTSIDE THE PATIENT'S BODY. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11815558 ↗
Other Coronary drug-eluting stent devices
- SYNERGY™ H7493926038300Boston Scientific Corporation · Coronary drug-eluting stent8 deaths124 reports
- SYNERGY™ H7493926016300Boston Scientific Corporation · Coronary drug-eluting stent7 deaths106 reports
- SYNERGY™ H7493926020300Boston Scientific Corporation · Coronary drug-eluting stent9 deaths90 reports
- SYNERGY™ H7493926016250Boston Scientific Corporation · Coronary drug-eluting stent6 deaths82 reports
- XIENCE Sierra™ 1550350-33Abbott Vascular Inc. · Coronary drug-eluting stent1 recall32 reports
- Resolute Onyx™ RONYX30022UXMedtronic, Inc. · Coronary drug-eluting stent6 deaths76 reports
- SYNERGY™ H7493926024300Boston Scientific Corporation · Coronary drug-eluting stent3 deaths75 reports
- Resolute Onyx™ RONYX30015UXMedtronic, Inc. · Coronary drug-eluting stent3 deaths73 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page