SYNERGY™ XD H7493941838400Recalls & reported problems
Boston Scientific Corporation · Everolimus-Eluting Platinum Chromium Coronary Stent System · Class 3 · Implant
All models of SYNERGY XD (65)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Boston Scientific Synergy XD everolimus-eluting coronary stent is used to hold open narrowed heart arteries. Most of the 31 reports describe the stent shifting or coming loose from its balloon during delivery, failing to cross tight or calcified lesions, or deforming, cracking or breaking. Five reports describe injuries; no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 31
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 31 adverse event reports that list this exact device (DI 08714729981176), received between Oct 23, 2020 and Apr 20, 2026 — classified as 5 injuries, 26 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (15 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated15
- Material Deformation8
- Failure to Advance6
- Break5
- Difficult to Remove4
- Adverse Event Without Identified Device or Use Problem1
- Deflation Problem1
- Difficult to Advance1
Patient problems
- No Clinical Signs, Symptoms or Conditions26
- Vascular Dissection2
- Dyspnea1
- Heart Failure/Congestive Heart Failure1
- Insufficient Information1
- Myocardial Infarction1
- No Consequences Or Impact To Patient1
- Renal Impairment1
Recent adverse event reports
The 25 most recent of 31 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 20, 2026Failure to Advance; Device Dislodged or Dislocated
IT WAS REPORTED THAT STENT MOVED ON BALLOON OCCURRED. A 4.00 X 38MM SYNERGY XD DRUG-ELUTING STENT WAS ADVANCE FOR TREATMENT, BUT IT COULD NOT CROSS THE LESION. HOWEVER, IT WAS NOTED THAT THE STENT WAS NOT COMPLETELY SEATED ON DEPLOYMENT SYSTEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24931814 ↗MalfunctionNov 17, 2025Device Dislodged or Dislocated; Material Deformation; Device Dislodged or Dislocated; Material Deformation
IT WAS REPORTED THAT STENT MOVED ON THE BALLOON. A 4.00 X 38MM SYNERGY? XD DRUG-ELUTING STENT WAS ADVANCED FOR TREATMENT. HOWEVER, THE STENT CAME OFF THE BALLOON AND WAS DAMAGED. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23572438 ↗MalfunctionNov 3, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT STENT DISLODGEMENT OCCURRED. A 4.00 X 38MM SYNERGY? XD DRUG-ELUTING STENT WAS SELECTED FOR USE. HOWEVER, DURING PREPARATION, THE STENT WAS OBSERVED TO BE DETACHED FROM THE SHAFT AND WAS UNABLE TO BE DEPLOYED. THE PROCEDURE WAS COMPLETED WITH ANOTHER SYNERGY DES. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23457975 ↗MalfunctionSep 15, 2025Leak/Splash; Leak/Splash
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 25-AUG-2025. IT WAS REPORTED THAT BALLOON LEAK WAS ENCOUNTERED. THE 70% STENOSED TARGET LESION WAS LOCATED IN A MILDLY TORTUOUS AND MILDLY CALCIFIED CORONARY ARTERY. A 4.00 X 38MM SYNERGY XD DRUG-ELUTING STENT WAS ADVANCED FOR TREATMENT. HOWEVER, THE BALLOON WAS INFLATED TO 8 ATMOSPHERES BEFORE DEFLATING IMMEDIATELY. THE STENT WAS OPPOSED TO THE VESSEL WALL ENOUGH TO REMOVE THE BALLOON. THE STENT WAS POST DILATED WITH A NON-COMPLIANT (NC) BALLOON AND PATIENT WAS FINE. THE PROCEDURE WAS COMPLETED USING AN ALTERNATE DEVICE. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED BALLOON TORN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23057715 ↗MalfunctionDec 9, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT STENT DISLODGEMENT OCCURRED. THE TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED RIGHT CORONARY ARTERY. A 4.00 X 38MM SYNERGY XD DRUG-ELUTING STENT WAS SELECTED FOR TREATMENT. HOWEVER, DURING INSERTION THROUGH A NON-BOSTON SCIENTIFIC GUIDE EXTENSION CATHETER, THE STENT WAS COMPLETELY DETACHED FROM THE BALLOON INSIDE THE GUIDE. EVERYTHING WAS REMOVED TOGETHER; THE STENT WAS REMOVED INSIDE THE GUIDE. THE PROCEDURE WAS COMPLETED WITH A NEW GUIDEWIRE AND STENT. THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20880564 ↗MalfunctionDec 2, 2024Device Dislodged or Dislocated; Material Deformation; Device Dislodged or Dislocated; Material Deformation
IT WAS REPORTED THAT STENT MOVED ON BALLOON OCCURRED. A 4.00 X 38MM SYNERGY XD DRUG-ELUTING STENT WAS SELECTED FOR TREATMENT. HOWEVER, DURING UNPACKING, THE STENT WAS SMASHED AND UN-CLEANLY ATTACHED TO THE BALLOON. A NEW STENT WAS THEN PULLED WITHOUT INCIDENT, AND THE PROCEDURE WAS COMPLETED WITH A NON-BOSTON SCIENTIFIC DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20819543 ↗MalfunctionAug 27, 2024Material Rupture; Material Rupture
IT WAS REPORTED VIA USER FACILITY MEDWATCH # 5157457 THAT BALLOON RUPTURE OCCURRED. A 4.00 MM X 38 MM SYNERGY XD DRUG-ELUTING STENT WAS INSERTED INTO PROXIMAL MID RIGHT CORONARY ARTERY (RCA). THE STENT WAS POSITIONED. HOWEVER, THE BALLOON BURST UPON INFLATION FOR STENT DEPLOYMENT AT 20 ATMOSPHERES. THE STENT DELIVERY SYSTEM WAS REMOVED WITH THE BALLOON REMAINING INTACT. A CONTRAST INJECTION TO THE RCA, OBSERVED UNDER FLUOROSCOPY, CONFIRMED THAT THE STENT WAS DEPLOYED IN THE APPROPRIATE LOCATION. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20075233 ↗MalfunctionJul 17, 2024Fracture; Fracture
IT WAS REPORTED THAT STENT FRACTURE OCCURRED. THE PATIENT PRESENTED WITH A CHRONIC TOTAL OCCLUSION OF THE LEFT ANTERIOR DESCENDING ARTERY. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE NON-TORTUOUS AND HEAVILY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 4.00 X 38MM SYNERGY XD DRUG-ELUTING STENT FRACTURED IN THE BODY OF THE STENT AFTER DEPLOYMENT. THE STENT REMAINED INTACT AROUND THE FRACTURE, AND THERE WAS NO PERFORATION. POST-DILATION WAS PERFORMED, AND IT WAS NOTICED THAT THE SIDE CELL APPEARED ABNORMAL, AS IF IT HAD "BLOWN OUT" ON THE SIDES OF THE CELL. THERE WAS NOTHING DISLODGED OR DETACHED. THE DEVICE WAS NOT REMOVED, AND THE PROCEDURE WAS COMPLETED. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19767567 ↗MalfunctionMay 1, 2024Break; Break
IT WAS REPORTED THAT SHAFT BREAK OCCURRED. THE TARGET LESION WAS LOCATED IN CORONARY ARTERY. A 4.00 X 38MM SYNERGY XD DRUG-ELUTING STENT WAS SELECTED TO TREAT THE LESION. HOWEVER, DURING INSERTION, THE DEVICE BROKE OFF IN THE GUIDE. ANOTHER STENT WAS OPENED, AND THE PROCEDURE WAS COMPLETED. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19220604 ↗MalfunctionApr 24, 2024Break; Break
IT WAS REPORTED THAT SHAFT BREAK OCCURRED. THE TARGET LESION WAS LOCATED IN CORONARY ARTERY. A 4.00 X 38MM SYNERGY XD DRUG-ELUTING STENT WAS SELECTED TO TREAT THE LESION. HOWEVER, DURING INSERTION, THE DEVICE BROKE OFF IN THE GUIDE. ANOTHER STENT WAS OPENED, AND THE PROCEDURE WAS COMPLETED. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19174938 ↗MalfunctionMar 8, 2024Device Dislodged or Dislocated; Material Deformation; Device Dislodged or Dislocated; Material Deformation
IT WAS REPORTED THAT THE STENT MOVED ON BALLOON. A 4.00 X 38MM SYNERGY DRUG-ELUTING STENT WAS SELECTED FOR TREATMENT. UPON OPENING, IT WAS NOTED THAT THE STENT WAS MANGLED ON THE BALLOON. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18869581 ↗MalfunctionFeb 23, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT STENT MOVED ON BALLOON OCCURRED. THE TARGET LESION WAS LOCATED IN THE CORONARY ARTERY. A 4.00 X 38MM SYNERGY XD DRUG ELUTING STENT (DES) WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE PROCEDURE, THE STENT PARTIALLY SHIFTED ON THE STENT BALLOON PRIOR TO DEPLOYMENT. THE DEVICE WAS REMOVED FROM THE PATIENT'S BODY SUCCESSFULLY, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18770199 ↗MalfunctionJan 23, 2024Break; Difficult to Remove; Difficult to Advance; Material Deformation; Break; Difficult to Remove; Difficult to Advance; Material Deformation
IT WAS REPORTED THAT SHAFT BREAK OCCURRED. THE TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS MID RIGHT CORONARY ARTERY (RCA). TWO STENTS HAD ALREADY BEEN PREVIOUSLY IMPLANTED IN THE PROXIMAL AND MID RCA. A NON-BOSTON SCIENTIFIC GUIDE EXTENSION CATHETER WAS ADVANCED TO THE MID RCA AND A 4.00 X 38MM SYNERGY XD DRUG ELUTING STENT DELIVERY SYSTEM (SDS) WAS INTRODUCED. RESISTANCE WAS ENCOUNTERED WHILE ADVANCING THE SDS AND ONLY 30% OF THE STENT WAS ADVANCED OUT OF THE GUIDE EXTENSION CATHETER. IT WAS NOTED ON ANGIO THAT THE STENT HAD AN ACCORDION APPEARANCE, BUT IT WAS NOT KNOWN IF THE UNDEPLOYED STENT HAD GOTTEN CAUGHT ON A STENT STRUT OR IF IT HAD GOTTEN CAUGHT COMING OUT OF THE GUIDE EXTENSION CATHETER. DURING AN ATTEMPT TO REMOVE THE SYNERGY XD, THE DEVICE WOULD NOT GO BACK INTO THE GUIDE EXTENSION CATHETER AND THE STENT DELIVERY SYSTEM (SDS) BROKE. THE DETACHED PORTION OF THE SDS WAS LODGED WELL ENOUGH INTO THE GUIDE EXTENSION CATHETER THAT ALL PIECES WERE ABLE TO BE REMOVED, ALONG WITH THE WIRE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18564475 ↗MalfunctionOct 24, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT STENT DISLODGEMENT OCCURRED. THE 4.0 X 38 MM SYNERGY XD DRUG-ELUTING STENT WAS SELECTED FOR USE. BEFORE ENTERING THE BODY, THE STENT CAME OFF THE BALLOON. THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18000451 ↗MalfunctionOct 24, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT STENT DISLODGEMENT OCCURRED. THE 4.0 X 38 MM SYNERGY XD DRUG-ELUTING STENT WAS SELECTED FOR USE. BEFORE ENTERING THE BODY, THE STENT CAME OFF THE BALLOON. THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18000450 ↗MalfunctionOct 24, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT STENT DISLODGEMENT OCCURRED. THE 4.0 X 38 MM SYNERGY XD DRUG-ELUTING STENT WAS SELECTED FOR USE. BEFORE ENTERING THE BODY, THE STENT CAME OFF THE BALLOON. THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18000449 ↗InjurySep 14, 2023Obstruction of Flow; Obstruction of Flow
IT WAS REPORTED VIA USER MEDWATCH MW5120195 THAT MULTIPLE PATIENT ISSUES OCCURRED. ON (B)(6) 2023, THE PATIENT PRESENTED TO THE EMERGENCY DEPARTMENT WITH CHEST PRESSURE AND INCREASING SHORTNESS OF BREATH AND WAS TRANSFERRED TO ANOTHER FACILITY FOR CARDIAC CATHETERIZATION. FILMS REVEALED MULTIVESSEL DISEASE AND THE DECISION WAS MADE TO PROCEED WITH PERCUTANEOUS INTERVENTION OF THE LEFT MAIN (LM)/PROXIMAL LEFT ANTERIOR DESCENDING ARTERY (LAD) AND THE LEFT DOMINANT PROXIMAL CIRCUMFLEX (CX) ARTERY. THE 90% STENOSED CX WAS TREATED WITH A 3.50 NON-BOSTON SCIENTIFIC (NON-BSC) NON-COMPLAINT (NC) BALLOON, AND PLACEMENT OF A 4.00 X 38 SYNERGY XD STENT. THE STENT WAS POST-DILATED WITH A 4.50 X 12 NON-BSC NC BALLOON AND FINAL FILMS REVEALED GOOD RESULT WITH TIMI III FLOW. THE PATIENT CONTINUED TO DO WELL AND WAS DISCHARGED ON ASPIRIN AND PLAVIX. THE PATIENT REPORTED BEING COMPLIANT WITH THEIR MEDICATIONS. ON (B)(6) 2023, THE PATIENT RETURNED TO THE EMERGENCY DEPARTMENT WITH INCREASING SHORTNESS OF BREATH AND NO CHEST DISCOMFORT. OVER THE NEXT WEEK THE PATIENT WAS TREATED FOR RESPIRATORY DISTRESS, ACUTE ON CHRONIC CONGESTIVE HEART FAILURE, ACUTE KIDNEY INJURY, AND NON-ST ELEVATION MYOCARDIAL IN
Patient problems: Dyspnea; Heart Failure/Congestive Heart Failure; Myocardial Infarction; Renal Impairment; Respiratory Insufficiency; Thrombosis/Thrombus
MAUDE report 17750255 ↗InjurySep 6, 2023Break; Difficult to Remove; Failure to Advance; Material Deformation; Adverse Event Without Identified Device or Use Problem; Break; Difficult to Remove; Failure to Advance; Material Deformation; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT SHAFT BREAK AND REMOVAL DIFFICULTIES OCCURRED. A PERCUTANEOUS CORONARY INTERVENTION (PCI) PROCEDURE WAS PERFORMED. THE STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED CORONARY ARTERY. A 4.00 X 38MM SYNERGY XD DRUG-ELUTING STENT (DES) WAS ADVANCED FOR TREATMENT. HOWEVER, DURING DELIVERY THROUGH A 6F NON-BOSTON SCIENTIFIC GUIDE EXTENSION CATHETER, THE STENT FAILED TO CROSS THE ENTIRE LESION. THE PHYSICIAN ATTEMPTED TO PULL THE DEVICE OUT, BUT THE SHAFT BROKE WITH THE STENT, BALLOON, AND PART OF THE SHAFT STILL ENTRAPPED IN THE CORONARY ARTERY. MULTIPLE UNSUCCESSFUL SNARING ATTEMPTS WERE MADE, SO THE PATIENT WAS SENT FOR OPEN-HEART SURGERY TO RETRIEVE THE BROKEN DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17692486 ↗MalfunctionFeb 8, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT STENT DISLODGMENT OCCURRED. A 4.00 X 38MM SYNERGY XD DRUG-ELUTING STENT WAS SELECTED FOR USE. HOWEVER, DURING PREPARATION, THE STYLET WAS PULLED AND THE STENT DETACHED FROM THE BALLOON. A NEW STENT WAS USED TO COMPLETE THE PROCEDURE SUCCESSFULLY. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16326464 ↗MalfunctionDec 2, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT STENT DISLODGEMENT OCCURRED. A 4.00 X 38MM SYNERGY XD DRUG-ELUTING STENT WAS SELECTED FOR USED. HOWEVER, UPON PULLING THE STENT PROTECTOR, THE STENT CAME WITH IT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15906432 ↗MalfunctionNov 14, 2022Failure to Advance; Device Dislodged or Dislocated; Failure to Advance; Device Dislodged or Dislocated
IT WAS REPORTED THAT STENT MOVED ON BALLOON OCCURRED. THE TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS MID CIRCUMFLEX ARTERY. A 4.00 X 38MM SYNERGY XD DRUG-ELUTING STENT (DES) WAS ADVANCED TO TREAT THE LESION. HOWEVER, DURING THE PROCEDURE, THE DEVICE FAILED TO CROSS THE LESION, AND THE PHYSICIAN NOTICED THAT THE STENT APPEARED TO BE PARTIALLY DETACHED FROM THE BALLOON. THE STENT WAS CAREFULLY REMOVED BACK INTO THE GUIDE AND WITHDRAWN. THE PROCEDURE WAS COMPLETED WITH ANOTHER SYNERGY XD DES. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15788635 ↗MalfunctionSep 13, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE STENT DISLODGEMENT OCCURRED. A 4.00 X 38MM SYNERGY XD STENT BALLOON WAS SELECTED FOR USE. HOWEVER, DURING THE PREPARATION, WHEN PULLING THE STYLET THE STENT CAME OFF THE BALLOON DELIVERY SYSTEM. THE PROCEDURE WAS COMPLETED WITH A NEW STENT. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15411017 ↗MalfunctionJul 27, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT STENT DISLODGEMENT OCCURRED. A 4.00 X 38MM SYNERGY XD DRUG-ELUTING STENT WAS ADVANCED TO TREAT THE LESION. HOWEVER, DURING THE PROCEDURE, THE STENT CAME OFF THE DELIVERY SYSTEM. THE DEVICE WAS REMOVED FROM THE PATIENT BODY ALONG WITH THE STENT DELIVERY SYSTEM. THE PROCEDURE WAS COMPLETED WITH A NEW STENT. NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15113847 ↗InjuryJun 9, 2022Failure to Advance; Failure to Advance
IT WAS REPORTED THAT AN ADVERSE PATIENT EVENT OCCURRED. A 4.00 X 38MM SYNERGY XD DRUG-ELUTING STENT WAS ADVANCED BUT FAILED TO CROSS THE LESION. THE INABILITY TO CROSS THE LESION CAUSED A SIGNIFICANT DELAY TO THE PROCEDURE THAT RESULTED IN AN ADVERSE PATIENT EVENT. DESPITE MULTIPLE EFFORTS, NO FURTHER INFORMATION WAS ABLE TO BE OBTAINED.
Patient problems: Insufficient Information
MAUDE report 14649308 ↗MalfunctionJan 3, 2022Material Deformation; Material Deformation
IT WAS REPORTED THAT STENT DAMAGE OCCURRED. A 4.00 X 38MM SYNERGY XD DRUG-ELUTING STENT WAS SELECTED FOR USE. WHEN THE DEVICE WAS PULLED FROM THE HOOP AND REMOVED FROM STENT PROTECTOR DURING PREPARATION, THE DISTAL END OF THE STENT WAS LIFTED AND DAMAGED. BUT THE PROXIMAL PORTION WAS INTACT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13147002 ↗
Other Coronary drug-eluting stent devices
- SYNERGY™ H7493926038300Boston Scientific Corporation · Coronary drug-eluting stent8 deaths124 reports
- SYNERGY™ H7493926016300Boston Scientific Corporation · Coronary drug-eluting stent7 deaths106 reports
- SYNERGY™ H7493926020300Boston Scientific Corporation · Coronary drug-eluting stent9 deaths90 reports
- SYNERGY™ H7493926016250Boston Scientific Corporation · Coronary drug-eluting stent6 deaths82 reports
- XIENCE Sierra™ 1550350-33Abbott Vascular Inc. · Coronary drug-eluting stent1 recall32 reports
- Resolute Onyx™ RONYX30022UXMedtronic, Inc. · Coronary drug-eluting stent6 deaths76 reports
- SYNERGY™ H7493926024300Boston Scientific Corporation · Coronary drug-eluting stent3 deaths75 reports
- Resolute Onyx™ RONYX30015UXMedtronic, Inc. · Coronary drug-eluting stent3 deaths73 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page