SYNERGY™ XD H7493941816450Recalls & reported problems
Boston Scientific Corporation · Everolimus-Eluting Platinum Chromium Coronary Stent System · Class 3 · Implant
All models of SYNERGY XD (65)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729981190), received between Apr 27, 2021 and Dec 13, 2023 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Difficult to Advance1
- Difficult to Remove1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 13, 2023Break; Break
IT WAS REPORTED THAT SHAFT BREAK OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED CORONARY ARTERY. A 4.50 X 16MM SYNERGY XD DRUG-ELUTING STENT (DES) WAS ADVANCED FOR TREATMENT. HOWEVER, DURING ADVANCING, THE DELIVERY SYSTEM WAS FRACTURED INTO 2 PIECES AFTER FORCEDLY PUSHING IT INTO THE LESION. THE DEVICE WAS COMPLETELY REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A 4.0X16 SYNERGY DES. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18319434 ↗MalfunctionFeb 4, 2022Difficult to Advance; Material Deformation; Difficult to Advance; Material Deformation
IT WAS REPORTED THAT DIFFICULTY REMOVAL, SHAFT BREAK, AND PROCEDURE RESCHEDULING OCCURRED. THE 80% STENOSED TARGET LESION WAS LOCATED IN THE TORTUOUS AND CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 4.50 X 20MM SYNERGY XD DRUG-ELUTING STENT WAS ADVANCED HOWEVER THEY MET RESISTANCE IN THE VESSEL RIGHT AS THE STENT EXITED THE GUIDE CATHETER. THEY PUSHED AND PULLED AND NOTICE THE HYPOTUBE CAME OUT WITHOUT THE DISTAL PART OF THE DELIVERY SYSTEM. THEY BROUGHT THE WHOLE SYSTEM OUT AND WAS ABLE TO RETRIEVE THE BROKEN PART OF THE SHAFT. SUBSEQUENTLY, A 4.50 X 16MM SYNERGY XD WAS ADVANCED BUT MET THE SAME RESISTANCE, THIS TIME THE SHAFT BENT BUT DIDN'T BREAK. THEY TRIED ONE MORE BALLOON AND IT DIDN'T CROSS SO THEY STOPPED THE PROCEDURE AND SENT THE PATIENT TO SURGERY. NO PATIENT ISSUES BECAUSE OF THE STENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13450384 ↗MalfunctionApr 27, 2021Difficult to Remove; Difficult to Remove
IT WAS REPORTED THAT REMOVAL DIFFICULTIES WERE ENCOUNTERED. THE 80% STENOSED TARGET LESION WAS LOCATED IN A NON-TORTUOUS AND MODERATELY CALCIFIED VESSEL. A 4.50 X 16MM SYNERGY XD DRUG-ELUTING STENT WAS ADVANCED TO TREAT THE LESION. HOWEVER, AFTER DEPLOYING THE STENT, THE FULLY DEFLATED BALLOON WAS DIFFICULT TO WITHDRAW FROM THE STENT. THE BALLOON WAS REMOVED BY PULLING BACK ON CATHETER SHAFT. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11735655 ↗
Other Coronary drug-eluting stent devices
- SYNERGY™ H7493926038300Boston Scientific Corporation · Coronary drug-eluting stent8 deaths124 reports
- SYNERGY™ H7493926016300Boston Scientific Corporation · Coronary drug-eluting stent7 deaths106 reports
- SYNERGY™ H7493926020300Boston Scientific Corporation · Coronary drug-eluting stent9 deaths90 reports
- SYNERGY™ H7493926016250Boston Scientific Corporation · Coronary drug-eluting stent6 deaths82 reports
- XIENCE Sierra™ 1550350-33Abbott Vascular Inc. · Coronary drug-eluting stent1 recall32 reports
- Resolute Onyx™ RONYX30022UXMedtronic, Inc. · Coronary drug-eluting stent6 deaths76 reports
- SYNERGY™ H7493926024300Boston Scientific Corporation · Coronary drug-eluting stent3 deaths75 reports
- Resolute Onyx™ RONYX30015UXMedtronic, Inc. · Coronary drug-eluting stent3 deaths73 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page