Skip to content

SYNERGY™ XD H7493941816450Recalls & reported problems

Boston Scientific Corporation · Everolimus-Eluting Platinum Chromium Coronary Stent System · Class 3 · Implant

All models of SYNERGY XD (65)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729981190), received between Apr 27, 2021 and Dec 13, 2023 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Difficult to Advance1
  • Difficult to Remove1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 13, 2023Break; Break

    IT WAS REPORTED THAT SHAFT BREAK OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED CORONARY ARTERY. A 4.50 X 16MM SYNERGY XD DRUG-ELUTING STENT (DES) WAS ADVANCED FOR TREATMENT. HOWEVER, DURING ADVANCING, THE DELIVERY SYSTEM WAS FRACTURED INTO 2 PIECES AFTER FORCEDLY PUSHING IT INTO THE LESION. THE DEVICE WAS COMPLETELY REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A 4.0X16 SYNERGY DES. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18319434 ↗
  • MalfunctionFeb 4, 2022Difficult to Advance; Material Deformation; Difficult to Advance; Material Deformation

    IT WAS REPORTED THAT DIFFICULTY REMOVAL, SHAFT BREAK, AND PROCEDURE RESCHEDULING OCCURRED. THE 80% STENOSED TARGET LESION WAS LOCATED IN THE TORTUOUS AND CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 4.50 X 20MM SYNERGY XD DRUG-ELUTING STENT WAS ADVANCED HOWEVER THEY MET RESISTANCE IN THE VESSEL RIGHT AS THE STENT EXITED THE GUIDE CATHETER. THEY PUSHED AND PULLED AND NOTICE THE HYPOTUBE CAME OUT WITHOUT THE DISTAL PART OF THE DELIVERY SYSTEM. THEY BROUGHT THE WHOLE SYSTEM OUT AND WAS ABLE TO RETRIEVE THE BROKEN PART OF THE SHAFT. SUBSEQUENTLY, A 4.50 X 16MM SYNERGY XD WAS ADVANCED BUT MET THE SAME RESISTANCE, THIS TIME THE SHAFT BENT BUT DIDN'T BREAK. THEY TRIED ONE MORE BALLOON AND IT DIDN'T CROSS SO THEY STOPPED THE PROCEDURE AND SENT THE PATIENT TO SURGERY. NO PATIENT ISSUES BECAUSE OF THE STENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13450384 ↗
  • MalfunctionApr 27, 2021Difficult to Remove; Difficult to Remove

    IT WAS REPORTED THAT REMOVAL DIFFICULTIES WERE ENCOUNTERED. THE 80% STENOSED TARGET LESION WAS LOCATED IN A NON-TORTUOUS AND MODERATELY CALCIFIED VESSEL. A 4.50 X 16MM SYNERGY XD DRUG-ELUTING STENT WAS ADVANCED TO TREAT THE LESION. HOWEVER, AFTER DEPLOYING THE STENT, THE FULLY DEFLATED BALLOON WAS DIFFICULT TO WITHDRAW FROM THE STENT. THE BALLOON WAS REMOVED BY PULLING BACK ON CATHETER SHAFT. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11735655 ↗

Other Coronary drug-eluting stent devices

All Coronary drug-eluting stent devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page