SYNERGY™ XD H7493941828450Recalls & reported problems
Boston Scientific Corporation · Everolimus-Eluting Platinum Chromium Coronary Stent System · Class 3 · Implant
All models of SYNERGY XD (65)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729981220), received between May 11, 2021 and Oct 2, 2023 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Adverse Event Without Identified Device or Use Problem1
- Difficult to Advance1
- Failure to Advance1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Perforation1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 2, 2023Break; Failure to Advance; Break; Failure to Advance
IT WAS REPORTED THAT SHAFT BREAK OCCURRED. FOLLOWING PRE-DILATATION USING A 4.50 X 20 NC QUANTUM APEX AND A 3.50 X 12 EMERGE BALLOON CATHETERS, A 4.50 X 28MM SYNERGY XD DRUG-ELUTING STENT WAS ADVANCED FOR TREATMENT. HOWEVER, DURING THE PROCEDURE, THE STENT FAILED TO CROSS THE LESION AND THE SHAFT SNAPPED NEAR THE PROXIMAL HUB. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17857829 ↗MalfunctionNov 28, 2022Break; Difficult to Advance; Break; Difficult to Advance
IT WAS REPORTED THAT SHAFT BREAK OCCURRED. THE PATIENT PRESENTED WITH CORONARY ARTERY DISEASE AND UNDERWENT PERCUTANEOUS CORONARY INTERVENTION. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND MODERATELY CALCIFIED RIGHT CORONARY ARTERY. A 4.50 X 28MM SYNERGY XD DRUG-ELUTING STENT WAS ADVANCED THROUGH A NON-BOSTON SCIENTIFIC GUIDE CATHETER. HOWEVER, DURING INTRODUCTION, ADVANCING DIFFICULTY WAS ENCOUNTERED AND THE DEVICE BROKE MID-SHAFT. THE DEVICE WAS PULLED OUT AND THE PROCEDURE WAS COMPLETED WITH ANOTHER SYNERGY XD DRUG-ELUTING STENT. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15869211 ↗InjuryMay 11, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT VESSEL PERFORATION OCCURRED. THE GREATER THAN 70% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY CALCIFIED AND NON-TORTUOUS LEFT ANTERIOR DESCENDING ARTERY (LAD). FOLLOWING DEPLOYMENT OF A 28MM X 4.50MM SYNERGY XD DRUG-ELUTING STENT AND POST DILATION WITH A 20MM X 5.00MM NC QUANTUM APEX BALLOON CATHETER, A VESSEL PERFORATION WAS NOTED IN THE LAD. THE PERFORATION WAS COVERED WITH A STENT AND PERICARDIOCENTESIS WAS PERFORMED. THE PATIENT WAS STABLE POST-PROCEDURE.
Patient problems: Perforation
MAUDE report 11810297 ↗
Other Coronary drug-eluting stent devices
- SYNERGY™ H7493926038300Boston Scientific Corporation · Coronary drug-eluting stent8 deaths124 reports
- SYNERGY™ H7493926016300Boston Scientific Corporation · Coronary drug-eluting stent7 deaths106 reports
- SYNERGY™ H7493926020300Boston Scientific Corporation · Coronary drug-eluting stent9 deaths90 reports
- SYNERGY™ H7493926016250Boston Scientific Corporation · Coronary drug-eluting stent6 deaths82 reports
- XIENCE Sierra™ 1550350-33Abbott Vascular Inc. · Coronary drug-eluting stent1 recall32 reports
- Resolute Onyx™ RONYX30022UXMedtronic, Inc. · Coronary drug-eluting stent6 deaths76 reports
- SYNERGY™ H7493926024300Boston Scientific Corporation · Coronary drug-eluting stent3 deaths75 reports
- Resolute Onyx™ RONYX30015UXMedtronic, Inc. · Coronary drug-eluting stent3 deaths73 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page