Skip to content

SYNERGY™ XD H7493941832450Recalls & reported problems

Boston Scientific Corporation · Everolimus-Eluting Platinum Chromium Coronary Stent System · Class 3 · Implant

All models of SYNERGY XD (65)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 08714729981237), received between Jun 28, 2021 and May 15, 2026 — classified as 2 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Break” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break5
  • Difficult to Advance1
  • Failure to Deflate1
  • Inflation Problem1
  • Material Rupture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Cardiac Arrest1
  • Foreign Body In Patient1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 15, 2026Break; Difficult to Advance

    IT WAS REPORTED THAT SHAFT BREAK OCCURRED. THE PATIENT PRESENTED WITH CORONARY ARTERY DISEASE AND UNDERWENT PERCUTANEOUS CORONARY INTERVENTION. AFTER THE LESION WAS PRE-DILATED WITH GOOD EXPANSION, A NON-BOSTON SCIENTIFIC (NON-BSC) GUIDE CATHETER AND A NON-BSC GUIDEWIRE WERE USED. A 4.50 X 32 MM SYNERGY? XD DRUG-ELUTING STENT WAS ADVANCED FOR TREATMENT. THE PHYSICIAN ENCOUNTERED RESISTANCE DURING ADVANCEMENT IN THE PROXIMAL PORTION OF THE GUIDE JUST PAST THE TOUHY VALVE BUT WAS EVENTUALLY ABLE TO ADVANCE THE SYSTEM. THE STENT WAS SUCCESSFULLY DEPLOYED. DURING WITHDRAWAL OF THE DELIVERY SYSTEM FROM THE GUIDE, THE SHAFT BROKE IN HALF. THE ENTIRE DEVICE WAS REMOVED FROM THE PATIENT WITHOUT ISSUE, AND THE PROCEDURE WAS COMPLETED. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25186857 ↗
  • MalfunctionJan 20, 2026Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 75% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND MILDLY CALCIFIED CORONARY ARTERY. A 4.50 X 32MM SYNERGY XD DRUG-ELUTING STENT WAS SELECTED FOR TREATMENT. HOWEVER, DURING INFLATION, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED SUCCESSFULLY WITH NO FRAGMENTS LEFT, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED. IT WAS FURTHER REPORTED THAT THE TARGET LESION WAS LOCATED IN THE RIGHT CORONARY ARTERY. DURING INFLATION, CONTRAST FROM THE STENT BALLOON WAS COMING OUT FROM THE SHAFT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24125310 ↗
  • InjuryAug 1, 2025Break; Failure to Deflate; Break; Failure to Deflate

    IT WAS REPORTED THAT DEFLATION FAILURE AND SHAFT BREAK OCCURRED RESULTING IN CARDIAC ARREST, ADDITIONAL INTERVENTION AND AN UNRETRIEVED DEVICE FRAGMENT. THE TARGET LESION WAS LOCATED IN THE CIRCUMFLEX ARTERY. A STENT WAS INITIALLY PLACED IN THE CIRCUMFLEX ARTERY, FOLLOWED BY A 4.50 X 32MM SYNERGY XD DRUG-ELUTING STENT PLACED IN THE CIRCUMFLEX ARTERY EXTENDING TO THE LEFT MAIN OSTIUM. AFTER DEPLOYMENT, THE BALLOON FAILED TO DEFLATE. MULTIPLE UNSUCCESSFUL ATTEMPTS WERE MADE TO DEFLATE THE BALLOON WITH THE INFLATION DEVICE. A HORNET 14 GUIDEWIRE WAS THEN USED TO ATTEMPT TO PUNCTURE THE BALLOON. ADDITIONALLY, THE BACK END OF THE BALLOON WAS CUT OFF IN AN ATTEMPT TO RELEASE THE PRESSURE. WHEN THE PHYSICIAN TRIED TO RETRACT THE BALLOON, THE BALLOON SHAFT BROKE, LEAVING THE BALLOON STUCK IN THE LEFT MAIN/CIRCUMFLEX ARTERY. THE PATIENT BECAME UNSTABLE AND EXPERIENCED CARDIAC ARREST AND A MINIMALLY INVASIVE, TEMPORARY HEART PUMP WAS PLACED. A SECOND PHYSICIAN SCRUBBED IN AND WAS ABLE TO ADVANCE A NON-BOSTON SCIENTIFIC GUIDEWIRE AND A MICROCATHETER PAST THE BALLOON TO THE DISTAL CIRCUMFLEX. THEN, A 2.0 BALLOON WAS ADVANCED AND WAS ABLE TO RESTORE FLOW. THE PHYSICIAN USED PROGRESSIVELY BIGGER

    Patient problems: Cardiac Arrest; Foreign Body In Patient

    MAUDE report 22690330 ↗
  • InjuryMay 2, 2025Break; Inflation Problem; Break; Inflation Problem

    IT WAS REPORTED THAT A SHAFT BREAK OCCURRED REQUIRING ADDITIONAL INTERVENTION. A 4.50 X 32 MM SYNERGY XD DRUG-ELUTING STENT WAS ADVANCED FOR TREATMENT. DURING THE PROCEDURE, THE STENT BALLOON FAILED TO INFLATE. WHILE ATTEMPTING TO REMOVE THE DEVICE, THE PHYSICIAN BROKE THE SHAFT. THE REMAINING COMPONENTS OF THE SHAFT AND BALLOON WERE SUCCESSFULLY RETRIEVED FROM THE PATIENTS BODY USING A SNARE. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21952957 ↗
  • MalfunctionJul 20, 2023Break; Break

    IT WAS REPORTED THAT SHAFT BREAK OCCURRED. THE TARGET LESION WAS LOCATED IN LEFT FEMORAL ARTERY. A 4.50 X 32MM SYNERGY XD WAS ADVANCED FOR TREATMENT. HOWEVER, DURING INSERTION INTO THE GUIDE, THE CATHETER WAS BROKEN AND COULD NOT BE DELIVERED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17368503 ↗
  • MalfunctionJun 28, 2021Break; Break

    IT WAS REPORTED THAT SHAFT BREAK OCCURRED. THE TARGET LESION WAS LOCATED IN THE CORONARY ARTERY. A 4.50 X 32MM SYNERGY XD DRUG-ELUTING STENT WAS ADVANCED FOR TREATMENT. HOWEVER, DURING THE INSERTION, THE SHAFT OF THE DELIVERY SYSTEM BROKE. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12081715 ↗

Other Coronary drug-eluting stent devices

All Coronary drug-eluting stent devices →

Sources and updates