SYNERGY™ XD H7493941828500Recalls & reported problems
Boston Scientific Corporation · Everolimus-Eluting Platinum Chromium Coronary Stent System · Class 3 · Implant
All models of SYNERGY XD (65)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08714729981282), received between May 24, 2022 and Aug 13, 2024 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Failure to Advance2
- Difficult to Remove1
- Material Rupture1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 13, 2024Break; Material Rupture; Break; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE AND SHAFT BREAK OCCURRED. THE PATIENT UNDERWENT A CHRONIC TOTAL OCCLUSION (CTO) PROCEDURE. A 5.00 X 28MM SYNERGY XD DRUG-ELUTING STENT WAS ADVANCED FOR TREATMENT. HOWEVER, DURING POST-DILATION, THE BALLOON MAINTAINED A POCKET OF FLUID AT THE DISTAL END OF THE STENT BALLOON, AND IT WAS ALSO NOTED THAT THE SHAFT SHEARED AND BROKE INTO TWO PIECES. THE PATIENT CONDITION WAS STABLE POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19978283 ↗MalfunctionJun 25, 2024Difficult to Remove; Difficult to Remove
IT WAS REPORTED THAT REMOVAL DIFFICULTIES WERE ENCOUNTERED. THE PATIENT WAS BEING TREATED FOR BLOCKAGE AND UNDERWENT PERCUTANEOUS CORONARY INTERVENTION. A 5.00 X 28MM SYNERGY XD DRUG-ELUTING STENT WAS ADVANCED FOR TREATMENT. HOWEVER, RETRACTING THE DEVICE FROM THE CATHETER WAS MORE DIFFICULT THAN USUAL. THE DEVICE WAS PULLED OUT AND NO PHYSICAL DAMAGE WAS OBSERVED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19606717 ↗MalfunctionJun 30, 2022Break; Failure to Advance; Break; Failure to Advance
IT WAS REPORTED THAT SHAFT BREAK OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY-SEVERELY TORTUOUS AND MILDLY CALCIFIED RIGHT CORONARY ARTERY. A 5.00 X 28MM SYNERGY XD DRUG-ELUTING STENT WAS ADVANCED FOR TREATMENT. HOWEVER, AS THE PHYSICIAN WAS DRIVING THE STENT CATHETER UP THE GUIDE, RESISTANCE WAS ENCOUNTERED AND THE PROXIMAL PORTION OF THE SHAFT SNAPPED WHEN REACHING A TORTUOUS PART OF ANATOMY. FURTHERMORE, MULTIPLE ATTEMPTS WERE MADE TO CROSS THE LESION. THE PHYSICIAN WAS ABLE TO PULL AND REMOVE IT OUT FROM THE GUIDE. A NEW SYNERGY XD STENT WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14877269 ↗MalfunctionMay 24, 2022Break; Failure to Advance; Break; Failure to Advance
IT WAS REPORTED THAT SHAFT BREAK OCCURRED. A 5.00 X 28MM SYNERGY XD DRUG-ELUTING STENT WAS ADVANCED BUT FAILED TO CROSS THE LESION AND IT WAS FOUND THAT THE SHAFT BROKE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14490969 ↗
Other Coronary drug-eluting stent devices
- SYNERGY™ H7493926038300Boston Scientific Corporation · Coronary drug-eluting stent8 deaths124 reports
- SYNERGY™ H7493926016300Boston Scientific Corporation · Coronary drug-eluting stent7 deaths106 reports
- SYNERGY™ H7493926020300Boston Scientific Corporation · Coronary drug-eluting stent9 deaths90 reports
- SYNERGY™ H7493926016250Boston Scientific Corporation · Coronary drug-eluting stent6 deaths82 reports
- XIENCE Sierra™ 1550350-33Abbott Vascular Inc. · Coronary drug-eluting stent1 recall32 reports
- Resolute Onyx™ RONYX30022UXMedtronic, Inc. · Coronary drug-eluting stent6 deaths76 reports
- SYNERGY™ H7493926024300Boston Scientific Corporation · Coronary drug-eluting stent3 deaths75 reports
- Resolute Onyx™ RONYX30015UXMedtronic, Inc. · Coronary drug-eluting stent3 deaths73 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page