SYNERGY™ XD H7493941832500Recalls & reported problems
Boston Scientific Corporation · Everolimus-Eluting Platinum Chromium Coronary Stent System · Class 3 · Implant
All models of SYNERGY XD (65)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 08714729981299), received between Dec 16, 2020 and Aug 29, 2025 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Device Dislodged or Dislocated2
- Failure to Advance2
- Adverse Event Without Identified Device or Use Problem1
- Deflation Problem1
- Difficult to Advance1
- Material Rupture1
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Insufficient Information1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 29, 2025Nonstandard Device; Difficult to Advance; Device Dislodged or Dislocated; Adverse Event Without Identified Device or Use Problem; Nonstandard Device; Difficult to Advance; Device Dislodged or Dislocated; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE STENT MOVED ON THE BALLOON. THE PATIENT UNDERWENT PERCUTANEOUS CORONARY INTERVENTION (PCI). A 5.00 X 32MM SYNERGY XD DRUG-ELUTING STENT (DES) WAS SELECTED FOR TREATMENT. DURING INSERTION, THE STENT FAILED TO ADVANCE IN THE 6FR NON-BOSTON SCIENTIFIC GUIDE CATHETER, AND IT WAS NOTED THAT THE STENT UNRAVELED FROM THE CATHETER. THE STENT DID REMAIN ON THE BALLOON WITH NO ISSUES TO THE PATIENT. A NON-BSC STENT WAS SUCCESSFULLY DEPLOYED TO TREAT THE LESION AND COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22929213 ↗MalfunctionFeb 10, 2023Break; Failure to Advance; Break; Failure to Advance
IT WAS REPORTED THAT SHAFT BREAK OCCURRED. THE PATIENT PRESENTED FOR PERCUTANEOUS CORONARY INTERVENTION. A 5.00 X 32MM SYNERGY XD DRUG-ELUTING STENT WAS ADVANCED FOR TREATMENT. HOWEVER, DURING THE PROCEDURE, THE DEVICE FAILED TO CROSS THE LESION AND THE SHAFT BROKE INSIDE THE PATIENT. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16350895 ↗MalfunctionFeb 8, 2023Break; Break
IT WAS REPORTED THAT SHAFT BREAK OCCURRED. THE PATIENT UNDERWENT PERCUTANEOUS CORONARY INTERVENTION. A 5.00 X 32MM SYNERGY XD DRUG-ELUTING STENT WAS ADVANCED TO TREAT THE LESION. HOWEVER, DURING THE PROCEDURE, THE DELIVERY SYSTEM BROKE WHILE ATTEMPTING DEPLOYMENT. THE PROCEDURE WAS COMPLETED. NO PATIENT COMPLICATIONS NOR INJURIES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16321685 ↗MalfunctionSep 23, 2021Deflation Problem; Failure to Advance; Deflation Problem; Failure to Advance
IT WAS REPORTED THAT FAILURE TO DEFLATE WAS ENCOUNTERED. A PERCUTANEOUS CORONARY INTERVENTION WAS BEING PERFORMED. A 5.00 X 32MM SYNERGY XD DRUG ELUTING STENT WAS ADVANCED FOR TREATMENT BUT THE STENT WAS UNABLE TO CROSS THE LESION. HOWEVER, IT WAS NOTED THAT THE BALLOON WOULD NOT DEFLATE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12521395 ↗MalfunctionApr 28, 2021Material Separation; Material Separation
IT WAS REPORTED THAT BALLOON TEAR OCCURRED. A 5.00 X 32MM SYNERGY XD DRUG-ELUTING STENT WAS USED, HOWEVER; IT WAS NOTED THAT THE BALLOON WAS TORN OFF FROM THE CATHETER. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11740757 ↗MalfunctionMar 8, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE BALLOON ALMOST DETACHED. DURING A PROCEDURE INVOLVING A 5.00 X 32MM SYNERGY XD IT WAS NOTED THAT THE BALLOON ALMOST DETACHED. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 11440434 ↗MalfunctionDec 16, 2020Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 5.00 X 32MM SYNERGY XD DRUG-ELUTING STENT WAS ADVANCED FOR TREATMENT. HOWEVER, DURING PROCEDURE, IT WAS NOTICED THAT THERE WAS A BLOOD CAME BACK UP INTO THE ENDOFLATOR WHICH WAS ATTACHED TO THE MONORAIL PORTION OF THE STENT. WHEN THE DEVICE WAS PULLED OUT, A HOLE IN THE BALLOON PORTION OF THE STENT DELIVERY SYSTEM WAS NOTED. THE PROCEDURE WAS COMPLETED WITH ANOTHER STENT AND NO PATIENT COMPLICATIONS WERE REPORTED. IT WAS FURTHER REPORTED THAT THE 80%-90% STENOSED, NON-TORTUOUS, ECCENTRIC TARGET LESION WAS LOCATED IN THE MID RIGHT CORONARY ARTERY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11017916 ↗
Other Coronary drug-eluting stent devices
- SYNERGY™ H7493926038300Boston Scientific Corporation · Coronary drug-eluting stent8 deaths124 reports
- SYNERGY™ H7493926016300Boston Scientific Corporation · Coronary drug-eluting stent7 deaths106 reports
- SYNERGY™ H7493926020300Boston Scientific Corporation · Coronary drug-eluting stent9 deaths90 reports
- SYNERGY™ H7493926016250Boston Scientific Corporation · Coronary drug-eluting stent6 deaths82 reports
- XIENCE Sierra™ 1550350-33Abbott Vascular Inc. · Coronary drug-eluting stent1 recall32 reports
- Resolute Onyx™ RONYX30022UXMedtronic, Inc. · Coronary drug-eluting stent6 deaths76 reports
- SYNERGY™ H7493926024300Boston Scientific Corporation · Coronary drug-eluting stent3 deaths75 reports
- Resolute Onyx™ RONYX30015UXMedtronic, Inc. · Coronary drug-eluting stent3 deaths73 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page