OrcaPod™ SUV-628Recalls & reported problems
Boston Scientific Corporation · Single Use Air/Water, Suction and Biopsy Valves Set · Class 2
All models of OrcaPod (4)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Boston Scientific OrcaPod suction and biopsy valve set is used for irrigation and suction during endoscopy. The 71 reports mostly describe the suction button leaking fluid, sticking when pressed, suction cutting in and out, and parts coming loose, with no deaths and no injuries. No recalls name this device. Reports rose from 2 in the earlier 12 months to 41 in the most recent 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 46
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 46 adverse event reports that list this exact device (DI 08714729981558), received between Jun 18, 2019 and Sep 11, 2026 — classified as 46 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (30 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Leak/Splash30
- Detachment of Device or Device Component19
- Physical Resistance/Sticking16
- Suction Problem13
- Fluid/Blood Leak12
- Difficult to Remove5
- Defective Device3
- Break2
Patient problems
- No Clinical Signs, Symptoms or Conditions32
- No Consequences Or Impact To Patient13
- No Patient Involvement1
Recent adverse event reports
The 25 most recent of 46 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 11, 2026Difficult to Remove; Physical Resistance/Sticking
It was reported to Boston Scientific Corporation that an Orca Valve was used during an Esophagogastroduodenoscopy (EGD) procedure for the treatment of tissue resection performed on (b)(6) 2026.\\During the procedure, the physician observed that the button became stuck within the scope. Upon completion of the procedure, the button was difficult to remove. The procedure was completed with another of the same device. \\There were no patient complications reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50735035 ↗MalfunctionSep 11, 2026Difficult to Remove; Physical Resistance/Sticking
It was reported to Boston Scientific Corporation that an Orca Valve was used in the esophagus during an Esophagogastroduodenoscopy (EGD) procedure for the treatment of tissue resection performed on(b)(6) 2026.\\During the procedure, the physician observed that the button became stuck within the scope. Upon completion of the procedure, the button could not be removed. The procedure was completed with another of the same device. \\There were no patient complications reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50734691 ↗MalfunctionSep 11, 2026Break; Difficult to Remove; Material Integrity Problem; Physical Resistance/Sticking
It was reported to Boston Scientific Corporation that an Orca Valve was used in the esophagus during an Esophagogastroduodenoscopy (EGD) procedure for the treatment of tissue resection performed on (b)(6) 2026.\\During the procedure, the physician observed that the button became stuck within the scope. Upon completion of the procedure, the button could not be removed and ultimately broke off inside the scope. The procedure was completed with another of the same device. \\There were no patient complications reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50734550 ↗MalfunctionAug 18, 2026Physical Resistance/Sticking
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50310224 ↗MalfunctionAug 18, 2026Physical Resistance/Sticking
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50310219 ↗MalfunctionAug 18, 2026Fluid/Blood Leak; Suction Problem; Physical Resistance/Sticking
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50291261 ↗MalfunctionAug 18, 2026Fluid/Blood Leak; Suction Problem; Physical Resistance/Sticking
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50291254 ↗MalfunctionAug 18, 2026Fluid/Blood Leak; Suction Problem; Physical Resistance/Sticking
It was reported to Boston Scientific Corporation that OrcaPod 3 was used in the colon during a Colonoscopy procedure performed on an unknown procedure date.During the procedure, the red suction button exhibited leakage, sticking when depressed, and intermittent suction problem. When suction stopped functioning, the physician would remove the suction button and occlude the suction port with a finger to restore suction. Although this temporarily resolved the issue, the problem repeatedly recurred throughout the procedure. The procedure was completed with this device.There were no patient complications reported as a result of this event.Attempts to obtain additional information regarding the circumstances surrounding this event have been unsuccessful to date. Should additional relevant details become available, a supplemental report will be submitted.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50291253 ↗MalfunctionAug 14, 2026Fluid/Blood Leak; Suction Problem; Physical Resistance/Sticking
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50245629 ↗MalfunctionAug 14, 2026Fluid/Blood Leak; Suction Problem; Physical Resistance/Sticking
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50245615 ↗MalfunctionAug 14, 2026Fluid/Blood Leak; Suction Problem; Physical Resistance/Sticking
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50245614 ↗MalfunctionAug 14, 2026Fluid/Blood Leak; Suction Problem; Physical Resistance/Sticking
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50245611 ↗MalfunctionAug 14, 2026Fluid/Blood Leak; Suction Problem; Physical Resistance/Sticking
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50245610 ↗MalfunctionAug 13, 2026Fluid/Blood Leak; Suction Problem; Physical Resistance/Sticking
It was reported to Boston Scientific Corporation that OrcaPod 3 was used in the colon during a Colonoscopy procedure performed on an unknown procedure date.During the procedure, the red suction button exhibited leakage, sticking when depressed, and intermittent suction problem. When suction stopped functioning, the physician would remove the suction button and occlude the suction port with a finger to restore suction. Although this temporarily resolved the issue, the problem repeatedly recurred throughout the procedure. The procedure was completed with this device.There were no patient complications reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50243509 ↗MalfunctionJul 10, 2026Suction Problem; Infusion or Flow Problem; Physical Resistance/Sticking
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT ORCA VALVE WAS USED DURING A PROCEDURE PERFORMED ON (B)(6) 2026. IT WAS REPORTED THAT THE ORCA BUTTON IS NOT INSUFFLATING PROPERLY. THE SUCTION BUTTON BECOMES STUCK IN THE DEPRESSED POSITION, RESULTING IN INADEQUATE SUCTION. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25787076 ↗MalfunctionMar 4, 2026Fluid/Blood Leak; Leak/Splash
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SEAL BIOPSY VALVE WAS USED DURING AN ENDOSCOPIC MUCOSAL RESECTION (EMR) PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, ONCE THE SNARE WAS INSERTED, BACKFLOW AND BODY FLUID SPURTED OUT. THE PROCEDURE WAS COMPLETED USING ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24506808 ↗MalfunctionAug 14, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ORCA WAS USED DURING A ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE AIR/WATER VALVE WAS OBSERVED TO BE LEAKING WATER. THE PROCEDURE WAS COMPLETED USING ANOTHER ORCA. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22807758 ↗MalfunctionAug 1, 2025Fluid/Blood Leak; Leak/Splash; Suction Problem; Defective Device; Physical Resistance/Sticking; Fluid/Blood Leak; Leak/Splash; Suction Problem; Defective Device; Physical Resistance/Sticking
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ORCAPOD WAS USED DURING AN UNKNOWN PROCEDURE PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, IT WAS NOTICED THAT THE ORCA BUTTONS WERE STIFFER. THE A/W AND SUCTION BUTTONS WERE OBSERVED TO BE STICKING, DIFFICULT TO PRESS DURING USE, AND CLOGGING MORE EASILY. THERE WAS NO LEAK OBSERVED, HOWEVER, WHEN REMOVING THE ORCA BUTTONS, THE PHYSICIAN WAS SPLASHED WITH BILE. THE PROCEDURE WAS COMPLETED USING A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION REGARDING THE CIRCUMSTANCES SURROUNDING THIS EVENT HAVE BEEN UNSUCCESSFUL TO DATE. SHOULD ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22687012 ↗MalfunctionNov 18, 2021Leak/Splash; Detachment of Device or Device Component; Leak/Splash; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SEAL BIOPSY VALVE WAS USED DURING A COLONOSCOPY PROCEDURE PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, THE SEAL BIOPSY VALVE CAP WAS POPPING OFF DURING THE DEVICE PASSAGE AND SPRAYING/LEAKING. THE PROCEDURE WAS COMPLETED USING THE ORIGINAL SEAL BIOPSY VALVE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12834855 ↗MalfunctionNov 17, 2021Leak/Splash; Detachment of Device or Device Component; Leak/Splash; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SEAL BIOPSY VALVE WAS USED DURING A COLONOSCOPY PROCEDURE PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, THE SEAL BIOPSY VALVE CAP WAS POPPING OFF DURING THE DEVICE PASSAGE AND SPRAYING/LEAKING. THE PROCEDURE WAS COMPLETED USING THE ORIGINAL SEAL BIOPSY VALVE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12830131 ↗MalfunctionNov 17, 2021Leak/Splash; Detachment of Device or Device Component; Leak/Splash; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SEAL BIOPSY VALVE WAS USED DURING A COLONOSCOPY PROCEDURE PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, THE SEAL BIOPSY VALVE CAP WAS POPPING OFF DURING THE DEVICE PASSAGE AND SPRAYING/LEAKING. THE PROCEDURE WAS COMPLETED USING THE ORIGINAL SEAL BIOPSY VALVE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12827551 ↗MalfunctionNov 17, 2021Leak/Splash; Detachment of Device or Device Component; Leak/Splash; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SEAL BIOPSY VALVE WAS USED DURING A COLONOSCOPY PROCEDURE PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, THE SEAL BIOPSY VALVE CAP WAS POPPING OFF DURING THE DEVICE PASSAGE AND SPRAYING/LEAKING. THE PROCEDURE WAS COMPLETED USING THE ORIGINAL SEAL BIOPSY VALVE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12827549 ↗MalfunctionNov 17, 2021Leak/Splash; Detachment of Device or Device Component; Leak/Splash; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SEAL BIOPSY VALVE WAS USED DURING A COLONOSCOPY PROCEDURE PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, THE SEAL BIOPSY VALVE CAP WAS POPPING OFF DURING THE DEVICE PASSAGE AND SPRAYING/LEAKING. THE PROCEDURE WAS COMPLETED USING THE ORIGINAL SEAL BIOPSY VALVE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12827548 ↗MalfunctionNov 17, 2021Leak/Splash; Detachment of Device or Device Component; Leak/Splash; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SEAL BIOPSY VALVE WAS USED DURING A COLONOSCOPY PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, THE SEAL BIOPSY VALVE CAP WAS POPPING OFF DURING THE DEVICE PASSAGE AND SPRAYING/LEAKING. THE PROCEDURE WAS COMPLETED USING THE ORIGINAL SEAL BIOPSY VALVE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12827547 ↗MalfunctionNov 16, 2021Leak/Splash; Detachment of Device or Device Component; Leak/Splash; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SEAL BIOPSY VALVE WAS USED DURING A COLONOSCOPY PROCEDURE PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, THE SEAL BIOPSY VALVE CAP WAS POPPING OFF DURING THE DEVICE PASSAGE AND SPRAYING/LEAKING. THE PROCEDURE WAS COMPLETED USING THE ORIGINAL SEAL BIOPSY VALVE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12824503 ↗
FDA clearance and classification
The FDA regulates product code OCX (endoscopic irrigation/suction system) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the OrcaPod™ SUV-628
Has the OrcaPod™ SUV-628 been recalled?
No FDA recall record in our data names this exact device identifier (DI 08714729981558), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the OrcaPod™ SUV-628?
The FDA's MAUDE database lists 46 adverse event reports for this device, received between Jun 18, 2019 and Sep 11, 2026. In those reports, the most common specific device problems are “Leak/Splash” (30), “Detachment of Device or Device Component” (19) and “Physical Resistance/Sticking” (16). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the OrcaPod™ SUV-628?
None of the reports are classified as deaths, 0 as injuries and 46 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the OrcaPod™ SUV-628, and is it still sold?
The FDA's device database lists Boston Scientific Corporation as the company responsible for it (brand OrcaPod™, model SUV-628). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other endoscopic irrigation/suction system devices
- ERBEFLO CleverCap® 20325-203Erbe Medical, LLC · Endoscopic Irrigation/Suction System1 recall0 reports
- andorate® A/W Channel Cleaning Adapter + 4-piece Valve Kit GAR3846GA Health Company Limited · Endoscopic Irrigation/Suction System1 recall0 reports
- ERBEFLO CleverCap® 20325-201Erbe Medical, LLC · Endoscopic Irrigation/Suction System1 recall0 reports
- 4-piece valve kit(Suction, Air/Water, Biopsy Valves & Auxiliary Water Connector) EN10229.1GA Health Company Limited · Endoscopic Irrigation/Suction System1 recall0 reports
- ERBEFLO® 2 20325-222Erbe Medical, LLC · Endoscopic Irrigation/Suction System1 recall0 reports
- 2-piece valve kit (Suction and Air/Water Valve) EN10229.2GA Health Company Limited · Endoscopic Irrigation/Suction System1 recall0 reports
- BullDog® 1203-71Diversatek Healthcare, Inc. · Endoscopic Irrigation/Suction System1 recall0 reports
- Suction Valve EN10232GA Health Company Limited · Endoscopic Irrigation/Suction System1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page