OrcaPod™ SUV-629Recalls & reported problems
Boston Scientific Corporation · Single Use Air/Water, Suction and Biopsy Valve Set with Water Jet Connector · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
Quick summary
This Boston Scientific OrcaPod is a single-use valve set for endoscopes. Its 101 FDA reports include 3 injuries and no deaths, and mostly describe fluid leaking from the air and water button and suction buttons sticking, with some reports of parts detaching. Reports rose to 48 in the last 12 months from 4 before. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 112
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 112 adverse event reports that list this exact device (DI 08714729981572), received between May 23, 2019 and Sep 29, 2026 — classified as 3 injuries, 109 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (44 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak44
- Leak/Splash41
- Physical Resistance/Sticking36
- Detachment of Device or Device Component25
- Infusion or Flow Problem15
- Break4
- Suction Problem4
- Material Fragmentation3
Patient problems
- No Clinical Signs, Symptoms or Conditions86
- No Consequences Or Impact To Patient16
- No Known Impact Or Consequence To Patient5
- Exposure to Body Fluids2
- Discomfort1
- No Patient Involvement1
- Perforation1
Recent adverse event reports
The 25 most recent of 112 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026Physical Resistance/Sticking
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50935996 ↗MalfunctionSep 14, 2026Physical Resistance/Sticking
It was reported to Boston Scientific Corporation that an OrcaPod 4 was used in the colon during various endoscopy procedures performed on (b)(6) 2026. During the procedure, the physician reported that the suction button stuck in the depressed position. There were no patient complications reported as a result of this event and the patient's condition is stable.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50758400 ↗MalfunctionSep 14, 2026Physical Resistance/Sticking
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50757628 ↗MalfunctionSep 14, 2026Nonstandard Device; Physical Resistance/Sticking
It was reported to Boston Scientific Corporation that an OrcaPod 4 was used in the colon during various endoscopy procedures performed on (b)(6) 2026. During the procedure, the physician reported that the suction button stuck in the depressed position. It was also reported that the button detached and popped off the scope.There were no patient complications reported as a result of this event and the patient's condition is stable.Attempts to obtain additional information regarding the circumstances surrounding this event have been unsuccessful to date. Should additional relevant details become available, a supplemental report will be submitted.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50757275 ↗MalfunctionSep 10, 2026Break; Difficult to Remove; Material Integrity Problem; Physical Resistance/Sticking
It was reported to Boston Scientific Corporation that an Orca valve was used during a colonoscopy procedure performed on an unknown date.During the procedure, the red suction button became stuck in the depressed position and did not return to its standard position. Attempts to remove and replace the suction button were unsuccessful, and a hemostat was required to extract the device. The procedure was completed using another Orca suction button of the same type.Following the procedure, the endoscope was sent for repair. Upon return of the scope, facility personnel performed a functional test and observed that the suction button could not be properly seated in the suction port. Further inspection of the suction well revealed a broken fragment of the suction button shaft lodged within the port. No visible damage had been identified at the time of the original procedure, and the shaft fracture was not recognized until the subsequent inspection. The issue was isolated to the red suction button; the blue air/water button was not involved.No patient complication was reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50714637 ↗MalfunctionSep 7, 2026Physical Resistance/Sticking
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50658830 ↗MalfunctionSep 1, 2026Fluid/Blood Leak
It was reported to Boston Scientific Corporation that an OrcaPod 4 was used during procedure performed on an unknown procedure date.During the procedure, the nurse reported that the air/water button leaked continuously throughout the case; requiring her to place a towel beneath the physician to absorb the excess water. The procedure was completed with this device.There were no patient complications reported as a result of this event and the patient's condition is stable.Attempts to obtain additional information regarding the circumstances surrounding this event have been unsuccessful to date. Should additional relevant details become available, a supplemental report will be submitted.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50534788 ↗MalfunctionSep 1, 2026Break; Difficult to Remove; Device Contamination with Chemical or Other Material
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50534786 ↗MalfunctionAug 24, 2026Physical Resistance/Sticking
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50419128 ↗MalfunctionAug 19, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50326193 ↗MalfunctionAug 19, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50326188 ↗MalfunctionAug 18, 2026Fluid/Blood Leak
It was reported to Boston Scientific Corporation that an OrcaPod 4 was used in the gastrointestinal tract during a Gastroscopy procedure for the treatment of Polyps performed on (b)(6) 2026. During the procedure, the nurse observed that the water was escaping from the Air / Watter button. The procedure was completed using another of the same device.There was no patient complications reported as a result of this event and the patient's condition is stable.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50291898 ↗MalfunctionAug 18, 2026Fluid/Blood Leak
It was reported to Boston Scientific Corporation that an OrcaPod 4 was used during a Colonoscopy procedure on (b)(6) 2026. During the procedure, the nurse observed that the water was escaping from the Air / Watter button. The procedure was completed using this device.There were no patient complications reported as a result of this event and the patient's condition is stable.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50291897 ↗MalfunctionAug 18, 2026Suction Problem; Physical Resistance/Sticking
It was reported to Boston Scientific that an Orca valve was used for colon during a Colonoscopy procedure performed on July 21, 2026, for colonoscopy screening.During procedure, the suction button sticking. The procedure was completed using the same device.There were no patient complications as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50291249 ↗MalfunctionAug 5, 2026Unsealed Device Packaging; Device Contaminated During Manufacture or Shipping; Packaging Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50136764 ↗MalfunctionJul 27, 2026Physical Resistance/Sticking
It was reported to Boston Scientific Corporation that an OrcaPod 4 was used in the colon during a Colonoscopy procedure for the treatment of Polyps performed on (b)(6)2026. During the procedure, the suction button was sticking and was not returning to its original position after being pressed. The procedure was completed using this device.There were no patient complications reported as a result of this event and the patient's condition is stable.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50020387 ↗MalfunctionJul 27, 2026Physical Resistance/Sticking
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50020075 ↗MalfunctionJul 26, 2026Fluid/Blood Leak; Use of Device Problem; Physical Resistance/Sticking
Patient problems: Exposure to Body Fluids
MAUDE report 50000031 ↗MalfunctionJul 20, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25880612 ↗MalfunctionJul 10, 2026Suction Problem; Infusion or Flow Problem; Physical Resistance/Sticking
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT ORCA VALVE WAS USED DURING A PROCEDURE PERFORMED ON (B)(6) 2026. IT WAS REPORTED THAT THE ORCA BUTTON IS NOT INSUFFLATING PROPERLY. THE SUCTION BUTTON BECOMES STUCK IN THE DEPRESSED POSITION, RESULTING IN INADEQUATE SUCTION. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25788511 ↗MalfunctionJul 10, 2026Break; Suction Problem; Infusion or Flow Problem; Physical Resistance/Sticking
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT ORCA VALVE WAS USED DURING COLONOSCOPY AND ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURES PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THEY REPORTED FREQUENT ISSUES WITH THE NEW BUTTONS, INCLUDING WEAK AND NO INSUFFLATION, INADEQUATE SUCTION, THE SUCTION BUTTON BECOMING STUCK IN THE DEPRESSED POSITION, AND ONE INSTANCE IN WHICH THE BUTTON STEM BROKE OFF INTO THE SCOPE VALVE DURING BEDSIDE CLEANING. THE PROCEDURE WAS COMPLETED USING AN ALTERNATIVE DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25788503 ↗MalfunctionMay 29, 2026Fluid/Blood Leak; Infusion or Flow Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT ORCA VALVE WAS USED DURING PROCEDURE PERFORMED ON UNKNOWN DATE. DURING THE PROCEDURE, THE AIR/WATER BUTTON WAS LEAKING WATER AND NOT PROVIDING ENOUGH INSUFFLATION. THE PROCEDURE WAS COMPLETED USING WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25333518 ↗MalfunctionMay 29, 2026Fluid/Blood Leak; Infusion or Flow Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT ORCA VALVE WAS USED DURING A PROCEDURE PERFORMED ON UNKNOWN DATE. DURING THE PROCEDURE, THE AIR/WATER BUTTON WAS LEAKING WATER AND NOT PROVIDING ENOUGH INSUFFLATION. THE PROCEDURE WAS COMPLETED USING WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25333515 ↗MalfunctionMay 29, 2026Fluid/Blood Leak; Infusion or Flow Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT ORCA VALVE WAS USED DURING PROCEDURE PERFORMED ON UNKNOWN DATE. DURING THE PROCEDURE, THE AIR/WATER BUTTON WAS LEAKING WATER AND NOT PROVIDING ENOUGH INSUFFLATION. THE PROCEDURE WAS COMPLETED USING WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25333514 ↗MalfunctionMay 29, 2026Fluid/Blood Leak; Infusion or Flow Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT ORCA VALVE WAS USED DURING PROCEDURE PERFORMED ON UNKNOWN DATE. DURING THE PROCEDURE, THE AIR/WATER BUTTON WAS LEAKING WATER AND NOT PROVIDING ENOUGH INSUFFLATION. THE PROCEDURE WAS COMPLETED USING WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25333498 ↗
Other Endoscope channel accessory devices
- RX Locking Device and Biopsy Cap M00545260Boston Scientific Corporation · Endoscope channel accessory216 reports
- Orca™ SUV-617Boston Scientific Corporation · Endoscope Channel Accessory1 recall64 reports
- andorate® Disposable Endoscope Valves Set GAR004GA Health Company Limited · Endoscope Channel Accessory1 recall1 report
- BullDog 1203-72Diversatek Healthcare, Inc. · Endoscope Channel Accessory1 recall0 reports
- 3-piece valve kit (Suction, Air/Water and Biopsy Valves) EN10229GA Health Company Limited · Endoscope Channel Accessory1 recall0 reports
- Ruhof Scope Valet Guardian 345SVVO4Ruhof Corporation · Endoscope Channel Accessory1 recall0 reports
- AutoCap™ RX M00545100Boston Scientific Corporation · Endoscope channel accessory26 reports
- OrcaPod™ SUV-633Boston Scientific Corporation · Endoscope channel accessory24 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page