SYNERGY MEGATRON™ H7493942824350Recalls & reported problems

Boston Scientific Corporation · Everolimus-Eluting Platinum Chromium Coronary Stent System · Class 3 · Implant

All models of SYNERGY MEGATRON (28)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729985693), received between Jan 23, 2023 and Mar 25, 2025 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Activation Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 25, 2025Break; Break

    IT WAS REPORTED THAT SHAFT BREAK OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. AN OPTICROSS IMAGING CATHETER WAS USED BUT BECAME KINKED INSIDE THE PATIENT DURING DELIVERY POST-STENT PLACEMENT, RESULTING IN A LOSS OF IMAGING. THE DEVICE WAS REMOVED, AND A 2.25 X 38 MM SYNERGY XD DRUG-ELUTING STENT (DES) WAS SELECTED FOR TREATMENT. HOWEVER, THE STENT BROKE AT THE PROXIMAL STAINLESS-STEEL SHAFT AND SNAPPED INSIDE THE PATIENT. THE DEVICE WAS REMOVED, AND ANOTHER 3.50 X 24 MM SYNERGY XD DES WAS THEN SELECTED, BUT IT ALSO BROKE AT THE PROXIMAL STAINLESS-STEEL SHAFT AND SNAPPED INSIDE THE PATIENT. THE DEVICE WAS ALSO REMOVED BY JUST PULLING THEM BACK OUT OVER THE WIRE. THE HUB OF BOTH STENTS WAS DETACHED OUTSIDE THE BODY. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE, AND NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21689224 ↗
  • InjuryJan 2, 2025Activation Failure; Activation Failure

    VITALYST EARLY FEASIBILITY STUDY. IT WAS REPORTED THAT STENT WAS NOT WELL APPOSED. DURING THE INDEX PROCEDURE, THE 80% STENOSIS IN THE LEFT OSTIAL TO PROXIMAL LEFT ANTERIOR DESCENDING ARTERY (LAD) AND IN THE LCX WAS TREATED BY ANGIOPLASTY USING 2.50 MM NON-COMPLIANT BALLOONS FOLLOWED BY INTRAVASCULAR ULTRASOUND IMAGING (IVUS) OF EACH VESSEL. SUBSEQUENTLY, SEVERE CALCIFICATION IN THE PROXIMAL LAD AND PROXIMAL LCX WAS TREATED WITH ANGIOPLASTY USING A 3.00 MM NON-COMPLIANT BALLOON AND A NON-BOSTON SCIENTIFIC LITHOTRIPSY BALLOON FOLLOWED BY ANGIOPLASTY USING A 3.50 MM WOLVERINE CUTTING BALLOON. TWO 3.50 MM X 24.00 MM SYNERGY MEGATRON DES WERE IMPLANTED FOLLOWED BY PROXIMAL OPTIMIZATION TECHNIQUE USING A 4.00 MM BALLOON. A PROXIMAL MAL APPOSITION NOTED IN THE LMCA SEGMENT OF THE STENT WAS POST-DILATED USING A 5.00 MM BALLOON. IN ADDITION, A CONTAINED PERFORATION WAS NOTED IN THE LMCA WHICH WAS TREATED BY BALLOON ANGIOPLASTY WITH SERIAL INFLATIONS USING A 4.00 MM PTCA BALLOON WHICH STOPPED ALL EXTRAVASATIONS. ECHOCARDIOGRAM REVEALED GOOD RESULT WITH NO PERICARDIAL EFFUSION. POST INTERVENTION LESION ASSESSMENT REVEALED, TIMI 3 FLOW AND 0% RESIDUAL STENOSIS. ESTIMATED BLOOD LOSS WAS NOTED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21060440 ↗
  • MalfunctionJan 23, 2023Break; Break

    IT WAS REPORTED THAT SHAFT BREAK OCCURRED. THE PATIENT PRESENTED WITH ANGINA AND WAS BEING TREATED FOR PERCUTANEOUS CORONARY INTERVENTION (PCI). THE TARGET LESION WAS MODERATELY CALCIFIED VESSEL. A 3.00 X 28MM SYNERGY XD AND A 3.50 X 24MM SYNERGY MEGATRON DRUG-ELUTING STENTS WERE ADVANCED FOR TREATMENT. HOWEVER, DURING INSERTION OF BOTH STENTS THROUGH GUIDELINER, THE MID SHAFT OF BOTH SYNERGY MEGATRON STENT BROKE. THE WHOLE DEVICES WERE PULLED AND REMOVED FROM THE PATIENT. A NON-BSC DEVICES WERE USED TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS WERE REPORTED. IT WAS FURTHER REPORTED THAT A 6 FRENCH GUIDE CATHETER WITH A 6 FRENCH GUIDELINER WERE USED DURING INSERTION OF THE STENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16229833 ↗