Skip to content

SYNERGY MEGATRON™ H7493942828350Recalls & reported problems

Boston Scientific Corporation · Everolimus-Eluting Platinum Chromium Coronary Stent System · Class 3 · Implant

All models of SYNERGY MEGATRON (28)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08714729985730), received between Apr 18, 2023 and Jul 11, 2025 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Adverse Event Without Identified Device or Use Problem1
  • Device Damaged by Another Device1
  • Failure to Advance1
  • Fracture1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 11, 2025Fracture; Failure to Advance; Fracture; Failure to Advance

    IT WAS REPORTED THAT STENT FRACTURE OCCURRED. THE 50-60% STENOSED TARGET LESION WAS LOCATED IN A MODERATELY TORTUOUS AND SEVERELY CALCIFIED CORONARY ARTERY. A 3.50 X 28MM SYNERGY MEGATRON DRUG-ELUTING STENT WAS ADVANCED FOR TREATMENT. HOWEVER, A STENT STRUT BROKE WHEN TRYING TO CROSS THE CALCIFIED LESION. THE PROCEDURE WAS COMPLETED USING AN ALTERNATIVE DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22486844 ↗
  • MalfunctionMay 9, 2024Break; Material Deformation; Break; Material Deformation

    IT WAS REPORTED THAT SHAFT BREAK OCCURRED. THE TARGET LESION WAS LOCATED IN CORONARY ARTERY. A 3.50 X 28MM SYNERGY MEGATRON DRUG-ELUTING STENT WAS ADVANCED FOR TREATMENT. HOWEVER, DURING THE PROCEDURE, THE STENT WAS DEFORMED. WHEN THE DEVICE WAS REMOVED, THE MID SHAFT BROKE. EVERYTHING CAME OUT WITH NO ISSUES AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19279333 ↗
  • InjuryJul 10, 2023Device Damaged by Another Device; Device Damaged by Another Device

    IT WAS REPORTED THAT STENT DAMAGE POST-DEPLOYMENT OCCURRED. THE 80% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED AORTO-OSTIAL RIGHT CORONARY ARTERY (RCA). A 3.50 X 28MM SYNERGY MEGATRON DRUG-ELUTING STENT WAS ADVANCED FOR TREATMENT. HOWEVER, AFTER INITIAL DEPLOYMENT OF THE STENT, INTERACTION WITH THE GUIDE CATHETER CAUSED AXIAL DEFORMATION OF THE STENT. AS A RESULT, SUBSEQUENT POST INFLATION BALLOONING WAS MADE DIFFICULT. A NEW WIRE WAS PLACED OUTSIDE THE FIRST ROW OF MEGATRON STENT STRUTS AND BALLOONS WERE USED IN A "CRUSH" TECHNIQUE TO CREATE AN OPENING. THEN, A NEW STENT WAS DEPLOYED AT THE OSTIUM RCA AND THE PROCEDURE WAS SUCCESSFUL. THE PATIENT THEN UNDERWENT ANGIOGRAM OF THE TARGET LESION LOCATED IN THE NON-TORTUOUS MID TO DISTAL RIGHT CORONARY ARTERY. A 2.50MM X 12MM NC EMERGE WAS ADVANCED AND DEPLOYED FOR TREATMENT. HOWEVER, DURING WITHDRAWAL, THE HYPOTUBE BROKE JUST PROXIMAL TO THE BALLOON AND REMAINED INSIDE THE PATIENT. A NEW STENT WAS PLACED TO PIN THE BALLOON FRAGMENT ONTO ANOTHER STENT TO AVOID MIGRATION. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17290591 ↗
  • MalfunctionApr 18, 2023Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT SHAFT BREAK OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED CORONARY ARTERY. A 3.50 X 28MM SYNERGY MEGATRON WAS ADVANCED FOR TREATMENT. HOWEVER, WHEN INSERTING THE MEGATRON INTO THE GUIDE, THE SHAFT SNAPPED IN HALF AT THE MIDDLE OF THE DELIVERY SYSTEM. THE PIECES WERE REMOVED WITH NO ISSUES AND THE PROCEDURE WAS COMPLETED WITH A NEW DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16765597 ↗

Other Coronary drug-eluting stent devices

All Coronary drug-eluting stent devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page