SYNERGY MEGATRON™ H7493942832450Recalls & reported problems

Boston Scientific Corporation · Everolimus-Eluting Platinum Chromium Coronary Stent System · Class 3 · Implant

All models of SYNERGY MEGATRON (28)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 08714729985792), received between Jun 5, 2022 and Apr 28, 2026 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Difficult to Remove” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Remove4
  • Material Integrity Problem2
  • Device Dislodged or Dislocated1
  • Failure to Deflate1
  • Material Rupture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 28, 2026Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE PATIENT PRESENTED WITH ST-ELEVATION MYOCARDIAL INFARCTION AND UNDERWENT PERCUTANEOUS CORONARY INTERVENTION. THE TARGET LESION WAS LOCATED IN THE MODERATE TO SEVERELY CALCIFIED VESSEL. A 4.50 X 32MM SYNERGY MEGATRON DRUG-ELUTING STENT WAS ADVANCED FOR TREATMENT. HOWEVER, DURING THE INFLATION FOR STENT EXPANSION AT SUB NOMINAL PRESSURE, THE STENT BALLOON BURST. THE DEVICE WAS REMOVED AND THERE WERE NO DEVICE FRAGMENTS LEFT INSIDE THE PATIENT'S BODY. THE STENT WAS RE-PREPPED, AND THE RUPTURED STENT BALLOON WAS NOTED. THE PROCEDURE WAS COMPLETED WITH AN ALTERNATE STENT OF SIMILAR DIAMETER AND LENGTH. THERE WERE NO PATIENT COMPLICATIONS AND THE PATIENT FULLY RECOVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25023712 ↗
  • MalfunctionAug 7, 2025Failure to Deflate; Failure to Deflate

    IT WAS REPORTED THAT DEFLATION ISSUE OCCURRED. THE PATIENT PRESENTED WITH A CORONARY ARTERY DISEASE. A 4.50 X 32MM SYNERGY MEGATRON DRUG-ELUTING STENT WAS ADVANCED AND DEPLOYED. POST DEPLOYMENT, THE STENT DELIVERY SYSTEM BALLOON WOULD NOT DEFLATE. THE STENT BALLOON WAS REMOVED AND REPLACED; AND THE PROCEDURE WAS COMPLETED. THERE WERE NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22745695 ↗
  • MalfunctionJul 5, 2024Difficult to Remove; Difficult to Remove

    IT WAS REPORTED VIA USER MEDWATCH MW5155116 THAT CATHETER DIFFICULTY REMOVAL OCCURRED. A 4.50 X 32 MM SYNERGY MEGATRON DRUG-ELUTING STENT WAS DEPLOYED SUCCESSFULLY. HOWEVER, DURING THE REMOVAL OF THE STENT DELIVERY SYSTEM, IT BECAME DIFFICULT TO REMOVE FROM THE PATIENT'S BODY. THE PROCEDURE WAS COMPLETED WITH THE ORIGINAL DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19681399 ↗
  • MalfunctionJul 3, 2024Difficult to Remove; Difficult to Remove

    IT WAS REPORTED VIA USER MEDWATCH MW5155117 THAT CATHETER REMOVAL DIFFICULTY OCCURRED. THE PATIENT PRESENTED WITH A ST-SEGMENT ELEVATION MYOCARDIAL INFARCTION (STEMI). THE 70% STENOSED TARGET LESION WAS LOCATED IN THE MINIMALLY TORTUOUS AND NON-CALCIFIED RIGHT CORONARY ARTERY. THROMBUS WAS NOTED AND A 4.50 X 32MM SYNERGY MEGATRON DRUG-ELUTING STENT WAS DEPLOYED FOR 30 SECONDS AT 20 ATMOSPHERES. HOWEVER, UPON DEFLATION, THE STENT BALLOON BECAME DIFFICULT TO REMOVE. THE PHYSICIAN THEN PULLED BACK NEGATIVELY TWICE, DEFLATED THE BALLOON FOR 69 SECONDS, AND SUCCESSFULLY REMOVED IT. THE STENT BALLOON DID NOT PULL THROUGH THE GUIDE AND EVERYTHING HAD TO BE REMOVED. THE PROCEDURE WAS COMPLETED WITHOUT AFFECTING PATIENT CARE BUT DID DELAY FURTHER NECESSARY STENTING. THE STENT ITSELF WAS DEPLOYED WITHOUT ISSUE. THE PATIENT RETURNED TO THE FLOOR WITH TIMI 3 FLOW. NO OTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19671773 ↗
  • MalfunctionFeb 7, 2023Difficult to Remove; Material Integrity Problem; Difficult to Remove; Material Integrity Problem

    IT WAS REPORTED THAT REMOVAL DIFFICULTIES WERE ENCOUNTERED. THE PATIENT PRESENTED FOR PERCUTANEOUS CORONARY INTERVENTION IN A SAPHENOUS VEIN GRAFT. AFTER INSERTING A 6F SHEATH AND PLACING A 6F GUIDE CATHETER, A 4.50 X 32MM SYNERGY MEGATRON DRUG-ELUTING STENT WAS ADVANCED AND DEPLOYED. HOWEVER, DURING WITHDRAWAL, THE STENT BALLOON WOULD NOT REWRAP TIGHT ENOUGH TO FIT THROUGH THE GUIDE CATHETER OPENING. AFTER A STRUGGLE TO GET IT THROUGH THE GUIDE CATHETER, IT WOULD THEN NOT FIT THROUGH THE SHEATH. THE PHYSICIAN "SNAPPED" THE BALLOON THROUGH THE SHEATH AND WAS THEN ABLE TO REMOVE IT. THE SAME SCENARIO HAPPENED WITH THE SECOND 4.50 X 32MM SYNERGY MEGATRON DRUG-ELUTING STENT DEPLOYED. THE PHYSICIAN NEEDED TO DEPLOY A THIRD SYNERGY MEGATRON DRUG-ELUTING STENT AND UPSIZED TO 8F. THE PROCEDURE WAS COMPLETED WITH NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16313725 ↗
  • MalfunctionFeb 7, 2023Difficult to Remove; Material Integrity Problem; Difficult to Remove; Material Integrity Problem

    IT WAS REPORTED THAT REMOVAL DIFFICULTIES WERE ENCOUNTERED. THE PATIENT PRESENTED FOR PERCUTANEOUS CORONARY INTERVENTION IN A SAPHENOUS VEIN GRAFT. AFTER INSERTING A 6F SHEATH AND PLACING A 6F GUIDE CATHETER, A 4.50 X 32MM SYNERGY MEGATRON DRUG-ELUTING STENT WAS ADVANCED AND DEPLOYED. HOWEVER, DURING WITHDRAWAL, THE STENT BALLOON WOULD NOT REWRAP TIGHT ENOUGH TO FIT THROUGH THE GUIDE CATHETER OPENING. AFTER A STRUGGLE TO GET IT THROUGH THE GUIDE CATHETER, IT WOULD THEN NOT FIT THROUGH THE SHEATH. THE PHYSICIAN "SNAPPED" THE BALLOON THROUGH THE SHEATH AND WAS THEN ABLE TO REMOVE IT. THE SAME SCENARIO HAPPENED WITH THE SECOND 4.50 X 32MM SYNERGY MEGATRON DRUG-ELUTING STENT DEPLOYED. THE PHYSICIAN NEEDED TO DEPLOY A THIRD SYNERGY MEGATRON DRUG-ELUTING STENT AND UPSIZED TO 8F. THE PROCEDURE WAS COMPLETED WITH NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16313723 ↗
  • MalfunctionJun 5, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT A STENT DISLODGED. A PERCUTANEOUS CORONARY INTERVENTION (PCI) WAS PERFORMED ON A TORTUOUS CIRCUMFLEX (CX). A GUIDEZILLA, A 3.00MM X 40MM APEX BALLOON, AND A 4.50 X 32MM SYNERGY MEGATRON STENT WERE SELECTED FOR TREATMENT. HOWEVER, THE STENT CAME OFF THE BALLOON, AND THE BALLOON WAS NOTED TO BE SHEARED. THE DEVICES WERE REMOVED FROM THE PATIENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS RESULTED IN RELATION TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14599684 ↗

Other Coronary drug-eluting stent devices

All Coronary drug-eluting stent devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page