Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729992721), received between Dec 11, 2023 and Mar 27, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Gas/Air Leak” (2 reports).

7 other recalls cover devices cleared under the same FDA submission (P220032); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Gas/Air Leak2
  • Defective Device1
  • Infusion or Flow Problem1
  • Temperature Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 27, 2024Defective Device; Gas/Air Leak; Infusion or Flow Problem; Temperature Problem; Defective Device; Gas/Air Leak; Infusion or Flow Problem; Temperature Problem

    DURING A ATRIAL FIBRILLATION (A FIB) PROCEDURE A SMARTFREEZE CRYO CONSOLE WAS SELECTED FOR USE. AFTER PERFORMING THREE SUCCESSFUL ABLATIONS ON THE LEFT SIDE, THEN TO THE RIGHT SUPERIOR PULMONARY VEIN (RSPV) WITHIN THE FIRST FREEZE A LOW TEMPERATURE ERROR OCCURRED. THE TEMPERATURE REACHED -72 WITHIN 20 SECONDS. THE PHYSICIAN MUST HAVE BEEN TOO DEEP IN THE VEIN AND WAS NOT USING FLUORO AND GOING OFF OF WAVEFORMS. THE LOW TEMPERATURE ERROR WAS CLEARED AND NOW WAITING FOR BALLOON TO THAW. DURING THE THAWING PROCESS A REFRIGERANT FLOW OBSTRUCTION ERROR OCCURRED. TROUBLESHOOTING INCLUDES A COMBINATION OF TURNING OFF THE VACUUM, ROTATING THE CRYO CABLE 180 DEGREES, RECONNECTING THE CABLE, AND REINITIATING THE VACUUM. THE ERROR CODE COULD NOT BE CLEARED. ADDITIONAL TROUBLESHOOTING INCLUDES REPLACING THE CRYO CABLE AND REINITIATING THE VACUUM, STILL THE ERROR CODE PERSISTED. THEN A HARDWARE PROBLEM ERROR MESSAGE OCCURRED. DURING THIS TIME OF TROUBLESHOOTING, TECHNICAL SUPPORT WAS CALLED FOR ADDITIONAL ASSISTANCE. AFTER WAITING 5 MINUTES THE VACUUM WAS REINITIATED AGAIN AND THE HARDWARE ERROR PERSISTED. A NEW BALLOON AND CRYO CABLE WERE ALSO OPENED DURING THIS TIME AS A METHOD OF TROUBLESHOO

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18987714 ↗
  • MalfunctionDec 11, 2023Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED THAT DURING ONGOING USE THERE WAS A LEAK FROM THE CONSOLE. THE SITE CONTACTED TECH SERVICES TO NOTIFY THEM THAT THE SMARTFREEZE CONSOLE WAS LEAKING. THIS WAS NOT DURING A PROCEDURE AND THERE WAS NO PATIENT INVOLVEMENT, NO INJURY TO ANY OTHER PERSONAL WAS REPORTED. THE ISSUE COULD NOT BE TROUBLESHOOTED OVER THE PHONE SO A FIELD SERVICE ENGINEER WENT TO THE SITE. DAMAGE TO THE HOSE THAT RUNS BETWEEN THE N2O TANK AND THE PROCESSING PLATE WAS IDENTIFIED. REPLACEMENT OF THE HOSE WAS REQUIRED. THE CONSOLE IS NOT EXPECTED TO BE RETURNED AS IT WAS SERVICED ON SITE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18301132 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P220032. One submission often covers many models, so check the product and lot details in each record.

  • Z-0301-2025 · Oct 10, 2024 · Open, Classified
    Boston Scientific POLARx FIT BALLOON CATHETER LT OUS, Material Number M004CRBS2110
  • Z-0303-2025 · Oct 10, 2024 · Open, Classified
    Boston Scientific SMARTFREEZE CRYO CONSOLE: Console: Material Number M004CRBS4000; CONSOLE REFURB: Material Number M004CRBS400R0; CONSOLE HOSPITAL, Material Number M004CRBS400H
  • Z-0298-2025 · Oct 10, 2024 · Open, Classified
    Boston Scientific POLARx BALLOON CATHETER LT 28MM OUS, Material Number M004CRBS2100
  • Z-0299-2025 · Oct 10, 2024 · Open, Classified
    Boston Scientific POLARx FIT BALLOON CATHETER ST OUS, Material Number M004CRBS2010
  • Z-0297-2025 · Oct 10, 2024 · Open, Classified
    Boston Scientific POLARx BALLOON CATHETER ST 28MM OUS, Material Number M004CRBS2000.
  • Z-0300-2025 · Oct 10, 2024 · Open, Classified
    Boston Scientific POLARx FIT BALLOON CATHETER ST US, Material Number M004CRBS2060
  • Z-0302-2025 · Oct 10, 2024 · Open, Classified
    Boston Scientific POLARx FIT BALLOON CATHETER LT US, Material Number M004CRBS2160

Other Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation devices

All Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page