Skip to content

Flexitron HDR 136149Recalls & reported problems

Nucletron B.V. · Flexitron HDR · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08717213051126), received between Nov 12, 2019 and Sep 27, 2024 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Connection Problem” (1 report).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 18, 2019Terminated

    A component failure may lead to treatment interruption or incorrect source positioning.

    Z-0772-2020 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Connection Problem1
  • Improper or Incorrect Procedure or Method1
  • Manufacturing, Packaging or Shipping Problem1
  • Misassembled1

Patient problems

  • Insufficient Information2
  • No Clinical Signs, Symptoms or Conditions1
  • No Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjurySep 27, 2024Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    IT WAS REPORTED THAT THE USER MISREAD FLEXITRON SPS..

    Patient problems: Insufficient Information

    MAUDE report 20321937 ↗
  • MalfunctionNov 5, 2020Connection Problem; Manufacturing, Packaging or Shipping Problem; Connection Problem; Manufacturing, Packaging or Shipping Problem

    THE CUSTOMER REPORTED THAT THE CHECK CABLE BRACKET TUBE GOT DISCONNECTED FROM THE MOUNTING BLOCK.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 10791012 ↗
  • MalfunctionNov 12, 2019Misassembled; Misassembled

    ONE CUSTOMER COMPLAINT HAS BEEN REPORTED FROM THE FIELD CONCERNING A SYSTEM IN WHICH THE INTERNAL CHECK CABLE GUIDE MECHANISM FAILED. DUE TO AN OBSTRUCTION DURING THE CHECK CABLE OUT DRIVE, THE INTERNAL GUIDING TUBE BECAME DISPLACED AND THE CHECK CABLE BECAME DAMAGED.

    Patient problems: No Information

    MAUDE report 9309326 ↗

Other System, applicator, radionuclide, remote-controlled devices

All System, applicator, radionuclide, remote-controlled devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page