Copilot 1003331Recalls & reported problems
Abbott Vascular Inc. · COPILOT Bleedback Control Valve · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Abbott Copilot bleedback control valve is a small valve used in heart procedures to control blood flow in catheter lines. Many of its 188 reports describe leaks or splashes, loose connections that would not stay attached, and sticking parts. Some describe a damaged seal. There are 3 injury reports and no deaths. No recalls name this device. Reports fell from 38 in the year before to 9 in the latest year.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 189
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 189 adverse event reports that list this exact device (DI 08717648013997), received between Mar 3, 2017 and Sep 21, 2026 — classified as 3 injuries, 186 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (149 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Leak/Splash149
- Loose or Intermittent Connection45
- Physical Resistance/Sticking15
- Difficult to Advance2
- Product Quality Problem2
- Unsealed Device Packaging2
- Contamination /Decontamination Problem1
- Device Operates Differently Than Expected1
Patient problems
- No Clinical Signs, Symptoms or Conditions166
- No Consequences Or Impact To Patient17
- Hemorrhage/Blood Loss/Bleeding5
- Air Embolism3
- No Patient Involvement2
- Bradycardia1
- Cardiac Arrest1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 25 most recent of 189 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 21, 2026Loose or Intermittent Connection; Physical Resistance/Sticking
It was reported that the CoPilot cap was not pushing in and the rotator was not tightening correctly to the sheath. Another device was used to continue the procedure. There were no adverse patient effects. No additional information was provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50841060 ↗MalfunctionJul 22, 2026Loose or Intermittent Connection
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25909754 ↗MalfunctionMay 14, 2026Loose or Intermittent Connection
IT WAS REPORTED THAT DURING PREPARATION, THE SIDEPORT OF A COPILOT BBCV DEVICE WAS UNABLE TO BE SCREWED ONTO THE STOPCOCK. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25174142 ↗MalfunctionApr 9, 2026Difficult to Advance; Material Split, Cut or Torn
IT WAS REPORTED THE PROCEDURE WAS TO IMPLANT AN AMPLATZER AMULET LEFT ATRIAL APPENDAGE OCCLUDER. THE AMULET WAS PUSHED THROUGH THE COPILOT BLEEDBACK CONTROL VALVE (BBCV) HOWEVER RESISTANCE WAS MET AND IT WAS NOTED THE O-RING INSIDE THE COPILOT WAS DISFIGURED. THE AMULET WAS REMOVED HOWEVER A PIECE OF THE O-RING WAS NOTED ON THE AMULET DEVICE. THE COPILOT WAS REPLACED WITH A NEW ONE TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24829712 ↗MalfunctionApr 8, 2026Loose or Intermittent Connection
IT WAS REPORTED THAT DURING PREP AND DURING THE PROCEDURE, THE COPILOT DEVICE WOULD NOT STAY CONNECTED TO THE SYRINGE. WHEN THE TECH TURNED THE SYRINGE ON TO THE SIDE PORT OF THE COPILOT AND LET GO, THE SYRINGE WOULD TWIST OFF AND DROP TO THE TABLE. WHEN CONNECTING THE SIDE PORT TO PRESSURE TUBING FOR INJECTIONS A LOT OF FORCE WAS NEEDED TO MAKE SURE PRESSURE TUBING WOULD STAY CONNECTED. THE SAME DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24818664 ↗MalfunctionFeb 23, 2026Leak/Splash
IT WAS REPORTED THAT DURING USE OF THE COPILOT, BUBBLES WERE NOTED. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24421715 ↗MalfunctionFeb 23, 2026Leak/Splash
IT WAS REPORTED THAT DURING USE OF THE COPILOT, BUBBLES WERE NOTED. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24420638 ↗MalfunctionFeb 12, 2026Loose or Intermittent Connection
IT WAS REPORTED THE COPILOT BLEEDBACK CONTROL VALVE (BBCV) WAS BLOCKED AT THE SCREW THREAD. THIS ISSUE OCCURRED WITH TWO DEVICES. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24324334 ↗MalfunctionFeb 12, 2026Loose or Intermittent Connection
IT WAS REPORTED THE COPILOT BLEEDBACK CONTROL VALVE (BBCV) WAS BLOCKED AT THE SCREW THREAD. THIS ISSUE OCCURRED WITH TWO DEVICES. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24324283 ↗MalfunctionJan 2, 2026Contamination /Decontamination Problem
IT WAS REPORTED THAT THE COPILOT WAS USED IN THE PROCEDURE. WHEN THE PROCEDURE WAS COMPLETED, IMPURITIES WERE NOTED IN THE DEVICE. THERE WERE NO ADVERSE PATIENT EFFECTS AND THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23945737 ↗MalfunctionSep 12, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT THE COPILOT DEVICE LEAKED DURING CONTRAST AGENT INJECTIONS FROM THE BLEEDBACK CONTROL SEAL. THERE WERE NO ADVERSE PATIENT EFFECTS AND THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. ANOTHER COPILOT WAS USED TO COMPLETE THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23042491 ↗InjuryJun 17, 2025Leak/Splash; Leak/Splash
USER FACILITY MEDWATCH REPORT RECEIVED THAT STATES "PATIENT WAS BROUGHT FOR AN ELECTIVE LAA [LEFT ATRIAL APPENDAGE] OCCLUSION DEVICE. TRANSSEPTAL PUNCTURE PERFORMED. TRANSSEPTAL SHEATH EXCHANGED FOR DEVICE SHEATH. DEVICE SHEATH USES A DEVICE CALLED A COPILOT BLEEDBACK CONTROL VALVE. THE COPILOT BLEEDBACK CONTROL VALVE WAS FOUND TO BE DEFECTIVE. WHEN BLOOD WAS DRAWN BACK, THE BLEEDBACK VALVE INTRODUCED AIR TO THE DEVICE SHEATH, WHICH ULTIMATELY DELIVERED AIR TO THE LEFT ATRIUM, WHICH BROUGHT AIR TO THE CORONARY ARTERIES. TROUBLESHOOTING MANEUVERS WERE PERFORMED, AND ST ELEVATION RESOLVED, WITH THE PATIENT EVENTUALLY BEING SENT TO THE ICU [INTENSIVE CARE UNIT]. USED A MANUAL TUOHY." SUBSEQUENT TO THE MEDWATCH REPORT, IT WAS REPORTED THAT ONE DEVICE WAS IN THE COPILOT BBCV WHEN THE LEAK WAS NOTED. OXYGEN WAS ADMINISTERED TO THE PATIENT TO RESOLVE THE EKG CHANGES. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Air Embolism; Non specific EKG/ECG Changes
MAUDE report 22234199 ↗MalfunctionMay 30, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT THE RIGHT CORONARY ARTERY (RCA) WITH MODERATE CALCIFICATION, MODERATE TORTUOSITY AND 90% STENOSIS. THE COPILOT WAS USED IN THE PROCEDURE BUT THERE WAS POOR VALVE CLOSURE AND BLOOD LEAKAGE. A NON-ABBOTT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22118797 ↗MalfunctionApr 22, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A MODERATELY CALCIFIED LESION IN THE LEFT ANTERIOR DESCENDING ARTERY. DURING THE USE OF THE COPILOT BLEEDBACK CONTROL VALVE (BBCV ), CONTRAST WAS NOTED TO BE LEAKING FROM THE CAP OF THE BBC AS WELL AS PRESSURE WAS NOTED TO BE LOST. THEREFORE, THE CAP WAS TIGHTENED AND THE LEAK WAS NOTED TO SUBSIDE, AND THE SAME BBCV WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS NOR CLINICAL DELAY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21883939 ↗MalfunctionApr 14, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT THE COPILOT BLEED BACK CONTROL VALVE (BBCV) WAS BEING USED IN THE DIAGNOSTIC PROCEDURE; HOWEVER, THE VALVE WAS NOTED TO LEAKING SALINE AND BLOOD FROM THE ROTOR OF THE DEVICE. THE PROCEDURE WAS COMPLETED WITH THE DEVICE. THERE WAS NO ADVERSE PATIENT EFFECT AND NO CLINICALLY SIGNIFICANT DELAY REPORTED IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21835342 ↗MalfunctionMar 27, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A RIGHT CORONARY ARTERY. THE COPILOT BLEEDBACK CONTROL VALVE WAS NOTED TO HAVE CONSTANT AIR GETTING INTO SYRINGE WHEN DRAWING BACK ON THE SYRINGE. THE SAME DEVICE WAS USED THROUGHOUT THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21712408 ↗MalfunctionMar 21, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT THE COPILOT VALVE WAS CLOSED AND DID NOT LOCK PROPERLY AND WAS NOTED TO LEAK. THE WIRE DID NOT MOVE OR STAY FIXED. ANOTHER COPILOT WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21667698 ↗MalfunctionMar 21, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT THE COPILOT WAS LEAKING BLOOD FROM THE BLEEDBACK CONTROL VALVE DURING USE. THE DEVICE WAS CONNECTED TO A CATHETER, BALLOON AND WIRE. THE SAME DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21665758 ↗MalfunctionMar 18, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT THE COPILOT WAS LEAKING WHEN CONNECTED TO AN UNSPECIFIED DEVICE AND WHEN THE DEVICE WAS TAKEN OUT. THERE WAS NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21632996 ↗MalfunctionMar 6, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION IN AN UNSPECIFIED CORONARY ARTERY. THE COPILOT BLEED BACK CONTROL VALVE (BBCV) WAS USED IN THE PROCEDURE, HOWEVER, A LEAK WAS OBSERVED AT THE HEMOSTATIC VALVE ALTHOUGH THE SEAL WAS CLOSED. THERE WAS NO ADVERSE PATIENT EFFECT REPORTED AND NO CLINICALLY SIGNIFICANT DELAY REPORTED IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21540134 ↗MalfunctionMar 6, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION IN AN UNSPECIFIED CORONARY ARTERY. THE COPILOT BLEED BACK CONTROL VALVE (BBCV) WAS USED IN THE PROCEDURE, HOWEVER, A LEAK WAS OBSERVED AT THE HEMOSTATIC VALVE ALTHOUGH THE SEAL WAS CLOSED. THERE WAS NO ADVERSE PATIENT EFFECT REPORTED AND NO CLINICALLY SIGNIFICANT DELAY REPORTED IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21539932 ↗MalfunctionFeb 20, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT DURING THE PROCEDURE, THE COPILOT WAS NOTED TO BE LEAKING BLOOD, MAKING IT DIFFICULT FOR THE PHYSICIAN TO MANIPULATE THE DEVICES AND HAVE VISIBILITY ON ALL THE DEVICES. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21420730 ↗MalfunctionJan 30, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT THE COPILOT DEVICE WAS USED DURING THE PROCEDURE. THE PHYSICIAN FOUND THAT BLOOD FLOWED BACK EVEN WHEN IT WAS LOCKED. IT CONTINUED AFTER THE GUIDEWIRE WAS INSERTED, AND THE SITUATION REMAINED EVEN WHEN ADDITIONAL DEVICES WERE INSERTED. A SECOND COPILOT DEVICE WAS USED BUT HAD THE SAME ISSUE. THE PHYSICIAN USED A THIRD NEW COPILOT, BUT IT ALSO DID NOT WORK PROPERLY. THE SAME ISSUE OCCURRED WITH THE THREE COPILOT DEVICES. THE PROCEDURE WAS COMPLETED WITH A FOURTH, NEW COPILOT DEVICE, AND NO PATIENT ADVERSE EVENTS WERE REPORTED. HOWEVER, THE USER WAS VERY UPSET. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21275979 ↗MalfunctionJan 30, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT THE COPILOT DEVICE WAS USED DURING THE PROCEDURE. THE PHYSICIAN FOUND THAT BLOOD FLOWED BACK EVEN WHEN IT WAS LOCKED. IT CONTINUED AFTER THE GUIDEWIRE WAS INSERTED, AND THE SITUATION REMAINED EVEN WHEN ADDITIONAL DEVICES WERE INSERTED. A SECOND COPILOT DEVICE WAS USED BUT HAD THE SAME ISSUE. THE PHYSICIAN USED A THIRD NEW COPILOT, BUT IT ALSO DID NOT WORK PROPERLY. THE SAME ISSUE OCCURRED WITH THE THREE COPILOT DEVICES. THE PROCEDURE WAS COMPLETED WITH A FOURTH, NEW COPILOT DEVICE, AND NO PATIENT ADVERSE EVENTS WERE REPORTED. HOWEVER, THE USER WAS VERY UPSET. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21275930 ↗MalfunctionJan 30, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT THE COPILOT DEVICE WAS USED DURING THE PROCEDURE. THE PHYSICIAN FOUND THAT BLOOD FLOWED BACK EVEN WHEN IT WAS LOCKED. IT CONTINUED AFTER THE GUIDEWIRE WAS INSERTED, AND THE SITUATION REMAINED EVEN WHEN ADDITIONAL DEVICES WERE INSERTED. A SECOND COPILOT DEVICE WAS USED BUT HAD THE SAME ISSUE. THE PHYSICIAN USED A THIRD NEW COPILOT, BUT IT ALSO DID NOT WORK PROPERLY. THE SAME ISSUE OCCURRED WITH THE THREE COPILOT DEVICES. THE PROCEDURE WAS COMPLETED WITH A FOURTH, NEW COPILOT DEVICE, AND NO PATIENT ADVERSE EVENTS WERE REPORTED. HOWEVER, THE USER WAS VERY UPSET. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21275927 ↗
FDA clearance and classification
The FDA regulates product code DTL (adaptor, stopcock, manifold, fitting, cardiopulmonary bypass) under 21 CFR 870.4290 as a Class II device. 21 CFR 870.4290 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Copilot 1003331
Has the Copilot 1003331 been recalled?
No FDA recall record in our data names this exact device identifier (DI 08717648013997), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Copilot 1003331?
The FDA's MAUDE database lists 189 adverse event reports for this device, received between Mar 3, 2017 and Sep 21, 2026. In those reports, the most common specific device problems are “Leak/Splash” (149), “Loose or Intermittent Connection” (45) and “Physical Resistance/Sticking” (15); the most common patient problems are “Hemorrhage/Blood Loss/Bleeding” (5), “Air Embolism” (3) and “Bradycardia” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Copilot 1003331?
None of the reports are classified as deaths, 3 as injuries and 186 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Copilot 1003331 FDA approved?
Devices of this type (product code DTL) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Abbott Vascular Inc.
Who makes the Copilot 1003331, and is it still sold?
The FDA's device database lists Abbott Vascular Inc. as the company responsible for it (brand Copilot, model 1003331). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Adaptor, stopcock, manifold, fitting, cardiopulmonary bypass devices
- Indeflator 1003327Abbott Vascular Inc. · Adaptor, stopcock, manifold, fitting, cardiopulmonary bypass2 recalls722 reports
- DLP® 11001GMedtronic, Inc. · Adaptor, stopcock, manifold, fitting, cardiopulmonary bypass2 recalls0 reports
- RHV 23242Abbott Vascular Inc. · Adaptor, stopcock, manifold, fitting, cardiopulmonary bypass72 reports
- DLP® 14000Medtronic, Inc. · Adaptor, stopcock, manifold, fitting, cardiopulmonary bypass1 recall2 reports
- PhD™ 00884450746745Merit Medical Systems, Inc. · Adaptor, stopcock, manifold, fitting, cardiopulmonary bypass1 recall2 reports
- PhD™ 00884450746714Merit Medical Systems, Inc. · Adaptor, stopcock, manifold, fitting, cardiopulmonary bypass1 recall1 report
- NAMIC 64038301Medline Industries, Inc. · Adaptor, stopcock, manifold, fitting, cardiopulmonary bypass1 recall1 report
- NAMIC 64038402Medline Industries, Inc. · Adaptor, stopcock, manifold, fitting, cardiopulmonary bypass1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page