MULTI-LINK ULTRA 1003381-13Recalls & reported problems

Abbott Vascular Inc. · MULTI-LINK RX ULTRA Coronary Stent System 4.50 mm x 13 mm · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08717648035265), received between Jan 27, 2017 and Jan 27, 2017 — classified as 1 injury.

The most frequently reported specific device problem is “Detachment Of Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Detachment Of Device Component1
  • Device Dislodged or Dislocated1
  • Failure to Advance1
  • Improper or Incorrect Procedure or Method1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 27, 2017Detachment Of Device Component; Improper or Incorrect Procedure or Method; Failure to Advance; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION LOCATED IN THE RIGHT CORONARY ARTERY (RCA). THE 4.5 X 13 MM ULTRA STENT DELIVERY SYSTEM (SDS) WAS ADVANCED; HOWEVER, WOULD NOT CROSS THE LESION. THE SDS WAS RETRACTED AND A 3.5 X 8 MM BALLOON CATHETER WAS THEN USED TO DILATE THE RCA. UPON EXAMINATION OF THE SDS IT WAS DETERMINED THAT THERE WAS NO DAMAGE TO THE 4.5 X 13 MM ULTRA SDS; THEREFORE, IT WAS REINSERTED; HOWEVER, AGAIN IT WOULD NOT CROSS THE LESION. UPON RETRACTION OF THE SDS A SECOND TIME IT WAS OBSERVED THAT THE STENT IMPLANT HAD DISLODGED AND REMAINED IN THE RCA. THERE WAS NO RESISTANCE FELT DURING REMOVAL OF THE SDS. AFTER NUMEROUS ATTEMPTS TO SNARE THE STENT FAILED, A 1.2 MM BALLOON CATHETER WAS INSERTED INTO THE STENT AND INFLATED ENOUGH TO GRAB THE STENT, WHICH THEN WAS ABLE TO BE BROUGHT DOWN ALMOST TO THE FEMORAL ARTERY INSERTION SITE. AT THIS TIME OPEN HEART SURGEONS WERE CALLED TO DO A GROIN EXPLORATION TO REMOVE THE STENT FROM THE BODY. THEY REMOVED THE STENT PIECE BY PIECE FROM THE PATIENT. THE PATIENT HAS RECOVERED WITH NO FURTHER COMPLICATIONS. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6284931 ↗

Other STENT, CORONARY devices

All STENT, CORONARY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page