MULTI-LINK MINI VISION 1007821-08Recalls & reported problems
Abbott Vascular Inc. · MULTI-LINK MINI VISION Coronary Stent and Stent Delivery System 2.00 mm x 08 mm / Rapid-Exchange · Class 3 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08717648049378), received between Nov 8, 2016 and Aug 23, 2019 — classified as 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Failure to Advance” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Failure to Advance3
- Device Dislodged or Dislocated2
- Adverse Event Without Identified Device or Use Problem1
- Detachment Of Device Component1
- Difficult to Remove1
- Kinked1
Patient problems
- No Consequences Or Impact To Patient2
- Angina1
- High Blood Pressure/ Hypertension1
- Intimal Dissection1
- Occlusion1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryAug 23, 2019Difficult to Remove; Failure to Advance; Device Dislodged or Dislocated; Difficult to Remove; Failure to Advance; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A LESION IN THE MID RIGHT CORONARY ARTERY WITH HEAVY CALCIFICATION AND HEAVY TORTUOSITY. A 2.0 X 08 MM MINI VISION STENT DELIVERY SYSTEM (SDS) WAS ADVANCED; HOWEVER FAILED TO CROSS THE LESION DUE TO THE PATIENT ANATOMY. ON REMOVAL OF THE SDS WITHOUT REPORTED FORCE, RESISTANCE WITH THE LESION WAS MET AND THE STENT DISLODGED. MEDICAL INTERVENTION WAS PERFORMED TO REMOVE THE STENT AND A DISSECTION WAS NOTED. AN ADDITIONAL UNSPECIFIED STENT WAS USED TO TREAT THE DISSECTION. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Intimal Dissection
MAUDE report 8927751 ↗InjuryMar 23, 2018Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT IN 2016, A MINI VISION STENT WAS IMPLANTED IN THE CORONARY ARTERY. ON (B)(6) 2018, THE PATIENT WAS HOSPITALIZED FOR CHEST PAIN AND ELEVATED BLOOD SUGARS. ON (B)(6) 2018, AN ANGIOGRAM WAS PERFORMED. THE PREVIOUSLY IMPLANTED STENT WAS TREATED AND ANOTHER STENT WAS IMPLANTED. THE EVENT RESOLVED AND THE PATIENT WAS DISCHARGED. THERE WAS NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: Angina; High Blood Pressure/ Hypertension; Occlusion
MAUDE report 7367928 ↗MalfunctionJun 14, 2017Detachment Of Device Component; Kinked; Failure to Advance
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A VERY HEAVILY CALCIFIED LESION IN THE VERY TORTUOUS MID DIAGONAL CORONARY ARTERY. A 2.0X8MM MINI VISION RX BARE METAL STENT (BMS) SYSTEM WAS ADVANCED, BUT FAILED TO CROSS THE LESION DUE TO PATIENT ANATOMY AND, THOUGH NO FORCE WAS APPLIED, THE PROXIMAL SHAFT KINKED. THE STENT SYSTEM WAS WITHDRAWN WITHOUT RESISTANCE, BUT THE SHAFT SEPARATED (SNAPPED INTO TWO PIECES) AT THE KINK AREA DURING WITHDRAWAL. THE LESION WAS THEN DILATED WITH A 2.0X8MM NON-ABBOTT BALLOON FOLLOWED BY ADDITIONAL ATTEMPTS TO CROSS THE LESION WITH OTHER UNSPECIFIED STENT SYSTEM; NO OTHER STENT SYSTEMS WERE ABLE TO CROSS THE LESION DUE TO PATIENT ANATOMY, SO THE PROCEDURE WAS ABORTED. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO OCCURRENCE OF A CLINICALLY SIGNIFICANT DELAY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6640767 ↗InjuryNov 8, 2016Failure to Advance; Device Dislodged or Dislocated
IT WAS REPORTED THAT DURING A PROCEDURE OF THE LEFT ANTERIOR DESCENDING (LAD), DUE TO THE ACUTE ANGLE OF THE VESSEL, THE 2.0 X 8 MM MINI-VISION STENT BECAME DISLODGED FROM THE BALLOON AND REMAINED ON THE GUIDE WIRE. A 1.5 X 8 MM BALLOON WAS USED TO RETRIEVE AND REMOVE THE STENT FROM THE ANATOMY. THE PROCEDURE WAS COMPLETED USING BALLOON ANGIOPLASTY. THERE WAS NO REPORTED ADVERSE PATIENT EFFECT OR A CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6087404 ↗
Other STENT, CORONARY devices
- WALLSTENT® Endoprosthesis H965403130Boston Scientific Corporation · STENT, CORONARY35 reports
- WALLSTENT® Endoprosthesis H965402130Boston Scientific Corporation · STENT, CORONARY23 reports
- Tryton Side Branch Stent T5-2530-191-USTryton Medical, Inc. · Stent, Coronary1 death22 reports
- WALLSTENT® Endoprosthesis H965403330Boston Scientific Corporation · STENT, CORONARY21 reports
- WALLSTENT® Endoprosthesis H965403320Boston Scientific Corporation · STENT, CORONARY10 reports
- WALLSTENT® Endoprosthesis H965403120Boston Scientific Corporation · STENT, CORONARY9 reports
- Tryton Side Branch Stent T5-3035-151-USTryton Medical, Inc. · Stent, Coronary2 deaths7 reports
- MULTI-LINK VISION 1007850-15Abbott Vascular Inc. · STENT, CORONARY7 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page