Skip to content

MULTI-LINK VISION 1007849-12Recalls & reported problems

Abbott Vascular Inc. · MULTI-LINK RX VISION Coronary Stent System 3.50 mm x 12 mm · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08717648050695), received between Feb 8, 2018 and Apr 2, 2021 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Device Dislodged or Dislocated2
  • Improper or Incorrect Procedure or Method1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Patient Involvement1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 2, 2021Improper or Incorrect Procedure or Method; Device Dislodged or Dislocated; Improper or Incorrect Procedure or Method; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LEFT ANTERIOR DESCENDING ARTERY. A 3.5X12MM RX VISION STENT WAS PREPPED AND THEN INSERTED INTO THE PATIENT. ONCE AT THE TARGET LESION, THE BALLOON WAS INFLATED AND IT WAS NOTED THERE WAS NO STENT ON THE BALLOON. THE ENTIRE CATH LAB SEARCHED BUT COULD NOT FIND THE STENT. A SCAN WAS PERFORMED OF THE PATIENT AND THE STENT WAS NOT FOUND INSIDE THE PATIENT. IT IS ASSUMED IT DISLODGED DURING PREPARATION AS THE STENT WAS NOT FOUND IN THE BODY. ANOTHER SAME SIZE STENT WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11609453 ↗
  • MalfunctionFeb 8, 2018Device Dislodged or Dislocated

    IT WAS REPORTED THAT DURING UNPACKAGING, AFTER THE 3.5 X 12 MM VISION STENT DELIVERY SYSTEM (SDS) WAS REMOVED FROM THE HOOP, THE NEW SCRUB TECH INADVERTENTLY PULLED THE STENT OFF THE SDS BALLOON WHEN REMOVING THE STYLET. THE SDS WAS NOT USED IN THE ANATOMY AND WAS REPLACED WITH A NEW VISION SDS. THERE WAS NO PATIENT INVOLVEMENT AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Patient Involvement

    MAUDE report 7256073 ↗