The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08717648050701), received between Jan 4, 2018 and Feb 6, 2020 — classified as 1 death, 3 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Device Dislodged or Dislocated2
- Difficult to Remove2
- Adverse Event Without Identified Device or Use Problem1
- Deflation Problem1
- Improper or Incorrect Procedure or Method1
- Physical Resistance1
- Torn Material1
Patient problems
- No Patient Involvement2
- Cardiac Arrest1
- No Consequences Or Impact To Patient1
- Occlusion1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 6, 2020Difficult to Remove; Device Dislodged or Dislocated; Difficult to Remove; Device Dislodged or Dislocated
IT WAS REPORTED THAT DURING DEVICE PREPARATION, RESISTANCE WAS MET DURING REMOVAL OF THE PROTECTIVE SHEATH AND ONCE REMOVED, IT WAS NOTED THAT THE STENT WAS DISLODGED. THERE WAS NO PATIENT INVOLVEMENT AND NO CLINICALLY SIGNIFICANT DELAY IN PROCEDURE. ANOTHER VISION STENT WAS IMPLANTED TO COMPLETE THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Patient Involvement
MAUDE report 9679431 ↗DeathAug 28, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT AN ACUTE MYOCARDIAL INFARCTION. FOLLOWING THE USE OF A NON-ABBOTT HEART PUMP, A 3.50X15 MM RX VISION STENT WAS SUCCESSFULLY DEPLOYED IN THE PROXIMAL LEFT ANTERIOR DESCENDING CORONARY ARTERY. POST DEPLOYMENT OF THE STENT, COMPLETE OCCLUSION OF THE STENT OCCURRED ALMOST IMMEDIATELY. THE PATIENT CODED. DESPITE CHEST COMPRESSIONS BEING PERFORMED, THE PATIENT EXPIRED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Cardiac Arrest; Occlusion
MAUDE report 7824796 ↗MalfunctionMay 1, 2018Deflation Problem; Difficult to Remove; Improper or Incorrect Procedure or Method; Physical Resistance; Torn Material
IT WAS REPORTED THAT THE 3.5X15 MM VISION STENT DELIVERY SYSTEM (SDS) WAS INSERTED FOR DIRECT STENTING IN THE HEAVILY CALCIFIED, MID RIGHT CORONARY ARTERY (RCA), BUT WAS UNABLE TO ADVANCE. PREDILATATION WAS PERFORMED AND A 6 FRENCH GUIDELINER WAS INSERTED, WHICH ALLOWED THE VISION SDS TO REACH THE VESSEL. THE STENT WAS SUCCESSFULLY DEPLOYED AT RATED BURST PRESSURE. (B)(6) PRESSURE WAS ONLY HELD FOR A FEW SECONDS ON THE SDS WHEN THE PARTIALLY DEFLATED SDS BALLOON WAS RETRACTED FROM THE RCA AND MET RESISTANCE WITH THE GUIDELINER. THE SDS BALLOON BECAME BUNCHED UP AND WAS REMOVED WITH THE GUIDELINER AS A UNIT. AFTER REMOVAL FROM THE ANATOMY, A TEAR WAS THOUGHT TO BE SEEN ON THE SDS BALLOON. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. THERE WAS NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7476920 ↗MalfunctionJan 4, 2018Device Dislodged or Dislocated
IT WAS REPORTED THAT DURING PREPARATION OF THE 3.5 X 15 MM RX VISION STENT DELIVERY SYSTEM (SDS), WHEN REMOVING THE PROTECTIVE SHEATH, THE STENT IMPLANT DISLODGED FROM THE BALLOON. THERE WAS NO REPORT OF RESISTANCE FELT DURING REMOVAL OF THE SHEATH. THE DEVICE COULD NOT BE USED ON THE PATIENT. A NEW SDS WAS USED TO COMPLETE THE CASE SUCCESSFULLY. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Patient Involvement
MAUDE report 7164988 ↗
Other STENT, CORONARY devices
- WALLSTENT® Endoprosthesis H965403130Boston Scientific Corporation · STENT, CORONARY35 reports
- WALLSTENT® Endoprosthesis H965402130Boston Scientific Corporation · STENT, CORONARY23 reports
- Tryton Side Branch Stent T5-2530-191-USTryton Medical, Inc. · Stent, Coronary1 death22 reports
- WALLSTENT® Endoprosthesis H965403330Boston Scientific Corporation · STENT, CORONARY21 reports
- WALLSTENT® Endoprosthesis H965403320Boston Scientific Corporation · STENT, CORONARY10 reports
- WALLSTENT® Endoprosthesis H965403120Boston Scientific Corporation · STENT, CORONARY9 reports
- PALMAZ Balloon Expandable Stent P308Cordis US Corp. · STENT, CORONARY7 reports
- Cobra PzF 170-03-35024Celonova Biosciences, Inc. · Stent, Coronary7 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page