MULTI-LINK VISION 1007849-23Recalls & reported problems
Abbott Vascular Inc. · MULTI-LINK RX VISION Coronary Stent System 3.50 mm x 23 mm · Class 3 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08717648050725), received between May 2, 2018 and Mar 22, 2021 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Detachment Of Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Detachment Of Device Component2
- Difficult to Remove2
- Component Missing1
- Failure to Advance1
Patient problems
- No Consequences Or Impact To Patient2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 22, 2021Component Missing; Component Missing
IT WAS REPORTED THAT THE 3.5X23 MM VISION STENT WAS NOT IN THE BOX. THERE WAS NO REPORTED DEVICE USE OR PATIENT INVOLVEMENT. TWO 3.0X23 MM VISION STENTS WERE USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11544532 ↗MalfunctionMay 2, 2018Detachment Of Device Component; Difficult to Remove
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A 95% STENOSED, MILDLY TORTUOUS AND MODERATELY CALCIFIED LESION IN THE PROXIMAL LEFT ANTERIOR DESCENDING ARTERY. FOLLOWING PRE-DILATATION, A 3.5X23MM RX VISION STENT DELIVERY SYSTEM (SDS) FAILED TO CROSS DUE TO THE ANATOMY AND THE PROXIMAL SHAFT SEPARATED. RESISTANCE REMOVING THE SDS WAS FELT. A SECOND 3.5X23MM RX VISION SDS WAS ADVANCED AND THE STENT WAS DEPLOYED WITHOUT ISSUES. UPON REMOVAL, THE SDS BECAME FROZEN ON THE ALLSTAR GUIDE WIRE AND THE PROXIMAL SHAFT SEPARATED. AFTER THE SEPARATION, THE SDS COULD BE REMOVED. A 3.0X18MM RX VISION SDS WAS BEING ADVANCED; HOWEVER, RESISTANCE ADVANCING THE SDS OVER THE GUIDE WIRE OUTSIDE OF THE PATIENT ANATOMY WAS FELT. THE DECISION WAS MADE NOT TO USE IT AND DURING REMOVAL, RESISTANCE WAS ALSO MET WITH THE GUIDE WIRE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A NEW 3.0X18MM RX VISION STENT AND THE SAME ALLSTAR GUIDE WIRE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7480614 ↗MalfunctionMay 2, 2018Detachment Of Device Component; Difficult to Remove; Failure to Advance
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A 95% STENOSED, MILDLY TORTUOUS AND MODERATELY CALCIFIED LESION IN THE PROXIMAL LEFT ANTERIOR DESCENDING ARTERY. FOLLOWING PRE-DILATATION, A 3.5X23MM RX VISION STENT DELIVERY SYSTEM (SDS) FAILED TO CROSS DUE TO THE ANATOMY AND THE PROXIMAL SHAFT SEPARATED. RESISTANCE REMOVING THE SDS WAS FELT. A SECOND 3.5X23MM RX VISION SDS WAS ADVANCED AND THE STENT WAS DEPLOYED WITHOUT ISSUES. UPON REMOVAL, THE SDS BECAME FROZEN ON THE ALLSTAR GUIDE WIRE AND THE PROXIMAL SHAFT SEPARATED. AFTER THE SEPARATION, THE SDS COULD BE REMOVED. A 3.0X18MM RX VISION SDS WAS BEING ADVANCED; HOWEVER, RESISTANCE ADVANCING THE SDS OVER THE GUIDE WIRE OUTSIDE OF THE PATIENT ANATOMY WAS FELT. THE DECISION WAS MADE NOT TO USE IT AND DURING REMOVAL, RESISTANCE WAS ALSO MET WITH THE GUIDE WIRE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A NEW 3.0X18MM RX VISION STENT AND THE SAME ALLSTAR GUIDE WIRE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7480585 ↗
Other STENT, CORONARY devices
- WALLSTENT® Endoprosthesis H965403130Boston Scientific Corporation · STENT, CORONARY35 reports
- WALLSTENT® Endoprosthesis H965402130Boston Scientific Corporation · STENT, CORONARY23 reports
- Tryton Side Branch Stent T5-2530-191-USTryton Medical, Inc. · Stent, Coronary1 death22 reports
- WALLSTENT® Endoprosthesis H965403330Boston Scientific Corporation · STENT, CORONARY21 reports
- WALLSTENT® Endoprosthesis H965403320Boston Scientific Corporation · STENT, CORONARY10 reports
- WALLSTENT® Endoprosthesis H965403120Boston Scientific Corporation · STENT, CORONARY9 reports
- Tryton Side Branch Stent T5-3035-151-USTryton Medical, Inc. · Stent, Coronary2 deaths7 reports
- MULTI-LINK VISION 1007850-15Abbott Vascular Inc. · STENT, CORONARY7 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page