Skip to content

MULTI-LINK VISION 1007850-28Recalls & reported problems

Abbott Vascular Inc. · MULTI-LINK RX VISION Coronary Stent System 4.00 mm x 28 mm · Class 3 · Implant

All models of MULTI-LINK VISION (19)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08717648050817), received between Jun 6, 2017 and Jun 6, 2017 — classified as 1 malfunction.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Device Dislodged or Dislocated1
  • Improper or Incorrect Procedure or Method1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 6, 2017Improper or Incorrect Procedure or Method; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PROCEDURE WAS TO LESION IN THE SAPHENOUS VEIN GRAFT TO THE OBTUSE MARGINAL. WHEN THE RX VISION 4.0 X 28 MM CORONARY STENT SYSTEM WAS ADVANCED IN THE PATIENT, IT WAS NOTICED UNDER FLUOROSCOPY THAT THE STENT HAD COME OFF THE BALLOON. THE STENT COULD NOT BE VISUALIZED IN THE PATIENT AND WAS LATER FOUND IN THE TRASH. IT WAS DETERMINED THAT THE STENT HAD COME OFF THE BALLOON WHEN THE PROTECTIVE SHEATH WAS REMOVED. THE STENT SYSTEM WAS NOT INSPECTED PRIOR TO ENTERING THE PATIENT TO ENSURE THE STENT WAS ON THE BALLOON. THE PROCEDURE WAS COMPLETED BY DEPLOYING THREE ADDITIONAL RX VISION STENTS. THERE WAS NO REPORTED ADVERSE PATIENT EFFECT. THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6615272 ↗

Other STENT, CORONARY devices

All STENT, CORONARY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page