Acculink™ 1011343-30Recalls & reported problems

Abbott Vascular Inc. · RX ACCULINK™ Carotid Stent System 6-8 mm x 30 mm x 132 cm Rapid-Exchange · Class 3 · Implant

All models of Acculink (13)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08717648076367), received between Nov 28, 2018 and Sep 11, 2023 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Activation Failure” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Activation Failure2
  • Adverse Event Without Identified Device or Use Problem1
  • Deformation Due to Compressive Stress1
  • Difficult or Delayed Activation1
  • Difficult to Advance1
  • Failure to Advance1
  • Material Separation1

Patient problems

  • No Consequences Or Impact To Patient3
  • No Clinical Signs, Symptoms or Conditions1
  • Thrombosis/Thrombus1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 11, 2023Activation Failure; Activation Failure

    IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A LESION IN THE INTERNAL CAROTID ARTERY WITH MILD TORTUOSITY AND 70% STENOSIS. THE 6-8MM .014 ACCULINK SELF EXPANDING STENT (SES) WAS INTRODUCED VIA RIGHT FEMORAL APPROACH AND ATTEMPTED TO BE DEPLOYED; HOWEVER, AFTER UNLOCKING THE HANDLE, THE STENT WAS PARTIALLY DEPLOYED. ANOTHER .014 ACCULINK WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17721275 ↗
  • InjuryNov 18, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING PREPARATION OF THE EMBOSHIELD NAV6 THERE WAS DIFFICULTY LOADING THE FILTER INTO THE DELIVERY CATHETER. THEREFORE, THE NAV6 WAS NOT USED IN THE PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT WITH THIS DEVICE. AN ACCULINK WAS ADVANCED TO TREAT A LESION IN THE CAROTID ARTERY. THE STENT WAS IMPLANTED SUCCESSFULLY WITHOUT ISSUE. HOWEVER, DURING POST DILATATION OF THE ACCULINK THROMBUS WAS NOTED IN THE ARTERY. ASPIRATION WAS PERFORMED AND ADDITIONAL HEPARIN WAS ADMINISTERED TO TREAT THE THROMBUS SUCCESSFULLY. THERE WAS ADVERSE PATIENT SEQUELA. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Thrombosis/Thrombus

    MAUDE report 12838973 ↗
  • MalfunctionDec 26, 2019Activation Failure; Activation Failure

    SUBSEQUENT TO THE INITIALLY FILED REPORT, THE FOLLOWING INFORMATION WAS RECEIVED: RETURNED GOODS LAB IDENTIFIED A SHEATH SEPARATION ON THE DEVICE, BUT THIS WAS NOT NOTED DURING DURING THE PROCEDURE. THERE WAS NO RESISTANCE NOTED DURING THE PROCEDURE, AND IT WAS CONFIRMED THAT NOTHING REMAINS IN THE PATIENT ANATOMY. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9520024 ↗
  • MalfunctionOct 20, 2019Material Separation; Difficult or Delayed Activation; Deformation Due to Compressive Stress; Difficult to Advance; Material Separation; Difficult or Delayed Activation; Deformation Due to Compressive Stress; Difficult to Advance

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A HEAVILY TORTUOUS LEFT COMMON CAROTID ARTERY. THE PATIENT PRESENTED WITH AN ANEURYSM. THE PHYSICIAN ACCESSED THE RIGHT GROIN WITH A 6 X 9 NON-ABBOTT SHEATH. A 6-8/30 ACCULINK STENT DEVICE WAS ADVANCED AND THE SHAFT MADE A 90 DEGREE TURN (LARGE BEND) AT THE ANEURYSM. A BEND AT THE SHAFT WAS NOTICED. THERE WAS A BIT OF A RESISTANCE DUE TO THE ANATOMY, BUT THE DEVICE WAS ABLE TO CROSS THE TARGET LESION. INITIALLY, THE PHYSICIAN COULD NOT GET THE STENT TO DEPLOY. BUT THEN THE STENT ULTIMATELY DID DEPLOY. HOWEVER, DURING REMOVAL, THE DEVICE BECAME SEPARATED AT THE SHAFT. ALTHOUGH A SEPARATION OCCURRED, THE SEPARATED PORTION REMAINED ON THE MICRO WIRE AND EVERYTHING WAS SIMPLY PULLED OUT. NOTHING REMAINED IN THE PATIENT. THE PATIENT IS STABLE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9212556 ↗
  • InjuryNov 28, 2018Failure to Advance; Failure to Advance

    IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A LONG CALCIFIED LESION IN THE RIGHT INTERNAL CAROTID ARTERY. FOLLOWING THE DIFFICULT PLACEMENT OF A NON- ABBOTT INTRODUCER SHEATH, THE FILTER OF A LARGE EMBOSHIELD NAV6 EMBOLIC PROTECTION SYSTEM (EPS) WAS ADVANCED AND SUCCESSFULLY DEPLOYED DISTAL TO THE LESION. A 6-8 MM ACCULINK STENT DELIVERY SYSTEM (SDS) WAS ADVANCED BUT FAILED TO CROSS THE LESION DUE TO THE ANATOMY. THE LESION WAS THEREFORE FURTHER PRE-DILATED WITH A NON-ABBOTT BALLOON CATHETER. WHILST THE BALLOON WAS INFLATED, THE PATIENT SUFFERED A TRANSIENT ISCHEMIC ATTACK AND AN EPISODE OF HYPOTENSION. THE SYMPTOMS WERE TRANSIENT WHILE THE BALLOON WAS INFLATED. FOLLOWING THIS THE ACCULINK SDS WAS RE-ADVANCED; HOWEVER DID NOT CROSS THE LESION AGAIN DUE TO THE ANATOMY. AS A RESULT OF THE TYPE 2 ANATOMY, THE NON-ABBOTT GUIDE SHEATH PROLAPSED BACK AND WAS NO LONGER IN THE CAROTID ARTERY. NO ATTEMPTS WERE MADE TO REMOVE THE ACCULINK SDS OR THE FILTER DUE TO THE PROLAPSED SHEATH. THE PATIENT UNDERWENT A SURGICAL CUT DOWN TO REMOVE THE PROLAPSED SHEATH, THE ACCULINK SDS AND THE FILTER. A CAROTID ENDARTERECTOMY WAS PERFORMED FOR THE LESION. POST PROCEDURE THE PATIENT WAS REP

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8112064 ↗

Other STENT, CAROTID devices

All STENT, CAROTID devices →