Herculink Elite 1011490-15Recalls & reported problems

Abbott Vascular Inc. · RX Herculink Elite Renal and Biliary Stent System 4.5 mm x 15 mm x 135 cm · Class 3 · Implant

All models of Herculink Elite (23)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08717648077920), received between Jun 8, 2018 and Apr 29, 2025 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated2
  • Improper or Incorrect Procedure or Method2
  • Off-Label Use2
  • Activation Failure1
  • Component Missing1

Patient problems

  • Device Embedded In Tissue or Plaque1
  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 29, 2025Off-Label Use; Improper or Incorrect Procedure or Method; Component Missing; Off-Label Use; Improper or Incorrect Procedure or Method; Component Missing

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT THE LEFT VERTEBRAL ARTERY WITH MILD CALCIFICATION, MILD TORTUOSITY AND 90% STENOSIS. THE 4.5X15 MM HERCULINK ELITE STENT DELIVERY SYSTEM (SDS) WAS ADVANCED TO THE LESION; HOWEVER NO STENT WAS SEEN ON THE DELIVERY SYSTEM. THE STENT WAS NOT SEEN ON ANGIOGRAPHY. THE PHYSICIAN COULD NOT RECALL IF THE STENT WAS ON THE DELIVERY SYSTEM DURING PREP. THE DEVICE WAS REMOVED AND REPLACED. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS AND THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21931083 ↗
  • InjuryAug 25, 2021Off-Label Use; Device Dislodged or Dislocated; Off-Label Use; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A HEAVILY TORTUOUS RIGHT VERTEBRAL ARTERY. THE PHYSICIAN CHOSE A 4.5X15MM HERCULINK ELITE BALLOON-EXPANDABLE STENT SYSTEM (BESS) AS THERE WAS NO OTHER PRODUCT OPTION AVAILABLE THAT WAS THE APPROPRIATE SIZE FOR TREATMENT BASED ON THE DIAMETER AND LENGTH OF THE VESSEL. THE BESS WAS ADVANCED OVER THE GUIDE WIRE INTO THE PATIENT; HOWEVER, THE STENT DISLODGED. CURRENTLY, THE STENT REMAINS EMBEDDED IN AN UNSPECIFIED VESSEL WITH NO FURTHER TREATMENT PROVIDED. THERE WAS NO CLINICALLY SIGNIFICANT DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 12365173 ↗
  • MalfunctionOct 2, 2020Activation Failure; Activation Failure

    IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A 95% STENOSED, MODERATELY TORTUOUS, AND HEAVILY CALCIFIED LESION IN THE RIGHT RENAL ARTERY. A 4.5X15MM HERCULINK ELITE BALLOON-EXPANDABLE STENT SYSTEM (BESS) WAS ADVANCED TO THE LESION; HOWEVER, THE BALLOON FAILED TO INFLATE, AND THE STENT FAILED TO DEPLOY. THE BESS WAS REMOVED WITHOUT ISSUES AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A NEW 4.5X15MM HERCULINK STENT. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10620883 ↗
  • InjuryJun 8, 2018Improper or Incorrect Procedure or Method; Device Dislodged or Dislocated

    IT WAS REPORTED THAT DURING A PROCEDURE OF THE NARROW, MILDLY CALCIFIED, OSTIUM OF THE VERTEBRAL ARTERY, THE 4.5 X 15 MM HERCULINK ELITE STENT DELIVERY SYSTEM (SDS) WAS ADVANCED TO THE TARGET LESION; HOWEVER, THE STENT DISLODGED FROM THE BALLOON AND MIGRATED TO THE DISTAL END OF THE BRACHIAL ARTERY. THE SDS WAS REMOVED. ANGIOGRAPHY SHOWED GOOD BLOOD FLOW, SO NO ATTEMPT WAS MADE TO RETRIEVE THE DISLODGED STENT. THERE WAS NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: Foreign Body In Patient

    MAUDE report 7585054 ↗

Other STENT, RENAL devices

All STENT, RENAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page