Skip to content

Herculink Elite 1011502-12Recalls & reported problems

Abbott Vascular Inc. · RX Herculink Elite Renal and Biliary Stent System 6.5 mm x 12 mm x 135 cm · Class 3 · Implant

All models of Herculink Elite (23)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08717648078279), received between Oct 8, 2018 and Oct 8, 2018 — classified as 1 injury.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Device Dislodged or Dislocated1
  • Difficult to Remove1
  • Failure to Advance1
  • Improper or Incorrect Procedure or Method1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 8, 2018Difficult to Remove; Improper or Incorrect Procedure or Method; Failure to Advance; Device Dislodged or Dislocated; Difficult to Remove; Improper or Incorrect Procedure or Method; Failure to Advance; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A HEAVILY TORTUOUS, HEAVILY CALCIFIED LESION IN THE RENAL ARTERY. DURING A PERCUTANEOUS INTERVENTION, THE HERCULINK ELITE STENT DELIVERY SYSTEM (SDS) FAILED TO CROSS THE LESION DUE TO ANATOMY. DURING SDS RETRACTION, RESISTANCE WAS MET RETRACTING THE DEVICE INTO THE GUIDE CATHETER. DURING RETRACTION, THE HERCULINK ELITE STENT DISLODGED INTO THE ILIAC ARTERY. A WIRE WAS ATTEMPTED TO RETRIEVE THE DISLODGED HERCULINK ELITE STENT, HOWEVER, THE STENT THEN TRAVELED INTO THE PROFUNDA ARTERY. IT WAS DECIDED TO LEAVE THE STENT FREE FLOATING IN THE PROFUNDA AND NO ADDITIONAL TREATMENT WAS PROVIDED. ANOTHER HERCULINK ELITE STENT WAS SUCCESSFULLY USED IN REPLACEMENT. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Foreign Body In Patient

    MAUDE report 7945252 ↗

Other STENT, RENAL devices

All STENT, RENAL devices →

Sources and updates