TREK™ 1012272-25Recalls & reported problems
Abbott Vascular Inc. · TREK™ Coronary Dilatation Catheter 2.50 mm x 25 mm / Rapid-Exchange · Class 2
All models of TREK (50)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08717648138317), received between May 23, 2017 and Jul 11, 2024 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Detachment Of Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment Of Device Component1
- Device Markings/Labelling Problem1
- Inflation Problem1
- Leak/Splash1
- Material Separation1
Patient problems
- Air Embolism1
- Cardiac Arrest1
- Foreign Body In Patient1
- No Clinical Signs, Symptoms or Conditions1
- No Patient Involvement1
- Occlusion1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 11, 2024Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT ONCE THE 2.5X25MM TREK BALLOON DILATATION CATHETER WAS REMOVED FROM THE CHIPBOARD BOX THE DEVICE INSIDE THE BOX WAS ACTUALLY A 2.5X12MM TREK BALLOON. ANOTHER 2.5X25MM TREK BALLOON WAS OPENED AND USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19723657 ↗InjuryApr 20, 2020Inflation Problem; Leak/Splash; Inflation Problem; Leak/Splash
IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A MILDLY TORTUOUS AND RESTENOSED LESION IN THE CIRCUMFLEX ARTERY. THE PATIENT PRESENTED WITH NON-ST SEGMENT ELEVATION MYOCARDIAL INFRACTION. A 2.5X25MM TREK RX BALLOON DILATATION CATHETER (BDC) WAS PREPARED PER NORMAL PROCEDURE. THE BDC WAS ADVANCED TO THE TARGET LESION; HOWEVER, THE BALLOON DID NOT INFLATE. THE BDC WAS REMOVED TO ACCESS THE ISSUE. AT THIS POINT, THE PATIENTS PRESSURE BEGAN DROPPING RAPIDLY. ANOTHER ANGIO WAS TAKEN AND IT WAS NOTICED THAT THE LEFT MARGINAL ARTERY BECAME OCCLUDED. THE PATIENT ARRESTED. MORE HEPARIN WAS GIVEN AND THE PATIENT WAS RECOVERED USING CARDIOPULMONARY RESUSCITATION. FLOW WAS RESTORED AND THE PROCEDURE WAS QUICKLY COMPLETED. FOLLOWING THE PROCEDURE, THE BDC WAS INVESTIGATED AND IT IS BELIEVED THAT THERE MAY HAVE BEEN A SLOW LEAK, WHICH COLLECTED ATMOSPHERIC AIR UPON PREPARATION, WHICH SUBSEQUENTLY CAUSED AN AIR EMBOLISM. THE BDC WAS TESTED UNDER WATER AND A LEAK ON THE SHAFT WAS NOTED. THERE WERE NO ADVERSE PATIENT SEQUELA AND A CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE WAS REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Air Embolism; Cardiac Arrest; Occlusion
MAUDE report 9978396 ↗InjuryMar 25, 2020Material Separation; Material Separation
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT AN UNSPECIFIED LESION. IT WAS NOTED THAT DURING INFLATION OF A 2.5X25MM TREK BALLOON THE TIP AND BALLOON BECAME SEPARATED. THE BALLOON WAS SIMPLY REMOVED BUT IT IS UNKNOWN IF THE TIP WAS RETRIEVED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Foreign Body In Patient
MAUDE report 9877985 ↗MalfunctionMay 23, 2017Detachment Of Device Component
IT WAS REPORTED THAT DURING UNPACKAGING, WHILE REMOVING A 2.5X25MM TREK RX BALLOON DILATATION CATHETER (BDC) FROM ITS DISPENSER HOOP PACKAGING WITHOUT DIFFICULTY, THE HYPOTUBE SHAFT REPORTEDLY SNAPPED IN HALF (SEPARATED). THERE WAS NO DAMAGE NOTED TO ANY OF THE PACKAGING MATERIALS. THE DEVICE WAS NOT USED IN THE PATIENT AND THIS ISSUE DID NOT CAUSE OR CONTRIBUTE TO ANY CLINICALLY SIGNIFICANT DELAY. ANOTHER TREK RX BDC WAS UNPACKAGED AND USED WITHOUT ISSUE TO PERFORM THE PLANNED PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Patient Involvement
MAUDE report 6585931 ↗
Other Catheters, transluminal coronary angioplasty, percutaneous devices
- MINI TREK™ 1012270-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous223 reports
- Maverick²™ H7493892815200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous1 death203 reports
- MINI TREK™ 1012270-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous162 reports
- Maverick²™ H7493892820200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous161 reports
- NC TREK 1012454-08Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 recalls59 reports
- TREK™ 1012272-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 deaths158 reports
- NC TREK 1012453-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall84 reports
- NC TREK 1012453-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall1 death83 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page