NC TREK 1012451-06Recalls & reported problems
Abbott Vascular Inc. · NC TREK Coronary Dilatation Catheter 3.5 mm x 6 mm / Rapid-Exchange · Class 2
All models of NC TREK (56)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08717648151958), received between Jan 29, 2020 and Jan 17, 2022 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Difficult to Remove” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Difficult to Remove2
- Leak/Splash2
- Difficult to Advance1
- Failure to Advance1
- Inflation Problem1
- Material Rupture1
- Material Separation1
- Material Split, Cut or Torn1
Patient problems
- No Consequences Or Impact To Patient3
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 17, 2022Difficult to Remove; Failure to Advance; Difficult to Remove; Failure to Advance
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT THE MODERATELY CALCIFIED, HEAVILY TORTUOUS, CONCENTRIC LESION IN THE MID RIGHT CORONARY ARTERY (RCA). THE 3.5X6 MM NC TREK BALLOON DILATATION CATHETER (BDC) WAS ATTEMPTED TO ADVANCE BUT IT COULD NOT CROSS THE LESION DUE TO THE ANATOMY. THE NC TREK BDC WAS REMOVED INDEPENDENTLY WITH RESISTANCE NOTED WITH THE ANATOMY. OTHER DEVICES ALSO DID NOT CROSS THE LESION SO IT WAS ENDED AND ROTA-BLATION WAS SCHEDULED FOR NEXT TIME. THERE WERE NO ADVERSE PATIENT EFFECTS AND THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13266436 ↗MalfunctionSep 28, 2020Leak/Splash; Material Twisted/Bent; Material Split, Cut or Torn; Leak/Splash; Material Twisted/Bent; Material Split, Cut or Torn
ABBOTT VASCULAR RECEIVED THIS DEVICE WITHOUT A REPORTED COMPLAINT. UPON DEVICE INVESTIGATION, IT WAS NOTED THAT THE BALLOON LEAKED FROM A TEAR. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10596950 ↗MalfunctionAug 20, 2020Inflation Problem; Leak/Splash; Material Rupture; Inflation Problem; Leak/Splash; Material Rupture
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION LOCATED IN AN UNSPECIFIED CORONARY ARTERY. THE NC TREK RX BALLOON COULD NOT BE INFLATED. ANOTHER BALLOON WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. RETURNED DEVICE ANALYSIS NOTED A LEAK IN THE BALLOON AND A LONGITUDINAL RUPTURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10433914 ↗MalfunctionJan 29, 2020Difficult to Remove; Material Separation; Difficult to Advance; Difficult to Remove; Material Separation; Difficult to Advance
IT WAS REPORTED THAT THE 3.5X6 MM NC TREK WAS FLUSHED IN THE HOOP AND WAS PREPARED FOR USE. THE NC TREK WAS LOADED ON TO THE 014 SION BLUE GUIDE WIRE WITHOUT ISSUE. THE NC TREK WAS ADVANCED WITHOUT RESISTANCE UNTIL RIGHT BEFORE THE TUOHY WHEN THE NC TREK WOULD NO LONGER ADVANCE ON THE WIRE. THE NC TREK WAS DIFFICULT TO REMOVE AND FRACTURED IN TWO PIECES DURING REMOVAL FROM THE GUIDE WIRE. ANOTHER NC TREK WAS SUCCESSFULLY ADVANCED ON THE SAME GUIDE WIRE WITHOUT INCIDENT. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9644185 ↗
Other Catheters, transluminal coronary angioplasty, percutaneous devices
- MINI TREK™ 1012270-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous223 reports
- Maverick²™ H7493892815200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous1 death203 reports
- MINI TREK™ 1012270-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous162 reports
- Maverick²™ H7493892820200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous161 reports
- NC TREK 1012454-08Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 recalls59 reports
- TREK™ 1012272-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 deaths158 reports
- NC TREK 1012453-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall84 reports
- NC TREK 1012453-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall1 death83 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page