Skip to content

TREK 1012404-12Recalls & reported problems

Abbott Vascular Inc. · TREK Coronary Dilatation Catheter 2.25 mm x 12 mm / Over-The-Wire · Class 2

All models of TREK (50)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08717648153907), received between Aug 24, 2021 and Aug 24, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Inflation Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Inflation Problem1
  • Material Rupture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 24, 2021Inflation Problem; Material Rupture; Inflation Problem; Material Rupture

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A HEAVILY CALCIFIED LESION IN THE DIAGONAL ARTERY. THE 2.25X12MM TREK OTW BALLOON DILATATION CATHETER (BDC) WAS ADVANCED TO THE TARGET LESION WITHOUT ISSUE. DURING THE FIRST INFLATION AT 6 ATMOSPHERES (ATM), IT WAS NOTICED THAT THE BALLOON DID NOT APPEAR TO BE FULLY INFLATING. NEGATIVE PRESSURE WAS APPLIED, AND DURING THE SECOND INFLATION AT 8 ATM, THE BALLOON STILL DID NOT APPEAR TO BE FULLY INFLATING. THE BDC WAS THEN REMOVED AND TESTED OUTSIDE THE PATIENT. A SMALL PINHOLE WAS OBSERVED ON THE DISTAL END OF THE BALLOON AND AIR WAS COMING OUT. THE PATIENT DID NOT NEED TO BE RE-BALLOONED AS THE TREK WAS ABLE TO DILATE THE LESION DESPITE THE BALLOON NOT FULLY INFLATING. THE TARGET LESION WAS TREATED, AND THE PATIENT REMAINED STABLE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12363080 ↗

Other Catheters, transluminal coronary angioplasty, percutaneous devices

All Catheters, transluminal coronary angioplasty, percutaneous devices →