Skip to content

Armada B1070-150Recalls & reported problems

Abbott Vascular Inc. · Armada 35 LL PTA Catheter 7 mm x 150 mm x 80 cm / Over-The-Wire · Class 2

All models of Armada (213)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08717648154836), received between May 12, 2022 and May 12, 2022 — classified as 1 injury.

The most frequently reported specific device problem is “Difficult to Remove” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Difficult to Remove1
  • Failure to Fold1
  • Material Rupture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 12, 2022Failure to Fold; Difficult to Remove; Material Rupture; Failure to Fold; Difficult to Remove; Material Rupture

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT THE HEAVILY CALCIFIED, 60% STENOSED, RIGHT SUPERFICIAL FEMORAL ARTERY (SFA). THE 7X150 MM ARMADA 35 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) CATHETER WAS INFLATED ONCE TO NOMINAL PRESSURE AND RUPTURED AND BUNCHED UP, CAUSING IT TO BE DIFFICULT TO REMOVE WITHOUT EXCESSIVE FORCE. THEREFORE, A 5 CM CUTDOWN IN THE SFA WAS PERFORMED TO REMOVE THE DEVICE. REPORTEDLY, THERE WAS A 30 MINUTE DELAY IN THE PROCEDURE TO PERFORM THE CUT DOWN. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14378347 ↗

Other CATHETER, PERCUTANEOUS devices

All CATHETER, PERCUTANEOUS devices →

Sources and updates