Skip to content

Armada B1090-020Recalls & reported problems

Abbott Vascular Inc. · Armada 35 PTA Catheter 9 mm x 20 mm x 80 cm / Over-The-Wire · Class 2

All models of Armada (213)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08717648154898), received between Feb 11, 2021 and Jun 8, 2023 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Leak/Splash” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Leak/Splash1
  • Material Rupture1
  • Material Split, Cut or Torn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 8, 2023Leak/Splash; Material Split, Cut or Torn; Leak/Splash; Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING PREPARATION OF AND BEFORE USE IN THE PATIENT, WHEN NEGATIVE PRESSURE WAS PULLED, THERE WAS A LEAK OBSERVED IN THE BALLOON [TEAR/HOLE IN BALLOON] ON A 9X20MM ARMADA 35 BALLOON CATHETER AN A 4.0X60MM ARMADA 14 BALLOON CATHETER. THE SAME ISSUE OCCURRED WITH BOTH BALLOONS. ANOTHER ARMADA WAS USED TO CONTINUE THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17092161 ↗
  • MalfunctionFeb 11, 2021Material Rupture; Material Rupture

    IT WAS REPORTED THAT THIS WAS A PROCEDURE TO TREAT A HIGHLY STENOSIS LEFT ARTERIOVENOUS (AV) FISTULA. AN ARMADA BALLOON CATHETER WAS INSERTED AND ADVANCED WITHOUT ISSUES. HOWEVER, DURING INFLATION OF THE BALLOON, IT RUPTURED AT 14ATM. THE BALLOON CATHETER WAS ABLE TO BE REMOVED WITHOUT ISSUES AND WAS REPLACED WITH A NON-ABBOTT DEVICE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11314137 ↗

Other CATHETER, PERCUTANEOUS devices

All CATHETER, PERCUTANEOUS devices →

Sources and updates