Armada B2040-020Recalls & reported problems
Abbott Vascular Inc. · Armada 35 PTA Catheter 4 mm x 20 mm x 135 cm / Over-The-Wire · Class 2
All models of Armada (213)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08717648155079), received between Sep 12, 2019 and Dec 23, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Deflation Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Deflation Problem1
- Inflation Problem1
- Leak/Splash1
- Material Rupture1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Patient Involvement1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 23, 2025Deflation Problem; Inflation Problem; Deflation Problem; Inflation Problem
SUBSEQUENTLY, RETURNED DEVICE ANALYSIS OBSERVED THAT THE BALLOON CATHETER INFLATED AND MAINTAINED PRESSURE. THE BALLOON INFLATED WITH LIQUID, THERE WAS NO ISSUE AS REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23881826 ↗MalfunctionJun 29, 2023Material Rupture; Material Rupture
USER FACILITY MEDWATCH REPORT RECEIVED THAT STATES: DURING ANGIOGRAM, THE ANGIOPLASTY BALLOON WITH A NOMINAL INFLATION RATING OF 8 ATM AND A MAXIMUM RATING OF 24 ATM BURST AT 20 ATM. BALLOON REMOVED INTACT. NO HARM TO THE PATIENT. ADDITIONAL INFORMATION RECEIVED: IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT AN 80% STENOSED LESION IN THE RIGHT COMMON ILIAC ARTERY WITH CALCIFICATION. THE 4.00X20MM ARMADA BALLOON DILATATION CATHETER (BDC) WAS ADVANCED TO THE TARGET LESION AND THE BALLOON WAS INFLATED TO 20 ATMOSPHERES (ATM) FOR 30 SECONDS; HOWEVER, THE BALLOON RUPTURED DURING THE FIRST INFLATION. THEREFORE, THE BDC WAS REMOVED FROM THE PATIENT. THERE WAS NOT ADVERSE PATIENT EFFECT AND NO CLINICALLY SIGNIFICANT DELAY REPORTED IN THE PROCEDURE. PRIOR TO USE, THE ARMADA BDC WAS PREPPED (AIR ASPIRATION) OUTSIDE THE ANATOMY WITHOUT ISSUE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17226638 ↗MalfunctionSep 12, 2019Leak/Splash; Leak/Splash
IT WAS REPORTED THAT DURING PREPARATION, AIR ASPIRATION, OF A 4.0 X 20 MM ARMADA 35 BALLOON CATHETER AN AIR LEAK WAS NOTED. THE BALLOON CATHETER WAS NOT USED AND THERE WAS NO PATIENT INVOLVEMENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH AN UNSPECIFIED ARMADA BALLOON CATHETER. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Patient Involvement
MAUDE report 9014129 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.