Skip to content

Armada B2040-020Recalls & reported problems

Abbott Vascular Inc. · Armada 35 PTA Catheter 4 mm x 20 mm x 135 cm / Over-The-Wire · Class 2

All models of Armada (213)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08717648155079), received between Sep 12, 2019 and Dec 23, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Deflation Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Deflation Problem1
  • Inflation Problem1
  • Leak/Splash1
  • Material Rupture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Patient Involvement1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 23, 2025Deflation Problem; Inflation Problem; Deflation Problem; Inflation Problem

    SUBSEQUENTLY, RETURNED DEVICE ANALYSIS OBSERVED THAT THE BALLOON CATHETER INFLATED AND MAINTAINED PRESSURE. THE BALLOON INFLATED WITH LIQUID, THERE WAS NO ISSUE AS REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23881826 ↗
  • MalfunctionJun 29, 2023Material Rupture; Material Rupture

    USER FACILITY MEDWATCH REPORT RECEIVED THAT STATES: DURING ANGIOGRAM, THE ANGIOPLASTY BALLOON WITH A NOMINAL INFLATION RATING OF 8 ATM AND A MAXIMUM RATING OF 24 ATM BURST AT 20 ATM. BALLOON REMOVED INTACT. NO HARM TO THE PATIENT. ADDITIONAL INFORMATION RECEIVED: IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT AN 80% STENOSED LESION IN THE RIGHT COMMON ILIAC ARTERY WITH CALCIFICATION. THE 4.00X20MM ARMADA BALLOON DILATATION CATHETER (BDC) WAS ADVANCED TO THE TARGET LESION AND THE BALLOON WAS INFLATED TO 20 ATMOSPHERES (ATM) FOR 30 SECONDS; HOWEVER, THE BALLOON RUPTURED DURING THE FIRST INFLATION. THEREFORE, THE BDC WAS REMOVED FROM THE PATIENT. THERE WAS NOT ADVERSE PATIENT EFFECT AND NO CLINICALLY SIGNIFICANT DELAY REPORTED IN THE PROCEDURE. PRIOR TO USE, THE ARMADA BDC WAS PREPPED (AIR ASPIRATION) OUTSIDE THE ANATOMY WITHOUT ISSUE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17226638 ↗
  • MalfunctionSep 12, 2019Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT DURING PREPARATION, AIR ASPIRATION, OF A 4.0 X 20 MM ARMADA 35 BALLOON CATHETER AN AIR LEAK WAS NOTED. THE BALLOON CATHETER WAS NOT USED AND THERE WAS NO PATIENT INVOLVEMENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH AN UNSPECIFIED ARMADA BALLOON CATHETER. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Patient Involvement

    MAUDE report 9014129 ↗

Other CATHETER, PERCUTANEOUS devices

All CATHETER, PERCUTANEOUS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page