Armada B2070-080Recalls & reported problems
Abbott Vascular Inc. · Armada 35 PTA Catheter 7 mm x 80 mm x 135 cm / Over-The-Wire · Class 2
All models of Armada (213)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 08717648155376), received between Feb 23, 2018 and Sep 29, 2026 — classified as 2 injuries, 7 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture7
- Break1
- Difficult to Advance1
- Difficult to Insert1
- Difficult to Remove1
- Inflation Problem1
- Leak/Splash1
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Consequences Or Impact To Patient3
- Vascular Dissection1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026Break; Inflation Problem; Leak/Splash
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50935933 ↗InjuryFeb 5, 2026Difficult to Advance
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A RIGHT EXTERNAL ILIAC ARTERY. PRE-DILATATION WAS PERFORMED WITH A 7X80MM ARMADA 35 BALLOON DILATATION (BDC. SUBSEQUENTLY, A DISSECTION WAS NOTED. THE BDC POSSIBLY CONTRIBUTED TO THE DISSECTION. THEN AN UNSPECIFIED STENT WAS IMPLANTED TO TREAT THE DISSECTION. POST STENT IMPLANTATION, THE BDC WAS ATTEMPTED TO BE RE-USED. HOWEVER, THE BDC COULD NOT ADVANCE THROUGH THE 5F SHEATH. THEREFORE, THE PROCEDURE WAS COMPLETED WITH ANOTHER ARMADA BDC DEVICE. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Vascular Dissection
MAUDE report 24265347 ↗MalfunctionNov 19, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A HEAVILY CALCIFIED LESION IN THE SUPERFICIAL FEMORAL ARTERY (SFA). A 7X80MM ARMADA 35 BALLOON DILATATION CATHETER (BDC) WAS ATTEMPTED TO BE USED FOR VESSEL DILATATION; HOWEVER, DURING THE THIRD INFLATION THE BALLOON WAS NOTED TO RUPTURE AT 7 ATMOSPHERES. THE BDC WAS REMOVED AND A NON-ABBOTT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED ADVERSE PATIENT EFFECT NOR CLINICALLY SIGNIFICANT DELAY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20724097 ↗InjuryNov 6, 2024Difficult to Remove; Material Rupture; Material Separation; Difficult to Remove; Material Rupture; Material Separation
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT AN 80% RE-STENOSED AV FISTULA. THE 7X80MM ARMADA 35 BALLOON DILATATION CATHETER (BDC) WAS ADVANCED TO THE TARGET LESION AND THE BALLOON WAS INFLATED ONCE; HOWEVER, THE BALLOON RUPTURED AT 12 ATMOSPHERES (ATM). THEREFORE, THE BDC WAS REMOVED AND RESISTANCE WAS NOTED WITH THE SHEATH. A PIECE OF THE BALLOON SEPARATED IN THE ANATOMY AND CUT DOWN SURGERY WAS PERFORMED TO REMOVE THE SEPARATED PIECE OF BALLOON. THERE WAS NO ADVERSE PATIENT SEQUELA. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20626017 ↗MalfunctionMar 27, 2024Material Rupture; Material Rupture
IT WAS REPORTED THE PROCEDURE WAS TO TREAT THE LEFT FEMORAL ARTERY WITH MODERATE CALCIFICATION AND NO TORTUOSITY. THE 9X60 MM ARMADA PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) CATHETER WAS ADVANCED TO THE LESION WITHOUT RESISTANCE AND INFLATED BUT RUPTURED AT 8 ATMOSPHERES (ATM) DURING THE FIRST INFLATION. A 7X80 MM ARMADA PTA CATHETER WAS ADVANCED TO THE LESION WITHOUT RESISTANCE AND WAS INFLATED BUT RUPTURED AT 8 ATM DURING THE SECOND INFLATION. A THIRD ARMADA PTA CATHETER WAS USED TO COMPLETE THE PROCEDURE WITHOUT ISSUE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18987588 ↗MalfunctionNov 1, 2021Material Rupture; Material Rupture
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A HEAVILY CALCIFIED, LESION IN THE LEFT SUPERFICIAL FEMORAL ARTERY. THE 7.0MMX80MMX135CM ARMADA 35 BALLOON DILATATION CATHETER (BDC) WAS ADVANCED TO THE TARGET LESION. DURING THE FIRST INFLATION, TO 8 ATMOSPHERES, THE BALLOON RUPTURED. A 7.0MMX60MMX135CM ARMADA 35 BDC WAS ADVANCED TO THE TARGET LESION AND THE BALLOON RUPTURED AT 8 ATMOSPHERES, DURING THE FIRST INFLATION. ANOTHER ARMADA BALLOON WAS USED TO COMPLETE THE PROCEDURE WITHOUT FURTHER ISSUES. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12732717 ↗MalfunctionOct 23, 2019Material Rupture; Material Rupture
IT WAS REPORTED THAT THE 7.0X80 MM ARMADA 35 WAS ADVANCED TO THE HEAVILY CALCIFIED LESION IN THE RIGHT SUPERFICIAL ARTERY. THE ARMADA WAS INFLATED AT 10 ATMOSPHERES BUT RUPTURED WITH THE FIRST INFLATION DUE TO THE HEAVY CALCIUM. THE ARMADA REMAINED INTACT AND WAS REMOVED FROM THE ARTERY WITHOUT INCIDENT. THE ABSOLUTE PRO VASCULAR WAS SUCCESSFULLY IMPLANTED. THE 8.0X60 ARMADA 35 WAS USED FOR POST DILATATION OF THE STENT. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9227306 ↗MalfunctionMay 17, 2019Material Rupture; Material Rupture
IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A 95% STENOSED AND HEAVILY CALCIFIED LESION IN THE SUPERFICIAL FEMORAL ARTERY. AFTER NINE PASSES OF ATHERECTOMY WERE PERFORMED, THE 7.0X80MM ARMADA PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) CATHETER WAS THEN ADVANCED WITHOUT RESISTANCE AND INFLATED AN UNKNOWN NUMBER OF TIMES WHEN THE BALLOON RUPTURED. THE RATED BURST PRESSURE WAS UNKNOWN BUT WAS NOTED TO BE AT OR BELOW NOMINAL. THE DEVICE WAS REMOVED WITHOUT REPORTED ISSUE AND REPLACED WITH ANOTHER ARMADA 35 PTA CATHETER TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT OR A CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8621583 ↗MalfunctionFeb 23, 2018Difficult to Insert; Material Rupture
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT LESION IN THE RIGHT SUPERFICIAL FEMORAL ARTERY. A 7X80MM ARMADA 35 BALLOON CATHETER WAS USED TO PRE-DILATE THE LESION. THE BALLOON CATHETER WAS REMOVED AND AN UNKNOWN STENT WAS IMPLANTED; HOWEVER, WHEN ATTEMPTING TO RE-INSERT THE BALLOON CATHETER FOR POST-DILATATION IT COULD NOT BE INSERTED IN THE SHEATH. A 7X60MM ARMADA 35 BALLOON CATHETER WAS USED TO POST-DILATE THE STENT. THE BALLOON CATHETER WAS REMOVED AND AN ATTEMPT TO INSERT IT AGAIN FOR ADDITIONAL POST-DILATATION WAS MADE; HOWEVER, IT COULD NOT BE INSERTED INTO THE SHEATH. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH 7X80MM NON-ABBOTT BALLOON CATHETER. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE AND NO ADVERSE PATIENT EFFECTS. NO ADDITIONAL INFORMATION WAS PROVIDED. THE RETURN DEVICE ANALYSIS NOTED PINHOLE RUPTURES IN BOTH OF THE ARMANDA BALLOONS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7293121 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page