Armada B2100-060Recalls & reported problems
Abbott Vascular Inc. · Armada 35 PTA Catheter 10 mm x 60 mm x 135 cm / Over-The-Wire · Class 2
All models of Armada (213)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 1
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08717648155529), received between Dec 15, 2021 and Dec 9, 2025 — classified as 1 death, 2 injuries.
The most frequently reported specific device problem is “Difficult to Remove” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult to Remove3
- Material Separation3
- Material Rupture2
Patient problems
- Device Embedded In Tissue or Plaque1
- No Clinical Signs, Symptoms or Conditions1
- Unspecified Heart Problem1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
DeathDec 9, 2025Difficult to Remove; Material Separation; Difficult to Remove; Material Separation
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION LEFT SUBCLAVIAN ARTERY AND CEPHALIC VEIN. THE 10X60MM ARMADA 35 BALLOON DILATATION CATHETER (BDC) WAS USED IN THE PROCEDURE; HOWEVER, DURING RETRACTION OF THE BDC RESISTANCE WAS NOTED AND THE BALLOON DISLODGED FROM THE BDC SHAFT INSIDE THE PATIENT. A SNARE WAS USED TO REMOVE THE DISLODGED BALLOON WAS REMOVED FROM THE PATIENT. THERE WAS A CLINICALLY SIGNIFICANT DELAY REPORTED IN THE PROCEDURE. POST PROCEDURE THE PATIENT REPORTEDLY DIED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Unspecified Heart Problem
MAUDE report 23759420 ↗InjuryJan 6, 2022Difficult to Remove; Material Rupture; Material Separation; Difficult to Remove; Material Rupture; Material Separation
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A HEAVILY CALCIFIED RE-THROMBOSED STENT. AFTER THE 10X60MM ARMADA 35 BALLOON DILATATION CATHETER (BDC) WAS ADVANCED TO THE LESION, AND AFTER SEVERAL INFLATIONS OF THE BALLOON, BETWEEN NOMINAL PRESSURE AND ABOVE NOMINAL, THE BALLOON RUPTURED. IT WAS NOTED THE RUPTURE OCCURRED DUE TO THE PREVIOUSLY IMPLANTED STENT HAD FRACTURED AND CAUSED THE BALLOON TO RUPTURE ON CONTACT. DURING REMOVAL OF THE BDC RESISTANCE WAS MET WITH THE INTRODUCER AND THE CENTRAL SHEATH OF THE BALLOON SEPARATED. IT WAS NOTED ANOTHER UNSPECIFIED ARMADA 35 BDC WAS ADVANCED; HOWEVER, THE SAME EXACT ISSUE OCCURRED AND BDC RUPTURED AT AN UNKNOWN PRESSURE. THEREFORE, TWO FRAGMENTS REMAINED IN THE ANATOMY AND THE FRAGMENTS WERE PRESSED AGAINST THE VASCULAR WALL. THERE WAS NO ADVERSE PATIENT SEQUELA REPORTED AND NO CLINICALLY SIGNIFICANT DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 13177583 ↗InjuryDec 15, 2021Difficult to Remove; Material Rupture; Material Separation; Difficult to Remove; Material Rupture; Material Separation
IT WAS REPORTED THAT DURING A PROCEDURE TO TREAT A HEAVILY CALCIFIED RE-THROMBOSED STENT, AFTER THE 10X60MM ARMADA 35 BALLOON DILATATION CATHETER (BDC) WAS ADVANCED TO THE LESION, AND AFTER SEVERAL INFLATIONS OF THE BALLOON, BETWEEN NOMINAL PRESSURE AND ABOVE NOMINAL PRESSURE, THE BALLOON RUPTURED WITH ONE FRAGMENT SEPARATING FROM THE DEVICE. RESISTANCE WAS NOTED WHEN PULLING THE BDC BACK INTO THE INTRODUCER SHEATH. SURGICAL INTERVENTION WAS REQUIRED TO REMOVE BALLOON FRAGMENT FROM THE PATIENT ANATOMY, WHICH WAS SUCCESSFULLY REMOVED. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13014690 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026