Skip to content

Armada A1040-060Recalls & reported problems

Abbott Vascular Inc. · Armada 14 PTA Catheter 4.0 mm x 60 mm x 90 cm / Over-The-Wire · Class 2

All models of Armada (213)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08717648155864), received between Jun 9, 2023 and Jun 9, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Difficult to Remove” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Difficult to Remove1
  • Improper or Incorrect Procedure or Method1
  • Material Rupture1
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 9, 2023Difficult to Remove; Material Rupture; Material Separation; Improper or Incorrect Procedure or Method; Difficult to Remove; Material Rupture; Material Separation; Improper or Incorrect Procedure or Method

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A FISTULA IN THE ARTERIAL VENOUS SHUNT FOR HEMODIALYSIS. THE 4.0X60MM ARMADA 14 BALLOON DILATATION CATHETER (BDC) WAS SOAKED IN SALINE AND PREPARED (AIR ASPIRATION) OUTSIDE THE ANATOMY PRIOR TO USE. THERE WAS NO RESISTANCE DURING ADVANCEMENT, HOWEVER, THE BALLOON WAS INFLATED TWO TIMES AT 10 ATMOSPHERES (ATMS) AND 14 ATMS AND RUPTURED DURING THE SECOND INFLATION. THERE WAS RESISTANCE WITH THE ANATOMY DURING REMOVAL, FORCE WAS APPLIED AND THE BALLOON TORE IN A CIRCULAR WAY. SURGICAL INTERVENTION WAS PERFORMED TO RECOVER THE END OF THE BALLOON REMAINING IN THE ANATOMY. THERE WAS NO ADVERSE PATIENT SEQUELA AND THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17099922 ↗