Skip to content

Armada A1040-120Recalls & reported problems

Abbott Vascular Inc. · Armada 14 PTA Catheter 4.0 mm x 120 mm x 90 cm / Over-The-Wire · Class 2

All models of Armada (213)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08717648155888), received between Jun 30, 2025 and Jun 30, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Deformation Due to Compressive Stress” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Deformation Due to Compressive Stress1
  • Difficult to Advance1
  • Material Rupture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 30, 2025Material Rupture; Deformation Due to Compressive Stress; Difficult to Advance; Material Rupture; Deformation Due to Compressive Stress; Difficult to Advance

    IT WAS REPORTED THAT THE PROCEDURE PERFORMED WAS TO TREAT AN OCCLUDED, HEAVILY CALCIFIED, MILDLY TORTUOUS POPLITEAL ARTERY USING A 4GX25CM HI-TORQUE (HT) COMMAND 14 GUIDEWIRE AND A 2.0X20MM ARMADA 14 XT BALLOON DILATATION CATHETER (BDC). THE HT COMMAND GUIDEWIRE AND ARMADA 14 BDC STUCK TOGETHER DURING REMOVAL AND WERE WITHDRAWN AS ONE UNIT. A 4.0X120MM ARMADA 14 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BDC WAS TO BE USED IN THE HEAVILY CALCIFIED, MILDLY TORTUOUS LEFT INTERIOR TIBIAL. DURING ADVANCEMENT OF THE ARMADA 14 PTA BDC, RESISTANCE WAS MET DUE TO ANATOMY. ONCE AT THE LESION, THE ARMADA 14 PTA BDC WAS INFLATED ONCE TO 6 ATMOSPHERES AND COLLAPSED ON ITSELF LIKE AN ACCORDION. THE PROCEDURE WAS TERMINATED AND IT IS UNKNOWN IF OTHER DEVICES WERE USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22362266 ↗

Other CATHETER, PERCUTANEOUS devices

All CATHETER, PERCUTANEOUS devices →

Sources and updates