Armada A2015-020Recalls & reported problems

Abbott Vascular Inc. · Armada 14 PTA Catheter 1.5 mm x 20 mm x 150 cm / Over-The-Wire · Class 2

All models of Armada (213)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08717648155901), received between Mar 17, 2022 and May 20, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture2
  • Difficult to Advance1
  • Material Split, Cut or Torn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 20, 2024Material Rupture; Difficult to Advance; Material Rupture; Difficult to Advance

    IT WAS REPORTED THE PROCEDURE WAS TO TREAT A HEAVILY CALCIFIED LEFT ANTERIOR TIBIAL ARTERY. DURING PRE-DILATATION, THE 1.5X20X150CM ARMADA 14 BALLOON WAS ADVANCED BUT THERE WAS NOTED RESISTANCE DURING ADVANCEMENT DUE TO CALCIUM, HOWEVER IT REACHED THE LESION. DURING THE FIRST INFLATION, THE BALLOON RUPTURED AT 6 ATMOSPHERES (ATM). A 1.5X120X150CM ARMADA 14 BALLOON WAS THEN USED BUT THERE WAS NOTED RESISTANCE DURING ADVANCEMENT DUE TO CALCIUM, HOWEVER IT REACHED THE LESION. DURING THE FIRST INFLATION, THE BALLOON RUPTURED AT 5ATM. A NON-ABBOTT DEVICE WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19353472 ↗
  • MalfunctionMay 4, 2022Material Rupture; Material Rupture

    IT WAS REPORTED THAT ON (B)(6) 2022, A PERCUTANEOUS INTERVENTION WAS PERFORMED ON THE PERONEAL ARTERY LESION. AN ARMADA BALLOON CATHETER ADVANCED WITHOUT ISSUE. ON THE INITIAL INFLATION ATTEMPT, THE BALLOON RUPTURED AT 12 ATMOSPHERES. THE DEVICE WAS REMOVED WITHOUT ISSUES. THERE WAS NO ADVERSE PATIENT EFFECT AND THERE WAS NO CLINICALLY SIGNIFICANT DELAY. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14283402 ↗
  • MalfunctionMar 17, 2022Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED DURING PREPARATION OF AN ARMADA 14 1.5MM/20MM/150CM BALLOON DILATATION CATHETER (BDC), THE PHYSICIAN REMOVED THE PROTECTIVE COVER AND NOTICED THE BALLOON WAS BROKEN. THEREFORE, THE DEVICE WAS NOT USED AND WAS REPLACED. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED. ANALYSIS OF THE RETURNED DEVICE IDENTIFIED A LONGITUDINAL TEAR IN BOTH SIDES OF THE BALLOON 5.5MM PROXIMAL TO THE PROXIMAL END OF THE DISTAL BALLOON MARKER FOR A LENGTH OF 8MM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13794903 ↗

Other CATHETER, PERCUTANEOUS devices

All CATHETER, PERCUTANEOUS devices →