Skip to content

XIENCE XPEDITION 1074250-08Recalls & reported problems

Abbott Vascular Inc. · XIENCE Xpedition Everolimus Eluting Coronary Stent System 2.50 mm x 8 mm / Rapid-Exchange · Class 3 · Implant

All models of XIENCE XPEDITION (41)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08717648162114), received between May 26, 2021 and May 26, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break1
  • Deformation Due to Compressive Stress1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 26, 2021Break; Deformation Due to Compressive Stress; Break; Deformation Due to Compressive Stress

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

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11886777 ↗

Other Coronary drug-eluting stent devices

All Coronary drug-eluting stent devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page