Absolute Pro 1012528-100Recalls & reported problems
Abbott Vascular Inc. · Absolute Pro Vascular Self-Expanding Stent System 6.0 mm x 100 mm x 80 cm / Over-The-Wire · Class 3 · Implant
All models of Absolute Pro (51)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08717648175572), received between May 22, 2020 and Jun 9, 2020 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Activation Failure” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Activation Failure2
- Deformation Due to Compressive Stress1
- Material Deformation1
- Material Separation1
- Premature Activation1
Patient problems
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 9, 2020Premature Activation; Deformation Due to Compressive Stress; Activation Failure; Premature Activation; Deformation Due to Compressive Stress; Activation Failure
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT THE RIGHT SUPERFICIAL FEMORAL ARTERY (SFA). THE PATIENT WAS HEPARINIZED AND A SHORT 6FR SHEATH WAS PLACED AT THE PEDAL ACCESS SITE. THE FIRST 6.0X100 ABSOLUTE PRO STENT WAS ATTEMPTED TO BE DEPLOYED BUT THE SHEATH OF THE STENT KINKED. THE STENT DID NOT DEPLOYED AT ALL. A NEW, SAME SIZE, ABSOLUTE PRO STENT WAS ADVANCED AND THE STENT PARTIALLY DEPLOYED FROM THE SHEATH. THE STENT WAS ABLE TO BE REMOVED, ALTHOUGH IT WAS DAMAGED. A NEW ABSOLUTE PRO STENT WAS ABLE TO BE SUCCESSFULLY DEPLOYED. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS AND THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. RETURNED DEVICE ANALYSIS IDENTIFIED THAT THE FIRST 6.0X100 ABSOLUTE PRO STENT WAS PARTIALLY EXPOSED FOR A LENGTH OF 1MM. THE SECOND ABSOLUTE PRO STENT WAS NOTED TO HAVE SEPARATED STRUTS AT THE DISTAL END OF THE STENT AND THE SEPARATED PORTION WAS NOT RETURNED. CONTACT WITH THE ACCOUNT CONFIRMED THAT THE STENT DID SEPARATE AND WAS REMOVED IN PIECES BY SNARE. THE PIECES WERE DISCARDED AND NONE WERE LEFT IN THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10136151 ↗InjuryMay 22, 2020Material Separation; Material Deformation; Activation Failure; Material Separation; Material Deformation; Activation Failure
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT THE RIGHT SUPERFICIAL FEMORAL ARTERY (SFA). THE PATIENT WAS HEPARINIZED AND A SHORT 6FR SHEATH WAS PLACED AT THE PEDAL ACCESS SITE. THE FIRST 6.0X100 ABSOLUTE PRO STENT WAS ATTEMPTED TO BE DEPLOYED BUT THE SHEATH OF THE STENT KINKED. THE STENT DID NOT DEPLOY AT ALL. A NEW, SAME SIZE, ABSOLUTE PRO STENT WAS ADVANCED AND THE STENT PARTIALLY DEPLOYED FROM THE SHEATH. THE STENT WAS ABLE TO BE REMOVED, ALTHOUGH IT WAS DAMAGED. A NEW ABSOLUTE PRO STENT WAS ABLE TO BE SUCCESSFULLY DEPLOYED. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS AND THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED. RETURNED DEVICE ANALYSIS IDENTIFIED THAT THE FIRST 6.0X100 ABSOLUTE PRO STENT WAS PARTIALLY EXPOSED FOR A LENGTH OF 1MM.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10083578 ↗
Other STENT, ILIAC devices
- Cordis S.M.A.R.T. Control Vascular Stent System C10040MLCordis US Corp. · STENT, ILIAC1 recall3 reports
- Absolute Pro 1012534-100Abbott Vascular Inc. · STENT, ILIAC51 reports
- Cordis S.M.A.R.T. Control Vascular Stent System C09060MLCordis US Corp. · STENT, ILIAC1 recall0 reports
- Protégé™ GPS™ SERB65-09-60-80Ev3, Inc · STENT, ILIAC1 recall0 reports
- Absolute Pro 1012535-100Abbott Vascular Inc. · STENT, ILIAC1 death41 reports
- Omnilink Elite 1012631-39Abbott Vascular Inc. · STENT, ILIAC27 reports
- Omnilink Elite 1012632-29Abbott Vascular Inc. · STENT, ILIAC25 reports
- Omnilink Elite 1012631-59Abbott Vascular Inc. · STENT, ILIAC21 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page