Absolute Pro 1012531-80Recalls & reported problems

Abbott Vascular Inc. · Absolute Pro Vascular Self-Expanding Stent System 9.0 mm x 80 mm x 80 cm / Over-The-Wire · Class 3 · Implant

All models of Absolute Pro (51)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08717648175800), received between Feb 10, 2017 and Apr 5, 2021 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Difficult or Delayed Activation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Difficult or Delayed Activation1
  • Physical Resistance/Sticking1
  • Positioning Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 5, 2021Difficult or Delayed Activation; Physical Resistance/Sticking; Difficult or Delayed Activation; Physical Resistance/Sticking

    IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A CALCIFIED LESION IN THE ILIAC VESSEL. THE 9.0 X 80 MM ABSOLUTE PRO SELF EXPANDING STENT SYSTEM WAS ADVANCED TO THE TARGET LESION. DURING STENT DEPLOYMENT, THE DEPLOYMENT WHEEL STOPPED HALF WAY THROUGH DEPLOYMENT, BUT WITH MANIPULATION THE STENT WAS ABLE TO COMPLETELY DEPLOY AT THE INTENDED SITE. POST DILATATION WAS COMPLETED SUCCESSFULLY. THERE WAS NO ADVERSE PATIENT EFFECT AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11616405 ↗
  • MalfunctionFeb 10, 2017Positioning Failure

    ADDITIONAL INFORMATION: RETURN DEVICE ANALYSIS FOUND THAT THE STENT HOLDER (INNER MEMBER) OF THE ABSOLUTE PRO WAS RETURNED SEPARATED ON A VERSACORE 035" GUIDE WIRE. FOLLOW-UP WITH THE ACCOUNT CONFIRMED THAT THERE WAS RESISTANCE DURING REMOVAL OF THE STENT DELIVERY SYSTEM WITH THE GUIDE WIRE; HOWEVER, THEY COULD NOT RECALL ANYTHING ABOUT THE SEPARATION. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient