Skip to content

Absolute Pro 1012531-100Recalls & reported problems

Abbott Vascular Inc. · Absolute Pro Vascular Self-Expanding Stent System 9.0 mm x 100 mm x 80 cm / Over-The-Wire · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08717648175817), received between Feb 20, 2026 and Feb 20, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Deformation Due to Compressive Stress” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Deformation Due to Compressive Stress1
  • Difficult or Delayed Activation1
  • Difficult to Remove1
  • Improper or Incorrect Procedure or Method1
  • Malposition of Device1
  • Material Separation1
  • Physical Resistance/Sticking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 20, 2026Difficult to Remove; Material Separation; Improper or Incorrect Procedure or Method; Difficult or Delayed Activation; Malposition of Device; Deformation Due to Compressive Stress; Physical Resistance/Sticking

    IT WAS REPORTED THAT THE PROCEDURE PERFORMED WAS TO TREAT A DE NOVO, 70% STENOSED, MODERATELY CALCIFIED COMMON ILIAC ARTERY. THE 9.0X100MM ABSOLUTE PRO SELF-EXPANDING STENT SYSTEM (SESS) WAS ADVANCED; HOWEVER, DURING DEPLOYMENT THE THUMBWHEEL BECAME STUCK AND THE STENT ONLY PARTIALLY DEPLOYED. THE HANDLE WAS DISMANTLED IN AN ATTEMPT TO MANUALLY DEPLOY THE SES; HOWEVER, THIS WAS UNSUCCESSFUL. MULTIPLE ATTEMPTS WERE MADE TO REMOVE THE SESS, BUT RESISTANCE WAS MET WITH THE SHARP BIFURCATION AND THE SHEATH. DURING THE REMOVAL ATTEMPTS THE SHAFT KINKED AND SEPARATED HALF OUTSIDE THE BODY. FORCEPS WERE USED TO REMOVE THE SHAFT. THE SHEATH WAS PULLED BACK INTO THE AORTA AND THE STENT ULTIMATELY DEPLOYED, PARTIALLY IN AN UNINTENDED LOCATION OF THE AORTA AND PARTIALLY IN THE INTENDED AREA COVERING THE LESION TO THE PHYSICIAN'S SATISFACTION. THERE WERE NO ADVERSE PATIENT SEQUELAE AND NO CLINICALLY SIGNIFICANT DELAY IN PROCEDURE. FURTHER REVIEW OF THE PART AND LOT NUMBER FOUND THAT THE ABSOLUTE PRO SESS WAS USED AFTER EXPIRATION. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24405925 ↗

Other STENT, ILIAC devices

All STENT, ILIAC devices →