Absolute Pro 1012532-60Recalls & reported problems
Abbott Vascular Inc. · Absolute Pro Vascular Self-Expanding Stent System 10.0 mm x 60 mm x 80 cm / Over-The-Wire · Class 3 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08717648175855), received between Jan 15, 2019 and Apr 8, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Activation Failure” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Activation Failure1
- Difficult or Delayed Activation1
- Difficult to Remove1
- Improper or Incorrect Procedure or Method1
- Mechanical Jam1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 8, 2024Difficult to Remove; Improper or Incorrect Procedure or Method; Mechanical Jam; Activation Failure; Difficult to Remove; Improper or Incorrect Procedure or Method; Mechanical Jam; Activation Failure
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION IN THE EXTERNAL ILIAC ARTERY WITH MODERATE CALCIFICATION AND HEAVY TORTUOSITY. THE 10X60MM ABSOLUTE VASCULAR PRO SELF EXPANDING STENT SYSTEM (SESS) WAS ADVANCED TO THE TARGET LESION USING THE CROSSOVER APPROACH ON A .018 GUIDE WIRE. DURING STENT DEPLOYMENT THE THUMBWHEEL BECAME STUCK AFTER 1 CLICK AND THE STENT WAS ONLY ABLE TO BE PARTIALLY DEPLOYED. A .014 GUIDE WIRE WAS ADVANCED IN AN ATTEMPT TO STRAIGHTEN THE VERY TORTUOUS ANATOMY; HOWEVER, THE THUMBWHEEL REMAINED JAMMED. THE INTRODUCER SHEATH WAS PUSHED FORWARD TO COLLAPSE THE DEPLOYED STENT AND THE SESS WITH THE STENT WAS REMOVED FROM THE ANATOMY. IN THE PHYSICIAN'S OPINION, THE THUMBWHEEL BECAME STUCK DUE TO PATIENT'S VERY TORTUOUS ANATOMY. ANOTHER ABSOLUTE PRO STENT WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT AND NO CLINICALLY SIGNIFICANT DELAY REPORTED IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19068160 ↗MalfunctionJan 15, 2019Difficult or Delayed Activation; Difficult or Delayed Activation
IT WAS REPORTED THE PROCEDURE WAS TO TREAT A MILDLY TORTUOUS AND HEAVILY CALCIFIED LESION IN THE COMMON ILIAC. THE 10X80MM ABSOLUTE PRO VASCULAR SELF EXPANDING STENT SYSTEM (SESS) WAS ADVANCED TO THE LESION HOWEVER THE STENT WAS DIFFICULT TO DEPLOY. A 10X60MM ABSOLUTE PRO VASCULAR WAS ADVANCED AND WAS ALSO DIFFICULT TO DEPLOY. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE AND NO ADVERSE PATIENT EFFECTS. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8249076 ↗
Other STENT, ILIAC devices
- Cordis S.M.A.R.T. Control Vascular Stent System C10040MLCordis US Corp. · STENT, ILIAC1 recall3 reports
- Absolute Pro 1012534-100Abbott Vascular Inc. · STENT, ILIAC51 reports
- Protégé™ GPS™ SERB65-09-60-80Ev3, Inc · STENT, ILIAC1 recall0 reports
- Cordis S.M.A.R.T. Control Vascular Stent System C09060MLCordis US Corp. · STENT, ILIAC1 recall0 reports
- Absolute Pro 1012535-100Abbott Vascular Inc. · STENT, ILIAC1 death41 reports
- Omnilink Elite 1012631-39Abbott Vascular Inc. · STENT, ILIAC27 reports
- Omnilink Elite 1012632-29Abbott Vascular Inc. · STENT, ILIAC25 reports
- Omnilink Elite 1012631-59Abbott Vascular Inc. · STENT, ILIAC21 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page