Skip to content

Absolute Pro 1012532-60Recalls & reported problems

Abbott Vascular Inc. · Absolute Pro Vascular Self-Expanding Stent System 10.0 mm x 60 mm x 80 cm / Over-The-Wire · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08717648175855), received between Jan 15, 2019 and Apr 8, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Activation Failure” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Activation Failure1
  • Difficult or Delayed Activation1
  • Difficult to Remove1
  • Improper or Incorrect Procedure or Method1
  • Mechanical Jam1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 8, 2024Difficult to Remove; Improper or Incorrect Procedure or Method; Mechanical Jam; Activation Failure; Difficult to Remove; Improper or Incorrect Procedure or Method; Mechanical Jam; Activation Failure

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION IN THE EXTERNAL ILIAC ARTERY WITH MODERATE CALCIFICATION AND HEAVY TORTUOSITY. THE 10X60MM ABSOLUTE VASCULAR PRO SELF EXPANDING STENT SYSTEM (SESS) WAS ADVANCED TO THE TARGET LESION USING THE CROSSOVER APPROACH ON A .018 GUIDE WIRE. DURING STENT DEPLOYMENT THE THUMBWHEEL BECAME STUCK AFTER 1 CLICK AND THE STENT WAS ONLY ABLE TO BE PARTIALLY DEPLOYED. A .014 GUIDE WIRE WAS ADVANCED IN AN ATTEMPT TO STRAIGHTEN THE VERY TORTUOUS ANATOMY; HOWEVER, THE THUMBWHEEL REMAINED JAMMED. THE INTRODUCER SHEATH WAS PUSHED FORWARD TO COLLAPSE THE DEPLOYED STENT AND THE SESS WITH THE STENT WAS REMOVED FROM THE ANATOMY. IN THE PHYSICIAN'S OPINION, THE THUMBWHEEL BECAME STUCK DUE TO PATIENT'S VERY TORTUOUS ANATOMY. ANOTHER ABSOLUTE PRO STENT WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT AND NO CLINICALLY SIGNIFICANT DELAY REPORTED IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19068160 ↗
  • MalfunctionJan 15, 2019Difficult or Delayed Activation; Difficult or Delayed Activation

    IT WAS REPORTED THE PROCEDURE WAS TO TREAT A MILDLY TORTUOUS AND HEAVILY CALCIFIED LESION IN THE COMMON ILIAC. THE 10X80MM ABSOLUTE PRO VASCULAR SELF EXPANDING STENT SYSTEM (SESS) WAS ADVANCED TO THE LESION HOWEVER THE STENT WAS DIFFICULT TO DEPLOY. A 10X60MM ABSOLUTE PRO VASCULAR WAS ADVANCED AND WAS ALSO DIFFICULT TO DEPLOY. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE AND NO ADVERSE PATIENT EFFECTS. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8249076 ↗

Other STENT, ILIAC devices

All STENT, ILIAC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page