Skip to content

Absolute Pro 1012537-100Recalls & reported problems

Abbott Vascular Inc. · Absolute Pro Vascular Self-Expanding Stent System 9.0 mm x 100 mm x 135 cm/ Over-The-Wire · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08717648176258), received between Apr 19, 2019 and Nov 10, 2025 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Activation Failure” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Activation Failure1
  • Break1
  • Device Damaged by Another Device1
  • Material Separation1
  • Patient-Device Incompatibility1
  • Stretched1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 10, 2025Activation Failure; Activation Failure

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A CALCIFIED EXTERNAL ILIAC ARTERY. WHEN DEPLOYING A 9.0 X 100 MM ABSOLUTE PRO SELF-EXPANDING STENT, IT WAS ONLY ABLE TO PARTIALLY DEPLOY. THERE WAS NO ADVERSE PATIENT EFFECTS REPORTED AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23512712 ↗
  • MalfunctionAug 20, 2022Stretched; Device Damaged by Another Device; Stretched; Device Damaged by Another Device

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A HEAVILY CALCIFIED, 98% STENOSED LESION IN THE LEFT EXTERNAL ILIAC ARTERY. A 9X100MM ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM (SESS) WAS IMPLANTED WITHOUT ISSUE. DURING POST-DILATATION, WHILE ADVANCING WITH A 9X80MM ARMADA 35 BALLOON DILATATION CATHETER (BDC), THE PHYSICIAN PUSHED THE EDGE OF THE TIP OF THE ARMADA BDC AGAINST THE IMPLANTED ABSOLUTE PRO STENT CAUSING RESISTANCE BETWEEN THE DEVICES, AND THE ABSOLUTE PRO SESS BECAME STRETCHED. THE BDC WAS REMOVED AND REPLACED WITH A VIATRAC 14 PLUS PERIPHERAL BALLOON DILATATION CATHETER TO COMPLETE POST-DILATATION, WHEN THE BALLOON WAS INFLATED TO 24 ATMOSPHERES (ATM) AND RUPTURED ON THE FIRST INFLATION. THE VIATRAC BDC WAS REMOVED AND THE ARMADA BDC WAS RE-INSERTED TO SUCCESSFULLY COMPLETE THE POST-DILATATION. THE STRETCHED ABSOLUTE SESS REMAINS IMPLANTED. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15263210 ↗
  • MalfunctionApr 29, 2022Material Separation; Material Separation

    IT WAS REPORTED THAT DURING DEVICE PREPARATION, THE SYRINGE BECAME STUCK ON THE FLUSH PORT OF A 9X100MM ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM (SESS), AND WHEN THEY TRYTO TWIST THE SYRINGE OFF FROM THE FLUSH PORT, THE CATHETER SEPARATED FROM THE DEVICE HANDLE. ANOTHER ABSOLUTE PRO SESS WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14255909 ↗
  • InjuryDec 29, 2021Break; Break

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT THE LEFT EXTERNAL ILIAC ARTERY. THE 9.0X100MM ABSOLUTE PRO SELF-EXPANDING STENT SYSTEM (SESS) WAS SUCCESSFULLY ADVANCED TO THE TARGET LESION AND THE STENT WAS DEPLOYED WITH NO ISSUES. POST DEPLOYMENT, A MID STENT FRACTURE WAS OBSERVED. A 7.0X39 OMNILINK STENT WAS SUCCESSFULLY DEPLOYED TO CORRECT THE STENT FRACTURE AND REMAINS IN ONE PIECE. THERE WAS GOOD FLOW. THERE WAS NO ADVERSE PATIENT SEQUELA AND THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13112277 ↗
  • InjuryApr 19, 2019Patient-Device Incompatibility; Patient-Device Incompatibility

    IT WAS REPORTED THAT THE 9.0X100 MM ABSOLUTE PRO VASCULAR STENT WAS DEPLOYED IN THE PRE-DILATED, RIGHT EXTERNAL ILIAC VIA THE RADIAL ARTERY ACCESS SITE. AFTER THE STENT WAS DEPLOYED, THE DISTAL END OF THE STENT WAS NARROWED. THE PHYSICIAN FELT THIS WAS A DEVICE ISSUE AND NOT RELATED TO THE LESION ITSELF. AN 8.0X80 MM ARMADA 35 BALLOON DILATATION CATHETER WAS ADVANCED THROUGH THE STENT LUMEN TO POST DILATE THE ABSOLUTE PRO STENT AND FULLY EXPAND THE DISTAL END OF THE STENT. A HERCULINK ELITE STENT WAS ALSO DEPLOYED ON THE DISTAL END OF THE ABSOLUTE PRO STENT TO KEEP IT OPEN. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8532157 ↗